Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Organic compounds Peak area

With the abovementioned technique Hendriks et al. [30] estimated that road dust resuspension contribute around 10-15% of modelled PM10 on a European scale and up to 30% in densely populated area of South Europe (Fig. 5). These estimates are relative to the modelled concentrations, which are lower than observed ones due to uncertainties in primary organic emissions and the lack of secondary organic compounds. However, this overestimate is probably compensated by the fact that in the model, peak contributions (in cities) are not captured due to the model resolution (25 x 25 km). [Pg.181]

The next logical step toward chromophore design was to conduct a spectral survey of commercially available organic compounds in order to learn some general structure-property relationships for minimization of the residual absorbance. As an easily measured figure of merit, the ratio between the minimum and maximum molar absorptivities has been used. In many cases, this ratio (expressed in percent, or more conveniently, as the minimum molar absorptivity per 100,000 L/mol-cm of maximum absorbance) is 5-10% (5000-10,000 per 100,000). (The lower the number the better the dye.) An improved figure of merit would take into account the area under the absorption curve as well as the location of the transparent window relative to the peak in the absorption. This is tantamount to calculating the dispersion from the absorption spectrum, which was too complex for this type of survey. [Pg.210]

In organic compound analysis, the instrument response is expressed as a response factor (RF), which is the ratio of the concentration (or the mass) of the analyte in a standard to the area of the chromatographic peak. Conversely, a calibration factor (CF) is the ratio of the peak area to the concentration (or the mass) of the analyte. Equation 1, Appendix 22, shows the calculation of RF and CF. In trace element and inorganic compound analyses, the calibration curve is usually defined with a linear regression equation, and response (calibration) factors are not used for quantitation. [Pg.243]

Different polar organic solvents were tested as background electrolytes, and Af-methy 1 formamide was found to have the best properties with respect to both electrophoretic behaviour and high solubility of the interested compounds. The method was found to be precise (1.8% RSD for normalized peak areas), with good linearity and a low detection limit. [Pg.249]

Headspace-GC-MS analysis is useful for the determination of volatile compounds in samples that are difficult to analyze by conventional chromatographic means, e.g., when the matrix is too complex or contains substances that seriously interfere with the analysis or even damage the column. Peak area for equilibrium headspace gas chromatography depends on, e.g., sample volume and the partition coefficient of the compound of interest between the gas phase and matrix. The need to include the partition coefficient and thus the sample matrix into the calibration procedure causes serious problems with certain sample types, for which no calibration sample can be prepared. These problems can, however, be handled with multiple headspace extraction (MHE) [118]. Headspace-GC-MS has been used for studying the volatile organic compounds in polymers [119]. The degradation products of starch/polyethylene blends [120] and PHB [121] have also been identified. [Pg.126]

Example Ethanol is separated from a mixture of organic compounds by gas chromatography. The concentration of each component is proportional to its peak area. However, the chromatograph detector has a variable sensitivity from one run to the next. Is an internal standard required to determine the concentration of ethanol ... [Pg.4]

Calculate the area(s) under the peak for each volatile organic compound, and convert it to milligrams per kilograms of gamma-Cyclodextrin using the response factor for 8-cyclo-hexadecen-l-one. The response factor is determined from a calibration curve using 8-cyclohexadecen-l-one concentrations of 0.1 to 6 mg/100 mL of hexane. [Pg.130]


See other pages where Organic compounds Peak area is mentioned: [Pg.774]    [Pg.898]    [Pg.411]    [Pg.765]    [Pg.29]    [Pg.267]    [Pg.108]    [Pg.16]    [Pg.8]    [Pg.497]    [Pg.401]    [Pg.124]    [Pg.181]    [Pg.429]    [Pg.541]    [Pg.211]    [Pg.274]    [Pg.159]    [Pg.5]    [Pg.940]    [Pg.35]    [Pg.46]    [Pg.449]    [Pg.411]    [Pg.44]    [Pg.401]    [Pg.432]    [Pg.218]    [Pg.62]    [Pg.345]    [Pg.532]    [Pg.301]    [Pg.399]    [Pg.62]    [Pg.589]    [Pg.101]    [Pg.710]    [Pg.281]    [Pg.216]    [Pg.4948]    [Pg.205]    [Pg.86]    [Pg.641]   
See also in sourсe #XX -- [ Pg.136 ]




SEARCH



Peak areas

© 2024 chempedia.info