Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Narrow-bore columns for

Meehan, E., Oakley, S. A., Warner, F. P., (1992). Narrow Bore Columns for Size Exclusion Chromatography. Presented at Pittcon 92, New Orleans. [Pg.366]

Generally, size exclusion chromatography is carried out using columns with an internal diameter of 7.8 mm. However, some SEC applications require the use of expensive solvents. For this purpose, size exclusion columns with a smaller internal diameter (4.6 mm) have been developed. Of course one should use proportionally lower flow rates with narrow-bore columns. If the standard column size uses a flow rate of 1 ml/min, then the smaller 4.6-mm columns should be used at a flow rate of 0.35 ml/min. This provides the same linear velocity as 1 ml/min on 7.8-mm columns. The decreased flow rate reduces solvent consumption and solvent disposal cost. The performance of the smaller diameter columns is not compromised if properly optimized instrumentation is used. [Pg.333]

Narrow-bore columns are most useful for the analysis of polymers that are difficult to analyze in inexpensive solvents. However, if the appropriate equipment is available, good results can be obtained for a broad range of standard analyses. A comparison of an analysis of standards between an equivalent bank of conventional 7.8-mm and solvent efficient 4.6-mm columns is shown in Fig. 11.4. The columns used were Styragel HR 0.5, 1, 2, and 3 columns at 35°C with tetrahydrofuran (THF) as the solvent. The flow rate was 1 ml/min for the conventional columns (Fig. 11.4A) and 0.35 ml/min for the solvent-efficient 4.6-mm columns (Fig. 11.4B). If the correct equipment is available, the reduced solvent consumption of these solvent-efficient Styragel columns is of value to the environmentally conscious user. [Pg.334]

Sample loading must be reduced in accordance with the column inside diameter. Polymers exhibit high solution viscosity, and in order to avoid band broadening due to viscous streaming the sample concentration must be reduced for narrow-bore columns. Overloading effects become noticeable at much lower concentrations using 4.6-mm columns compared to 7.5-mm columns because of the effective sample concentration in a smaller volume column. [Pg.365]

Such effects principally cannot be observed in multi band detectors such as a UV diode array detector or a Fourier transform infrared (FTIR) detector because all wavelengths are measured under the same geometry. For all other types of detectors, in principle, it is not possible to totally remove these effects of the laminar flow. Experiments and theoretical calculations show (8) that these disturbances can only be diminished by lowering the concentration gradient per volume unit in the effluent, which means that larger column diameters are essential for multiple detection or that narrow-bore columns are unsuitable for detector combinations. Disregarding these limitations can lead to serious misinterpretations of GPC results of multiple detector measurements. Such effects are a justification for thick columns of 8-10 mm diameter. [Pg.441]

Where Q, is the minimum detectable amount, R the detector noise level and s the detector sensitivity [135,146,151,152]. For a concentration sensitive detector the minimum detectable concentration is the product of Q, and the volumetric gas flow rate through the detector. The minimum detectable amount or concentration is proportional to the retention time, and therefore, directly proportional to the column radius for large values of n. it follows, then, that very small quantities can be detected on narrow-bore columns. [Pg.29]

Note that the interfacing of LC techniques with MS puts significant constraints on the solvents that can be used i.e., they must be volatile, with a low salt concentration, for MS compatibility. Narrow-bore columns, which use much smaller amounts of salt and organic modifier, appear to have potential for facilitating IEC-MS applications.40 Despite the excellent sensitivity of MS detection for most elements, however, there are cases where matrix effects can interfere. In this situation, combination of IEC with atomic emission spectrometry (AES) or atomic absorption spectrometry (AAS) may be preferable, and can also provide better precision.21 32 4142 Other types of... [Pg.288]

RPLC-ELSD has been used for analysis of polyolefin additives [678], Improvements in sensitivity by use of narrow-bore columns were reported. MAE-RPLC-ELSD/UV (273 nm) has been proposed as a universal one-step and/or two-step procedure for the analysis of additives in polyolefins (Section 3.4.5.1), with the reference mixture of standard additives for... [Pg.249]

The present pneumatically assisted ESI interface is optimised around flow rates of 50-300 pl7min. The use of analytical columns 3. 6 mm i.d. with flow rates between 0.5 to 2 ml/min and narrow bore columns of 1-2 mm id with flow rates of 0.2-0.5 ml7min is routine in most pharmaceutical laboratories for HPLC analysis [24]. Capillary LC columns, because of their limited commercial availability and special practical considerations are used more where there is limited sample available or when sensitivity issues are present [25]. [Pg.164]


See other pages where Narrow-bore columns for is mentioned: [Pg.634]    [Pg.862]    [Pg.1861]    [Pg.246]    [Pg.634]    [Pg.862]    [Pg.1861]    [Pg.246]    [Pg.275]    [Pg.284]    [Pg.312]    [Pg.566]    [Pg.741]    [Pg.32]    [Pg.38]    [Pg.46]    [Pg.81]    [Pg.177]    [Pg.178]    [Pg.599]    [Pg.694]    [Pg.791]    [Pg.166]    [Pg.186]    [Pg.187]    [Pg.463]    [Pg.464]    [Pg.554]    [Pg.733]    [Pg.35]    [Pg.1239]    [Pg.250]    [Pg.264]    [Pg.363]    [Pg.162]    [Pg.259]    [Pg.70]    [Pg.213]    [Pg.484]    [Pg.73]    [Pg.453]    [Pg.506]    [Pg.519]    [Pg.27]   
See also in sourсe #XX -- [ Pg.308 ]




SEARCH



Borings

Narrow

© 2024 chempedia.info