Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Columns smaller-diameter

The efficiency obtained from an open tubular column can be increased by reducing the column radius, which, in turn will allow the column length to be decreased and, thus, a shorter analysis time can be realized. However, the smaller diameter column will require more pressure to achieve the optimum velocity and thus the reduction of column diameter can only be continued until the maximum available inlet pressure is needed to achieve the optimum mobile phase velocity. [Pg.388]

Superfractionation is an extension of distillation using smaller diameter columns and 100 or more trays to achieve reflux ratios exceeding 5 1. This equipment separates a narrow range aunponents such as of high-purity solvents, e.g., isoparaffins or individual aromatic compounds foi use. IS petrochemicals. [Pg.288]

Generally, size exclusion chromatography is carried out using columns with an internal diameter of 7.8 mm. However, some SEC applications require the use of expensive solvents. For this purpose, size exclusion columns with a smaller internal diameter (4.6 mm) have been developed. Of course one should use proportionally lower flow rates with narrow-bore columns. If the standard column size uses a flow rate of 1 ml/min, then the smaller 4.6-mm columns should be used at a flow rate of 0.35 ml/min. This provides the same linear velocity as 1 ml/min on 7.8-mm columns. The decreased flow rate reduces solvent consumption and solvent disposal cost. The performance of the smaller diameter columns is not compromised if properly optimized instrumentation is used. [Pg.333]

Smaller diameter columns are especially useful when expensive solvents are used. Figure 11.3 shows the analysis of poly (1,4-butylene terephthalate) using a Waters Alliance narrow-bore GPC system, quantitated against narrow polymethylmethacrylate standards. In this case, the solvent used is hexaflu-oro-2-isopropanol with 0.05 M sodium trifluoroacetic acid at a flow rate of... [Pg.333]

Sample loading must be reduced in accordance with the column inside diameter. Polymers exhibit high solution viscosity, and in order to avoid band broadening due to viscous streaming the sample concentration must be reduced for narrow-bore columns. Overloading effects become noticeable at much lower concentrations using 4.6-mm columns compared to 7.5-mm columns because of the effective sample concentration in a smaller volume column. [Pg.365]

Lower pressure drop and greater usable capacity allow use of smaller diameter columns. Pressure drop per stage is 30-50% less than either size Pall ring. [Pg.306]

We use the second-dimension separation from Fig. 6.6 with a 25 pL injection volume and 2.5 min sampling time the separation is an RPLC method that uses a monolithic column. Thus, 10 pL/min is the maximum flow rate in the first-dimension. Fig. 6.7 shows the development of the first-dimension column that utilizes a hydrophilic interaction (or HILIC) column for the separation of proteins at decreasing flow rates. The same proteins were separated in Fig. 6.6 (RPLC) and 6.7 (HILIC) and have a reversed elution order, which is known from the basics of HILIC (Alpert, 1990). It is believed that HILIC and RPLC separations are a good pair for 2DLC analysis of proteins as they appear to have dissimilar retention mechanisms, much like those of NPLC and RPLC it has been suggested that HILIC is similar in retention to NPLC (Alpert, 1990). Because the HILIC column used in Fig. 6.7 gave good resolution at 0.1 mL/min and no smaller diameter column was available, the flow was split 10-fold to match the second-dimension requirement... [Pg.141]

We believe that new developments (also including the commercial availability of smaller diameter columns that reduce solvent consumption and are more suitable for MS work) will also increase the interest in monolith columns. [Pg.328]

However, those who have worked with smaller diameter columns have often experienced lower performance and other difficulties. This is primarily due to extra-column effects the bandspreading in the injector, detector and tubing, or the gradient delay volume of the instrument. Troubleshooting guidelines for sorting out the causes of these difficulties are available in Reference f. With proper care, 2-mm columns can be run on a standard modern HPLC instrument with few difficulties. Smaller i.d. columns require special instrumentation. [Pg.91]

In the past, thermal conductivity detectors were most common in gas chromatography because they are simple and universal They respond to all analytes. Unfortunately, thermal conductivity is not sensitive enough to detect minute quantities of analyte eluted from open tubular columns smaller than 0.53 mm in diameter. Thermal conductivity detectors are still used for 0.53-mm columns and for packed columns. [Pg.542]

Use shorter columns with smaller-diameter particles. [Pg.568]


See other pages where Columns smaller-diameter is mentioned: [Pg.169]    [Pg.169]    [Pg.561]    [Pg.565]    [Pg.616]    [Pg.82]    [Pg.82]    [Pg.84]    [Pg.417]    [Pg.215]    [Pg.307]    [Pg.74]    [Pg.141]    [Pg.227]    [Pg.326]    [Pg.340]    [Pg.99]    [Pg.239]    [Pg.39]    [Pg.556]    [Pg.691]    [Pg.693]    [Pg.104]    [Pg.160]    [Pg.111]    [Pg.873]    [Pg.26]    [Pg.33]    [Pg.360]    [Pg.411]    [Pg.71]    [Pg.91]    [Pg.519]    [Pg.618]    [Pg.58]    [Pg.582]    [Pg.116]    [Pg.178]    [Pg.665]    [Pg.365]    [Pg.120]   
See also in sourсe #XX -- [ Pg.68 ]




SEARCH



Column diameter

Smaller

Smaller diameter

© 2024 chempedia.info