Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Narrow-band

Fig. 4a shows a characteristic narrow banded signal (860 kHz center frequency) from a flat steel surface (reference signal). A steel block was milled in a way that the distance of the upper and graved surface varied from 0 to about 1300 microns (Fig. 5). Moving the probe along the edge (see Fig. 5) about 30 signals have been acquired equidistantly (all 4 mm). Fig. 4b and 4c show two characteristic signals (position 6 and 12). The 30 measured signals have been preprocessed and deconvolved. Fig. 6 shows the evident correlation between measured TOF difference and signal position (depth of milled grave). Fig. 4a shows a characteristic narrow banded signal (860 kHz center frequency) from a flat steel surface (reference signal). A steel block was milled in a way that the distance of the upper and graved surface varied from 0 to about 1300 microns (Fig. 5). Moving the probe along the edge (see Fig. 5) about 30 signals have been acquired equidistantly (all 4 mm). Fig. 4b and 4c show two characteristic signals (position 6 and 12). The 30 measured signals have been preprocessed and deconvolved. Fig. 6 shows the evident correlation between measured TOF difference and signal position (depth of milled grave).
Due to the rather stringent requirements placed on the monochromator, a double or triple monocln-omator is typically employed. Because the vibrational frequencies are only several hundred to several thousand cm and the linewidths are only tens of cm it is necessary to use a monochromator with reasonably high resolution. In addition to linewidth issues, it is necessary to suppress the very intense Rayleigh scattering. If a high resolution spectrum is not needed, however, then it is possible to use narrow-band interference filters to block the excitation line, and a low resolution monocln-omator to collect the spectrum. In fact, this is the approach taken with Fourier transfonn Raman spectrometers. [Pg.1164]

Recently, the state-selective detection of reaction products tluough infrared absorption on vibrational transitions has been achieved and applied to the study of HF products from the F + H2 reaction by Nesbitt and co-workers (Chapman et al [7]). The relatively low sensitivity for direct absorption has been circumvented by the use of a multi-pass absorption arrangement with a narrow-band tunable infrared laser and dual beam differential detection of the incident and transmission beams on matched detectors. A particular advantage of probing the products tluough absorption is that the absolute concentration of the product molecules in a given vibration-rotation state can be detenuined. [Pg.2085]

In this approach one uses narrow-band continuous wave (cw) lasers for continuous spectroscopic detection of reactant and product species with high time and frequency resolution. Figure B2.5.11 shows an experimental scheme using detection lasers with a 1 MFIz bandwidth. Thus, one can measure the energy spectrum of reaction products with very high energy resolution. In practice, today one can achieve an uncertainty-limited resolution given by... [Pg.2128]

Recent tlieoretical [35, 36 and 37] and experimental [38] research has revealed anomalous behaviour of tire dimer anisotropy under certain excitation conditions. If tire dimer is excited by broadband light tliat covers botli excitonic transitions, or by a relatively narrow band properly positioned between tire maxima of tire excitonic transitions, tire... [Pg.3025]

In most experiments, light sourees have a "spread" of frequeneies assoeiated with them that is, they provide photons of various frequeneies. To eharaeterize sueh sourees, it is eommon to introduee the speetral souree funetion g(co) dco whieh gives the probability that the photons from this souree have frequeney somewhere between co and co+dco. For narrow-band lasers, g(co) is a sharply peaked funetion about some "nominal" frequeney cOq broader band light sourees have mueh broader g(co) funetions. [Pg.382]

In experimental measurements, sueh sharp 5-funetion peaks are, of eourse, not observed. Even when very narrow band width laser light sourees are used (i.e., for whieh g(co) is an extremely narrowly peaked funetion), speetral lines are found to possess finite widths. Let us now diseuss several sourees of line broadening, some of whieh will relate to deviations from the "unhindered" rotational motion model introdueed above. [Pg.429]

As orbitals spread into bands, orbitals oriented for a or a bonds spread into the widest bands. 7t orbitals form narrower bands and 5 bonding orbitals form the narrowest bands. [Pg.266]

At the beginning of a chromatographic separation the solute occupies a narrow band of finite width. As the solute passes through the column, the width of its band... [Pg.552]

A Quadrupole Assembly in Narrow Band-Pass Mode... [Pg.169]

Schematic diagram of an orthogonal Q/TOF instrument. In this example, an ion beam is produced by electrospray ionization. The solution can be an effluent from a liquid chromatography column or simply a solution of an analyte. The sampling cone and the skimmer help to separate analyte ions from solvent, The RF hexapoles cannot separate ions according to m/z values and are instead used to help confine the ions into a narrow beam. The quadrupole can be made to operate in two modes. In one (wide band-pass mode), all of the ion beam passes through. In the other (narrow band-pass mode), only ions selected according to m/z value are allowed through. In narrow band-pass mode, the gas pressure in the middle hexapole is increased so that ions selected in the quadrupole are caused to fragment following collisions with gas molecules. In both modes, the TOF analyzer is used to produce the final mass spectrum. Schematic diagram of an orthogonal Q/TOF instrument. In this example, an ion beam is produced by electrospray ionization. The solution can be an effluent from a liquid chromatography column or simply a solution of an analyte. The sampling cone and the skimmer help to separate analyte ions from solvent, The RF hexapoles cannot separate ions according to m/z values and are instead used to help confine the ions into a narrow beam. The quadrupole can be made to operate in two modes. In one (wide band-pass mode), all of the ion beam passes through. In the other (narrow band-pass mode), only ions selected according to m/z value are allowed through. In narrow band-pass mode, the gas pressure in the middle hexapole is increased so that ions selected in the quadrupole are caused to fragment following collisions with gas molecules. In both modes, the TOF analyzer is used to produce the final mass spectrum.
If the initial quadrupole analyzer is operated in its RF/DC mode (narrow band-pass mode), any one m/z value can be selected to pass right through the quadrupole analyzer. Ions of all other m/z values are shut out. [Pg.404]

A detailed examination of the correlation between Vj and M is discussed in references on analytical chemistry such as Ref. 6. We shall only outline the problem, with particular emphasis on those aspects which overlap other topics in this book. To consider the origin of the calibration curve, we begin by picturing a narrow band of polymer solution being introduced at the top of a solvent-filled column. The volume of this solvent can be subdivided into two categories the stagnant solvent in the pores (subscript i for internal) and the interstitial liquid in the voids (subscript v) between the packing particles ... [Pg.646]

As in all Fourier transform methods in spectroscopy, the FTIR spectrometer benefits greatly from the multiplex, or Fellgett, advantage of detecting a broad band of radiation (a wide wavenumber range) all the time. By comparison, a spectrometer that disperses the radiation with a prism or diffraction grating detects, at any instant, only that narrow band of radiation that the orientation of the prism or grating allows to fall on the detector, as in the type of infrared spectrometer described in Section 3.6. [Pg.59]

From about 1970, but before the availability of suitable lasers, Parmenter and others obtained SVLF spectra, particularly of benzene, using radiation from an intense high-pressure xenon arc source (see Section 3.4.4) and passing it through a monochromator to select a narrow band ca 20 cm wide) of radiation to excite the sample within a particular absorption band. [Pg.377]


See other pages where Narrow-band is mentioned: [Pg.722]    [Pg.813]    [Pg.1111]    [Pg.1122]    [Pg.1122]    [Pg.1312]    [Pg.1324]    [Pg.1477]    [Pg.1564]    [Pg.2420]    [Pg.2483]    [Pg.2484]    [Pg.3000]    [Pg.3039]    [Pg.3039]    [Pg.547]    [Pg.166]    [Pg.376]    [Pg.376]    [Pg.548]    [Pg.41]    [Pg.120]    [Pg.122]    [Pg.123]    [Pg.155]    [Pg.169]    [Pg.170]    [Pg.173]    [Pg.227]    [Pg.629]    [Pg.293]    [Pg.290]   
See also in sourсe #XX -- [ Pg.336 ]

See also in sourсe #XX -- [ Pg.340 ]




SEARCH



A Quadrupole Assembly in Narrow Band-Pass Mode

Absorption band narrowing

Band narrowing

Band narrowing

Band-gap-narrowed semiconductors

Bloch Conduction for Narrow-Band Polymers

Chemical band narrowing

Electronic polaron model band narrowing

Elution of a Narrow Band with Competitive Langmuir Isotherms

Exitance narrow band

Fatigue Damage Under Narrow-Band Random Processes

Filter narrow-band

Librational band model, narrowing

Load spectra narrow band

Narrow

Narrow Band Magnetism and Spin-Polarization

Narrow band approximation

Narrow band imaging

Narrow band materials

Narrow band noise

Narrow band sensor

Narrow band systems

Narrow band-gap polymers

Narrow-band emitters

Narrow-band pulses

Narrow-band random process

Narrow-band sizing

Narrow-band solids

Narrow-band transducer

Neutrino narrow-band beams

Polaron band narrowing effects

Radiation narrow band

Superconductivity narrow-band systems

The Band Description of Electrons in Narrow Bands

Ultra narrow band

Weak narrow bands

© 2024 chempedia.info