Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Measurements at low temperatures

Hence, it is necessary to correct the temperature change observed to the value it would have been if there was no leak. This is achieved by measuring the temperature of the calorimeter for a time period both before and after the process and applying Newton s law of cooling. This correction can be reduced by using the teclmique of adiabatic calorimetry, where the temperature of the jacket is kept at the same temperature as the calorimeter as a temperature change occurs. This teclmique requires more elaborate temperature control and it is prunarily used in accurate heat capacity measurements at low temperatures. [Pg.1901]

Figure C2.17.10. Optical absorjDtion spectra of nanocrystalline CdSe. The spectra of several different samples in the visible and near-UV are measured at low temperature, to minimize the effects of line broadening from lattice vibrations. In these samples, grown as described in [84], the lowest exciton state shifts dramatically to higher energy with decreasing particle size. Higher-lying exciton states are also visible in several of these spectra. For reference, the band gap of bulk CdSe is 1.85 eV. Figure C2.17.10. Optical absorjDtion spectra of nanocrystalline CdSe. The spectra of several different samples in the visible and near-UV are measured at low temperature, to minimize the effects of line broadening from lattice vibrations. In these samples, grown as described in [84], the lowest exciton state shifts dramatically to higher energy with decreasing particle size. Higher-lying exciton states are also visible in several of these spectra. For reference, the band gap of bulk CdSe is 1.85 eV.
Figure C3.1.10. (a) Steady state IR difference spectmm (dark minus light) of cytoclirome c oxidase CO complex measured at low temperature (127 K). This protein contains a copper atom situated immediately adjacent to a haem... Figure C3.1.10. (a) Steady state IR difference spectmm (dark minus light) of cytoclirome c oxidase CO complex measured at low temperature (127 K). This protein contains a copper atom situated immediately adjacent to a haem...
More recent determinations of Gfj by Freeman and his associates use the clearing field technique first introduced by Schmidt and Allen (1968). In this method, one collects the total free charge irrespective of mobility (vide infra). Robinson, et. al. (1971a,b) made measurements at low temperatures and confirmed the increase of Gfi with sphericity of the molecule (see Table 9.1). Fuochi and Freeman (1972) measured the yields in propane, methyl-substituted propane, and liquid Ar, obtaining a still lower value (2.0) in the last liquid. [Pg.288]

A detailed discussion of the problems encountered in vapour pressure measurements at low temperature is given in ref. [46,47], where also the use of an in situ manometer is described. Vapour pressure gas thermometry with other liquids besides He is discussed in ref. [43, p. 49],... [Pg.214]

It is often useful to carry out voltammetric measurements at low temperatures in order to evaluate both the stability of an electrogenerated species (the decrease in temperature will slow down the kinetics of any decomposition processes of the species formed in the electrode process) and the variation in formal electrode potential of a redox couple as a function of temperature. The latter point regards thermodynamic considerations of the redox processes, which will be discussed in Chapter 13, Section 3. [Pg.147]

As in the case of cyclic voltammetry, the electrolysis cell can be built with a thermostatic jacket to carry out measurements at low temperatures. In this case, the apparatus is of an isothermic type (i.e. the compartment containing the reference electrode is also cooled). In this case the most suitable reference electrode is the silver/silver chloride electrode filled by the same solution that will be used to dissolve the electroactive substance. One cannot use the saturated calomel electrode or the aqueous Ag/AgCl electrode because the KC1 (or NaCl) solution would freeze. [Pg.149]

Estimated from diffraction measurements at low temperature. In the temperature range investigated Qn xi close to 1.3 A [122]. [Pg.77]

Detection Wavelength Dependence of the Kinetics. The decay time of 100 ps measured at low temperature in the 665-nm anion band appears to be the same as the kinetics of the absorption changes in the 545-and 765-nm ground-state bands of BPh (Table 1). These results are consistent with the view that the BPh that receives an electron has its Ox Qy bs ds at 545 and 765 nm respectively (resolved at low... [Pg.215]

Improved spectra can be obtained through the cooling of the sample which often produces a more distinct splitting of degenerate bands. Bands which are weakly seen or seen only as shoulders at ambiant temperature often become quite apparent at lower temperatures. However, chalcogenometallates of the transition metals have seldom been measured at low temperatures. A few examples are given in (244) and (243). [Pg.106]

Table 1 Debye temperatures for some elemental solids and simple compounds. Data are obtained from thermal measurements at low temperature [25]... Table 1 Debye temperatures for some elemental solids and simple compounds. Data are obtained from thermal measurements at low temperature [25]...
Another reason why high angle reflections are better measured at low temperature is the decreased thermal diffuse scattering (see Sect. 2.2) which allows a more accurate integration of those intensities. [Pg.57]

Measurements at low temperature (1.3K). See Ref. 51 for more recent results, at room temperature. [Pg.102]

Particles produced in the gas phase must be trapped in condensed media, such as on solid substrates or in liquids, in order to accumulate, stock, and handle them. The surface of newly formed metallic fine particles is very active and is impossible to keep clean in an ambient condition, including gold. The surface must be stabilized by virtue of appropriate surface stabilizers or passivated with controlled surface chemical reaction or protected by inert materials. Low-temperature technique is also applied to depress surface activity. Many nanoparticles are stabilized in a solid matrix such as an inert gas at cryogenic temperature. At the laboratory scale, there are many reports on physical properties of nanometer-sized metallic particles measured at low temperature. However, we have difficulty in handling particles if they are in a solid matrix or on a solid substrate, especially at cryogenic temperature. On the other hand, a dispersion system in fluids is good for handling, characterization, and advanced treatment of particles if the particles are stabilized. [Pg.513]

Fig. 14. Product patterns as a function of temperature in neohexane/H2 reactions (a) Pure Ni measured at low temperatures (b) pure Ni diluted by Si02 and self-poisoned by the running reactions (notice that methane > neopentane, i.e., multiple reactions are also running at the lowest conversions) (c) alloy of Ni/Cu in the ratio 65 35 (increase in methane < decrease in neopentane this indicates that molecules other than methane are formed, i.e., the role of ay is larger here). From V. Ponec el at., Faraday Discuss. No. 72, p. 33. Fig. 14. Product patterns as a function of temperature in neohexane/H2 reactions (a) Pure Ni measured at low temperatures (b) pure Ni diluted by Si02 and self-poisoned by the running reactions (notice that methane > neopentane, i.e., multiple reactions are also running at the lowest conversions) (c) alloy of Ni/Cu in the ratio 65 35 (increase in methane < decrease in neopentane this indicates that molecules other than methane are formed, i.e., the role of ay is larger here). From V. Ponec el at., Faraday Discuss. No. 72, p. 33.
Fig. 4 Differential conductance dl/dV versus applied voltage V at 100 K. The differential conductance manifests a clear peak structure. Good reproducibility can be seen from the six nearly overlapping curves. Peak structures were observed in four samples measured at low temperatures although details were different from sample to sample. Subsequent sets of I-V measurements can show a sudden change, possibly due to conformational changes of the DNA. The inset shows an example of two typical I-V curves that were measured before and after such an abrupt change. Switching between stable and reproducible shapes can occur upon an abrupt switch of the voltage or by high current (from [14], with permission Copyright 2000 by Nature Macmillan Publishers Ltd)... Fig. 4 Differential conductance dl/dV versus applied voltage V at 100 K. The differential conductance manifests a clear peak structure. Good reproducibility can be seen from the six nearly overlapping curves. Peak structures were observed in four samples measured at low temperatures although details were different from sample to sample. Subsequent sets of I-V measurements can show a sudden change, possibly due to conformational changes of the DNA. The inset shows an example of two typical I-V curves that were measured before and after such an abrupt change. Switching between stable and reproducible shapes can occur upon an abrupt switch of the voltage or by high current (from [14], with permission Copyright 2000 by Nature Macmillan Publishers Ltd)...
The reported ESR spectra of polyisobutylene, irradiated with high energy radiation and measured at low temperatures (31, 35, 47, 48, 49, 52) are similar in form, showing a broad doublet with a hyperfine splitting constant of 20 gauss. These spectra have been interpreted as caused by radicals (XIII) formed by hydrogen abstraction from the main chain methylene groups ... [Pg.273]

The life-time of the ionic intermediates in the liquid solutions is so short that it is difficult or impossible to observe them directly. One of the convenient ways to overcome this difficulty is to prolong the life-time by freezing the irradiated solution at low temperature. For convenience of measurement, organic substances frozen to a glassy state have been widely used as solvents for the study by means of optical absorption measurement at low temperature. A rigid glass matrix is preferably used also in the ESR study, because the experimental results are readily compared with those obtained by the optical study. [Pg.403]

This calculation is based on the assumptions that both the standard and the sample are completely liquid at 60°C, and that the standard is completely liquid at the final measurement temperature. Although these assumptions are probably justified in many cases, they may not always be correct, particularly when making measurements at low temperatures. As a general guideline, liquid oils are clear, so the reference should be clear at the measurement temperature, and both the reference and the samples should be clear at 60°C. If the (filtered) samples are at all turbid at 60°C, it may be possible to raise the reference temperature as noted in step 4. If the reference oil is turbid at the measurement temperatu re, consult the manufacturer of the NMR spectrometer for an alternative reference. [Pg.571]

It should be noted that the use of the simple dependence (4.2) with constant L is possible only atp< pcc t. If the pressure obtained from formula (4.5) becomes comparable to pcrit, it is necessary to use in the calculation a refined curve of vapor pressure. As we know, the evaporation heat decreases as the temperature increases formulas (4.4), (4.5) with constant L measured at low temperature overestimate the limiting pressure. Finally, if the pressure found exceeds the critical pressure, above which the transition from liquid to gas occurs continuously, the theory developed here is inapplicable. [Pg.348]


See other pages where Measurements at low temperatures is mentioned: [Pg.125]    [Pg.119]    [Pg.428]    [Pg.278]    [Pg.70]    [Pg.224]    [Pg.192]    [Pg.271]    [Pg.233]    [Pg.23]    [Pg.75]    [Pg.216]    [Pg.37]    [Pg.103]    [Pg.103]    [Pg.104]    [Pg.166]    [Pg.19]    [Pg.479]    [Pg.157]    [Pg.196]    [Pg.141]    [Pg.81]    [Pg.36]    [Pg.270]    [Pg.229]    [Pg.298]    [Pg.299]    [Pg.498]    [Pg.305]   
See also in sourсe #XX -- [ Pg.156 , Pg.157 , Pg.158 , Pg.159 ]




SEARCH



Low-temperature measurements

Temperature at low

Temperature measurement

© 2024 chempedia.info