Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Mass spectrometer readings

Figure 5 gives the simulation results with the model given for the conditions used by Briggs et al. to obtain Fig. 3. Data points are shown in Fig. 5b, but not in 5a. Mass spectrometer readings were not calibrated, and only normalized data are shown in Fig. 3a. The simulation estimates the shape of the midbed temperature and the SO3 vol% variations successfully. It also reproduces the initial bed temperature lag for the first minute after introduction of the S03/S02 reactant mixture (Fig. 5b), as well as the absence of a lag when air is introduced to the catalyst bed displacing the reactant mixture (Fig. 5a). The model also gives the slow adjustment of the bed temperature after the maximum and minimum temperatures, although the rates of cooling and heating are not correct. The most serious deficiency of the model is that it overestimates the temperature rise and drop by 15 and 8°C, respectively. Figure 5 gives the simulation results with the model given for the conditions used by Briggs et al. to obtain Fig. 3. Data points are shown in Fig. 5b, but not in 5a. Mass spectrometer readings were not calibrated, and only normalized data are shown in Fig. 3a. The simulation estimates the shape of the midbed temperature and the SO3 vol% variations successfully. It also reproduces the initial bed temperature lag for the first minute after introduction of the S03/S02 reactant mixture (Fig. 5b), as well as the absence of a lag when air is introduced to the catalyst bed displacing the reactant mixture (Fig. 5a). The model also gives the slow adjustment of the bed temperature after the maximum and minimum temperatures, although the rates of cooling and heating are not correct. The most serious deficiency of the model is that it overestimates the temperature rise and drop by 15 and 8°C, respectively.
Desorption was monitored with mass spectroscopy. The cracking patterns of 2-propanol, acetone, and propene were individually determined ( ). For quantitative analysis, masses 45, 45, 41, 18, and 2 were used for 2-propanol, acetone, propene, water, and hydrogen, respectively, after correction for cracking in a similar procedure as described (52 ) The mass spectrometer sensitivities were determined to be 5.26, 7 88, 5.07, 4 74, and 5.20 amp/torr, and the pumping speeds were 9.5, 15.1, 51.0, 1.7, 56.9 L sec"", respectively for the five species. These two latter quantities were used to convert the mass spectrometer readings into molecular fluxes. [Pg.208]

Not all the allowable molecular formulas in column (5) represents stable chemical substances. Nevertheless, they might possibly show up in a mass spectrometer reading, which details not only stable compounds but also ions and decomposition fragments.)... [Pg.30]

Mass spectrometer reading of gases after pumpdown and during processing. [Pg.134]

This chapter should be read in conjunction with Chapter 3, Electron Ionization. In electron ionization (El), a high vacuum (low pressure), typically 10 mbar, is maintained in the ion source so that any molecular ions (M +) formed initially from the interaction of an electron beam and molecules (M) do not collide with any other molecules before being expelled from the ion source into the mass spectrometer analyzer (see Chapters 24 through 27, which deal with ion optics). [Pg.1]

Most mass spectrometers for analytical work have access to a large library of mass spectra of known compounds. These libraries are in a form that can be read immediately by a computer viz., the data corresponding to each spectrum have been compressed into digital form and stored permanently in memory. Each spectrum is stored as a list of m/z values for all peaks that are at least 5% of the height of the largest peak. To speed the search process, a much shorter version of the spectrum is normally examined (e.g., only one peak in every fourteen mass units). [Pg.323]

This chapter is best read in conjunction with Chapter 35, Gas Chromatography and Liquid Chromatography, and any other chapters appropriate to the operation of a mass spectrometer. [Pg.414]

Special isotope ratio mass spectrometers are needed to measure the small variations, which are too small to be read off from a spectrum obtained on a routine mass spectrometer. Ratios of isotopes measured very accurately (usually as 0/00, i.e., as parts per 1000 [mil] rather than parts per 100 [percent]) give information on, for example, reaction mechanisms, dating of historic samples, or testing for drugs in metabolic systems. Such uses are illustrated in the main text. [Pg.425]

Tossing a mental coin, the decision was to analyze the case of noise proportional to the square root of the signal. This, as you will recall, is Poisson-distributed noise, characteristic of the noise encountered when the limiting noise source is the shot noise that occurs when individual photons are detected and represent the ultimate sensitivity of the measurement. This is a situation that is fairly commonly encountered, since it occurs, as mentioned previously, in UV-Vis instrumentation as well as in X-ray and gamma-ray measurements. This noise source may also enter into readings made in mass spectrometers, if the detection method includes counting individual ions. We have, in... [Pg.285]

Fig. 7. Protein identification with electrospray tandem mass spectrometry and a triple quadrupole mass spectrometer. Fragment spectra of several peptides are generated during one investigation. From the fragment spectra short sequence stretches can be read. Together with their mass location in the peptide of the measured mass, they can be used to specifically identify a protein in the database. Because the protein identification depends only on one peptide, several proteins can be identified from one sample. Fig. 7. Protein identification with electrospray tandem mass spectrometry and a triple quadrupole mass spectrometer. Fragment spectra of several peptides are generated during one investigation. From the fragment spectra short sequence stretches can be read. Together with their mass location in the peptide of the measured mass, they can be used to specifically identify a protein in the database. Because the protein identification depends only on one peptide, several proteins can be identified from one sample.
The properties of an ideal mass analyzer are well described, [2] but despite the tremendous improvements made, still no mass analyzer is perfect. To reach a deeper insight into the evolution of mass spectrometers the articles by Beynon, [3] Habfast and Aulinger, [4,5] Brunnee [6,7], Chapman et al. [8] and McLuckey [9] are recommended for further reading. In recent years, miniature mass analyzers have gained interest for in situ analysis, [10] e.g., in environmental [11] or biochemical applications, [12] for process monitoring, for detection of chemical warfare agents, for extraterrestrial applications, [13] and to improve Space Shuttle safety prior to launch. [14]... [Pg.112]

Dudenbostel and Priestly (D2) have reviewed the importance of the mass spectrometer in the petroleum industry, giving a brief account of the computational work involved. They also describe recent developments which have made it possible for readings from the spectrometer to be converted automatically into digital output. This output may then be fed either directly to a digital computer or through the medium of punched cards. In either case human intervention is minimized, with... [Pg.343]

Above all, it is worth remembering that if insufficient attention is paid to sample preparation then the validity of an analysis will be seriously undermined [20] no matter how sophisticated the measurement instrument is, whether it be a mass spectrometer or sensor. Comments on, and reviews of modem methods of sample preparation such as those of Jinno [21] and Pawliszyn [22,23] are essential reading whether one is a developer of sensors or an instrumental analyst. [Pg.670]

Figure 19 Comparison of the reading by a Pirani gauge (1) with the output of a mass spectrometer (2). (From [19].)... Figure 19 Comparison of the reading by a Pirani gauge (1) with the output of a mass spectrometer (2). (From [19].)...
England J. G., Zindler A., Reisberg L. C., Rubenstone J. L., Salters V., Marcantonio F., Bourdon B., Brueckner H., Turner P. J., Weaver S., and Read P. (1992) The Lamont-Doherty-Geological-Observatory Isolab-54 isotope ratio mass-spectrometer. Int. J. Mass Spectrom. Ion Process. 121, 201-240. [Pg.1766]

The instrumentation employed by Stapleton and Bowie was an ion cyclotron resonance mass spectrometer that had been modified to permit computer control of all drift voltages and to allow direct reading of the ion transit time (typically 10 to 10 seconds) similar to that described by McMahon and Beauchamp An emission current of about 0.2 microampere and a nominal 70 eV electron beam produced ion currents of 10 to 10 A at source pressures of approximately 10 Torr. The mass spectra were measured by magnetic field modulation. [Pg.123]


See other pages where Mass spectrometer readings is mentioned: [Pg.298]    [Pg.97]    [Pg.298]    [Pg.97]    [Pg.322]    [Pg.421]    [Pg.376]    [Pg.35]    [Pg.231]    [Pg.56]    [Pg.208]    [Pg.662]    [Pg.670]    [Pg.678]    [Pg.686]    [Pg.29]    [Pg.106]    [Pg.112]    [Pg.15]    [Pg.9]    [Pg.152]    [Pg.162]    [Pg.10]    [Pg.641]    [Pg.1665]    [Pg.112]    [Pg.143]    [Pg.22]    [Pg.376]    [Pg.316]    [Pg.344]    [Pg.3]    [Pg.197]   
See also in sourсe #XX -- [ Pg.91 ]




SEARCH



© 2024 chempedia.info