Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Light source modulated

Light valve A projection device that employs a fixed light source modulated by an optical-valve intervening the source and projection optics. [Pg.576]

SymApps converts 2D structures From the ChemWindow drawing program into 3D representations with the help of a modified MM2 force field (see Section 7.2). Besides basic visualization tools such as display styles, perspective views, and light source adjustments, the module additionally provides calculations of bond lengths, angles, etc, Moreover, point groups and character tables can be determined. Animations of spinning movements and symmetry operations can also he created and saved as movie files (. avi). [Pg.147]

For characterization purposes of bulk or thin-film semiconductors the features at Eq and E] are the most useflil. In a number of technologically important semiconductors (e.g., Hgi j d Te, and In Gai j ) the value of. ) is so small that it is not in a convenient spectral range for Modulation Spectroscopy, due to the limitations of light sources and detectors. In such cases the peak at E can be used. The features at Eq and are not useflil since they occur too far into the near-ultraviolet and are too broad. [Pg.388]

A schematic representation of a PR apparatus is shown in Figure 2. In PR a pump beam (laser or other light source) chopped at frequency 2 creates photo-injected electron-hole pairs that modulate the built-in electric field of the semiconductor. The photon energy of the pump beam must be larger than the lowest energy gap of the material. A typical pump beam for measurements at or below room temperature is a 5-mW He-Ne laser. (At elevated temperatures a more powerful pump must be employed.)... [Pg.389]

Light from an appropriate light source (a xenon arc or a halogen or tun ten lamp) passes through a monochromator (probe monochromator). The exit intensity at wavelength "k, IqCK), is focused onto the sample by means of a lens (or mirror). Tbe reflected light is collected by a second lens (mirror) and focused onto an appropriate detector (photomultiplier, photodiode, etc.). For simplicity, the two lenses (mirrors) are not shown in Figure 2. For modulated transmission the detector is placed behind the sample. [Pg.389]

The light produced by a laser has a much more narrow wavelength than the light of an LED or other light sources (Fig. 15.3). In addition, laser light is coherent, i.e., the photons travel in parallel paths from the source. Lasers made of thin-films are similar to bulk lasers (He-Ne, ruby) except that they are more compact and efficient. Due to the short lifetime of the photons, high-frequency modulations are possible. [Pg.391]

Pulsed method. Using a pulsed or modulated excitation light source instead of constant illumination allows investigation of the time dependence of emission polarization. In the case of pulsed excitation, the measured quantity is the time decay of fluorescent emission polarized parallel and perpendicular to the excitation plane of polarization. Emitted light polarized parallel to the excitation plane decays faster than the excited state lifetime because the molecule is rotating its emission dipole away from the polarization plane of measurement. Emitted light polarized perpendicular to the excitation plane decays more slowly because the emission dipole moment is rotating towards the plane of measurement. [Pg.189]

With a modulated light source, the phase difference between the perpendicular and parallel components of the emission is measured. The time delay of the perpendicular component of emission is longer than that of the parallel component because the molecule requires a certain period of time to rotate into the perpendicular orientation where it can be detected through the perpendicular polarizer. [Pg.189]

Luminescence lifetimes are measured by analyzing the rate of emission decay after pulsed excitation or by analyzing the phase shift and demodulation of emission from chromophores excited by an amplitude-modulated light source. Improvements in this type of instrumentation now allow luminescence lifetimes to be routinely measured accurately to nanosecond resolution, and there are increasing reports of picosecond resolution. In addition, several individual lifetimes can be resolved from a mixture of chromophores, allowing identification of different components that might have almost identical absorption and emission features. [Pg.259]

FLIM systems can be purchased as an add-on for a standard fluorescence microscope. Such a system will consist of a CCD camera coupled to a modulatable image intensifier, an LED light source, and driver electronics. This system will modulate the LED and image intensifier while shifting the phase between them as it takes a series of images (Fig. 2.1). [Pg.81]

This procedure involves selecting a fluorophore of known lifetime and placing it in the microscope and measuring the phase and modulation depth [11]. Rearranging Eqs. (2.5 and 2.6) allows the expected phase and modulation to be predicted. These may then be used to compute the position of zero phase and the modulation depth of the light source. An advantage of the method is that it may be done under conditions exactly matching those of a sample. [Pg.88]

Monochromatic light can also be obtained from other types of lasers solid state, gas, ion, dye. Among them argon ion laser with its many lines is an especially valuable light source used in many sensors. However, these types of lasers are expensive, the modulation of the light cannot be done internally and external modulators (e.g. choppers) should be used. Wavelengths emitted by some exemplary lasers are presented in Table 1. [Pg.53]

A problem encountered with atomic absorption is that emission from the flame may fall on the detector and be registered as negative absorption. This can be eliminated by modulating the light source, either mechanically or electronically, and using an a.c. detector tuned to the frequency of modulation of the source. D. C. radiation, such as emission from the flame, will then not be detected. A high intensity of emission, however, may overload the detector, causing noise fluctuations. [Pg.84]

Figure 3.1 Schematic diagram of an AAS spectrometer. A is the light source (hollow cathode lamp), B is the beam chopper (see Fig. 3.2), C is the burner, D the monochromator, E the photomultiplier detector, and F the computer for data analysis. In the single beam instrument, the beam from the lamp is modulated by the beam chopper (to reduce noise) and passes directly through the flame (solid light path). In a double beam instrument the beam chopper is angled and the rear surface reflective, so that part of the beam is passed along the reference beam path (dashed line), and is then recombined with the sample beam by a half-silvered mirror. Figure 3.1 Schematic diagram of an AAS spectrometer. A is the light source (hollow cathode lamp), B is the beam chopper (see Fig. 3.2), C is the burner, D the monochromator, E the photomultiplier detector, and F the computer for data analysis. In the single beam instrument, the beam from the lamp is modulated by the beam chopper (to reduce noise) and passes directly through the flame (solid light path). In a double beam instrument the beam chopper is angled and the rear surface reflective, so that part of the beam is passed along the reference beam path (dashed line), and is then recombined with the sample beam by a half-silvered mirror.
Instruments that utilize a flame for the atomizer can also utilize the graphite furnace. It is a matter of replacing the flame module with the furnace module and lining it up with the light beam. The same light source (usually a hollow cathode lamp) and the same optical path are used. See Figure 9.14 and compare with Figure 9.10. [Pg.259]


See other pages where Light source modulated is mentioned: [Pg.75]    [Pg.6]    [Pg.147]    [Pg.464]    [Pg.75]    [Pg.6]    [Pg.147]    [Pg.464]    [Pg.2498]    [Pg.37]    [Pg.392]    [Pg.315]    [Pg.372]    [Pg.267]    [Pg.294]    [Pg.422]    [Pg.447]    [Pg.8]    [Pg.537]    [Pg.525]    [Pg.24]    [Pg.24]    [Pg.261]    [Pg.14]    [Pg.76]    [Pg.79]    [Pg.82]    [Pg.83]    [Pg.86]    [Pg.411]    [Pg.459]    [Pg.54]    [Pg.265]    [Pg.272]    [Pg.461]    [Pg.465]    [Pg.248]    [Pg.324]    [Pg.290]    [Pg.215]   
See also in sourсe #XX -- [ Pg.57 ]




SEARCH



Light modulators

Light sources

Modulation light

Source modulation

© 2024 chempedia.info