Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Intensive responses

A quantitative correlation between the charges under the current and mass intensity signals can be carried out as suggested by Heitbaum and Wolter [11]. The magnitude of the mass intensity response depends not only on the electrochemical properties of the system under study but also on the permeability of the electrode to the volatile products in addition to mass spectrometer parameters. A calibration of the actual experimental setup is therefore necessary. The proportionality between mass intensity (MI), and faradaic current (/) can be formulated as follows ... [Pg.129]

In an attempt to avoid the need for derivatisation Rinehart and Cook -6 applied the technique of Field Desorption Mass Spectrometry to neomycin and successfully observed an intense response for the molecular ion at m/e 615(M + H) together with very small peaks at m/e 455, 307 and 206 representing the loss of sugar fragments. [Pg.407]

The value of an intensive response is not a function of the size or throughput of the system. Product purity is an example of an intensive response. If a manufacturing facility can consistently produce 95% pure material, then it will be 95% pure, whether we look at a pound or a ton of the material. Other examples of intensive responses might be color, density, percent yield, alcohol content, and flavor. [Pg.14]

Chemical image data sets are visualized as a three-dimensional cube spanning one wavelength and two spatial dimensions called a hypercube (Figure 7.2). Each element within the cube contains the intensity-response information measured at that spatial and spectral index. The hypercube can be treated as a series of spatially resolved spectra (called... [Pg.195]

Figure 2.3 Sensitivities of rods and cones of the mudpuppy (a large aquatic salamander) retina. The graph shows the intensity-response curves, which were measured intracellularly from the retina of a dissected eye. The response is shown as a fraction of the response at saturation. A fit to the data is also shown. Rods are approximately 25 times more sensitive than cones. (Reprinted with permission from Gordon L. Fain and John E. Dowling. Intracellular recordings from single rods and cones in the mudpuppy retina. Science, Vol. 180, pp. 1178-1181, June, Copyright 1973 AAAS). Figure 2.3 Sensitivities of rods and cones of the mudpuppy (a large aquatic salamander) retina. The graph shows the intensity-response curves, which were measured intracellularly from the retina of a dissected eye. The response is shown as a fraction of the response at saturation. A fit to the data is also shown. Rods are approximately 25 times more sensitive than cones. (Reprinted with permission from Gordon L. Fain and John E. Dowling. Intracellular recordings from single rods and cones in the mudpuppy retina. Science, Vol. 180, pp. 1178-1181, June, Copyright 1973 AAAS).
Figure 2.8 Light- and dark-adapted intensity-response curves recorded intracellularly from a red cone of turtle retina. The dark-adapted potential is 0 mV. This shift of the response curve enables the visual system to discriminate small increments or decrements about the background level. (Reproduced from R. A. Normann and I. Perlman. The effects of background illumination on the photoresponses of red and green cones. Journal of Physiology, Vol. 286, pp. 491-507, 1979, by permission of Blackwell Publishing, UK.)... Figure 2.8 Light- and dark-adapted intensity-response curves recorded intracellularly from a red cone of turtle retina. The dark-adapted potential is 0 mV. This shift of the response curve enables the visual system to discriminate small increments or decrements about the background level. (Reproduced from R. A. Normann and I. Perlman. The effects of background illumination on the photoresponses of red and green cones. Journal of Physiology, Vol. 286, pp. 491-507, 1979, by permission of Blackwell Publishing, UK.)...
When the two-channel state is manipulated so that they come up different from each other and thereafter channeled to another two-slit device, interference cannot be necessarily expected the relative intensity response at a given point is determined by the numeric value of the modulus square amplitude. Whenever a mechanism is set up to restore both channels quantum state sameness, then interference effects will show up. The way one stores (or restore) such sameness is the key to elicit interferences. Coherence can be lost and regained this issue is discussed below. For an experimental case, the reader is referred to Chapman s et al. [16]. [Pg.63]

The first event may happen anywhere on the TV screen you can prepare the system as many times as you want and check that the first event appears localized (almost) at random this randomness is only apparent if you use the theory presented here. What has happened was a change in amplitudes for a transition from state +) to —) by capturing energy from the I-frame system the relative coherent intensity response being ... [Pg.71]

This entangled quantum state states us the available possibilities. Probing at b-position will yield a response from either nm,A.mb) or , A.nb). The relative intensity responses are equal to 1/2. Moreover, Eq. (29) imposes correlations at a-position. The probing device will register correlations implied by the quantum state (Cf. Eq. (3)) the probing device for this case cannot impose these correlations. [Pg.80]

Detector D1 interacts with quantum state [1A/2 0]t yielding relative intensity response 11 /v/212 = 1/2. [Pg.84]

Detector D2 interacts with quantum state [0 -1 /v/2] yielding relative intensity response —1 / /2 2 = 1/2. [Pg.84]

The measured CARS signal Scoh is proportional to the time integral over the absolute value squared of the total third-order polarization, P = Piso + Paniso + Pnr> because of the slow intensity response of the detector ... [Pg.20]

Mean + SEM for the mouse N40 amplitude stimulus intensity response is shown (Reprinted from Maxwell et al., 2006a, with permission from Macmillan)... [Pg.532]

For the diffraction technique, —325-mesh shale powders were packed into the 7/8-in. diameter by 1/16-in. deep cavity of a 1-in. diameter aluminum holder and pressed smooth with a glass plate. A Norelco-Phillips diffractometer equipped with a Cu(Ka) source was used to analyze the samples. The diffractometer was driven at l°/min and the intensity response of the nonrotating sample recorded on a strip chart recorder. Peak assignments in both cases were made by visual comparisons with standards. [Pg.182]

Clofentazine can be analysed in positive-ion APCI, but a more intense response can be achieved in negative-ion electron-capture conditions, resulting in M"" [82],... [Pg.191]

To compound the problem, marcasite for reasons unknown to the author does not show the intensity response, volume for volume,as does pyrite. It has been the author s experience that coals that have been shown by optical examination to contain marcasite in significant concentrations show almost no indication of the mineral being present on a diffractogram generated from the low temperature ash. [Pg.58]

The results show that concentration of 1% ethanol could be detected within less than 10 min, and concentration as low as 1.6 ppm phenol could be detected within less than 6 min. Different intensity response of the various bacterial sensors, dnaK,... [Pg.173]


See other pages where Intensive responses is mentioned: [Pg.90]    [Pg.150]    [Pg.155]    [Pg.297]    [Pg.405]    [Pg.408]    [Pg.226]    [Pg.92]    [Pg.58]    [Pg.443]    [Pg.170]    [Pg.251]    [Pg.233]    [Pg.17]    [Pg.19]    [Pg.38]    [Pg.90]    [Pg.196]    [Pg.95]    [Pg.28]    [Pg.65]    [Pg.226]    [Pg.319]    [Pg.395]    [Pg.662]    [Pg.90]    [Pg.146]    [Pg.103]    [Pg.58]    [Pg.781]    [Pg.201]    [Pg.222]    [Pg.223]   
See also in sourсe #XX -- [ Pg.14 ]




SEARCH



Intensity-response function

Intensity/energy response function

© 2024 chempedia.info