Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Intensity setting

Figure 2.19 A detailed plot of the spectrum from a Cu X-ray tube, in the vicinity of the K lines, showing the area selected by the high resolution and high intensity settings of the beam conditioner shown in Figure 2,20... Figure 2.19 A detailed plot of the spectrum from a Cu X-ray tube, in the vicinity of the K lines, showing the area selected by the high resolution and high intensity settings of the beam conditioner shown in Figure 2,20...
The intensity setting of an input is called a level. It is possible for an input to be at different levels at different times. Thus, if x designates the x in the previous algebraic example, it might have had the value 0 when we were first interested in the system now we might want x to have the value 5. To designate these different sets of conditions, a second subscript is added. Thus, x, = 0 means that in the first instance, x, = 0 and x,2 = 5 means that in the second instance, x, = 5. [Pg.3]

Figure 7 Typical surface temperature profile for colored samples exposed to outdoor sunlight (45° south facing-backed exposure), xenon long-arc radiation (intensity set to 0.55 W/m2 at 340 nm, filtered to mimic outdoor sunlight), and a fluorescent UVA-340 lamp. Abbrevlatlorr. UV, ultraviolet. Source Courtesy of Atlas Materials Testing Technology, LLC, Chicago, IL, U.S.A. Figure 7 Typical surface temperature profile for colored samples exposed to outdoor sunlight (45° south facing-backed exposure), xenon long-arc radiation (intensity set to 0.55 W/m2 at 340 nm, filtered to mimic outdoor sunlight), and a fluorescent UVA-340 lamp. Abbrevlatlorr. UV, ultraviolet. Source Courtesy of Atlas Materials Testing Technology, LLC, Chicago, IL, U.S.A.
Approach One is UV intensity set-point approach. In this approach, measurements made by the UV intensity sensor are used to control the UV reactor. The sensor is located in a position that allows it to properly respond to both changes in UV output of the lamps and UVT of the water. The sensor output and the flow rate are used to monitor the dose delivery. The set-point value for UV intensity over a range of flow rates is determined during the validation. [Pg.352]

Some interpretation may be required to obtain parameters of interest from the values reported in compilations. Gamma-ray intensities may be given in fractions that can be used directly, or relative to a selected—usually the most intense— reference gamma-ray intensity set at a value of 100. In the latter case, all other gamma-ray intensities must be calculated by multiplying the given intensity of the reference gamma ray by the appropriate fraction. [Pg.166]

If fragmentation involves only the urethane group, one would expect sets of peaks for various combinations of n and nt. The smallest of these should be due to cleavage of the PPG units alone from either end of the molecule tiiese would correspond to m = 0. This is observed in the region below 500 Da where three sets of peaks occur, all due to PPG species cationized with Na. The peaks labeled (o) are due to PPG oligomers with n = 4-17 and n ax = 7. The original PPG also had = 7. The most intense set of peaks ( ) is for n = 4 -13 and n = 7 and corresponds to PPG minus two hydrogens. Because the spectra were run under low resolution, no additional structural information is available. The third set ( ) is presumably due to formation of cyclic ethers and has n = 5-12 and n = 7. [Pg.386]

There are four additional sets of regularly occurring patterns in the region from 500-3000 Da, corresponding to different combinations of PPG and MDI blocks. These are summarized in Table 8.12. The most intense set in Figure 8.42, centered at 755 Da ( ) is due to fragments consisting of one MDI and... [Pg.386]

Sodium, calcium, and silicon images of the skull and rib are displayed in a RGB false color overlay image with the sodium and calcium signal intensities set to the same scale for direct visual comparison. An example of the total ion images and the RGB overlay images of skull and rib is illustrated in Fig. 16.2. [Pg.273]

Since, today s computer devices are characterized by high reliability parameters, the intensity of server failure was set to one per three years. In case of hardware failures (assigned only to servers) the component replace time takes into consideration the time needed to delivery of a new system component by courier. The software components failure (with intensity set to one per year) with probability 0.2 results in the component inoperability, in other cases in the component performance degradation. [Pg.2081]


See other pages where Intensity setting is mentioned: [Pg.37]    [Pg.77]    [Pg.78]    [Pg.80]    [Pg.410]    [Pg.43]    [Pg.44]    [Pg.47]    [Pg.216]    [Pg.33]    [Pg.774]    [Pg.6259]    [Pg.245]    [Pg.187]    [Pg.109]    [Pg.6258]    [Pg.537]    [Pg.166]    [Pg.66]    [Pg.290]    [Pg.139]    [Pg.10]    [Pg.216]    [Pg.7]    [Pg.53]    [Pg.333]   
See also in sourсe #XX -- [ Pg.3 ]




SEARCH



Basis sets Raman band intensities

Intensive care units acute, setting

© 2024 chempedia.info