Fourier transform infrared wavenumber measured

Between the source and the detector is put either monochromators used in dispersive instruments or interferometers used in Fourier transform infrared (FT-IR) instruments. In a dispersive instrument the intensity at each wavenumber is measured one by one in sequence and only a small spectral range falls on the detector at any one time. In a FT-IR instrument the intensities of all the wavenumbers are measured simultaneously by the detector. Fourier transform infrared spectroscopy offers some advantages compared to dispersive instruments, namely (i) higher signal-to-noise ratios for spectra obtained under conditions of equal measurement time, and (ii) higher accuracy in frequency for spectra recorded over a wide range of frequencies. Therefore we will give below a brief picture of the principle of FT-IR spectroscopy, based on a Michelson interferometer (Fig. 2). [Pg.205]

Fig. 2(b) shows the results of fourier transform infrared (FTIR) absorption measurements on Be-doped bulk GaAs. This spectra was taken in normal incidence thus illustrating the three-dimensional nature of the confinement and shows two features corresponding to the C and D components of the Is—2p transition at 184 and 167 wavenumbers respectively (corresponding to 54 and 59 mm, note the 2p-ls energy separation is 3/4 of the 28 meV binding energy for Be in bulk). [Pg.536]

Fourier transform infrared (FTIR) spectra were measured using a Dlgilab FTS-14 spectrometer with the Real Time Disk Operating System (RDOS). The samples were scanned 256 times at a resolution of one point every four wavenumbers with double precision. [Pg.277]

The essential problem of the dispersive spectrometer lies with its monochromator. This contains narrow slits at the entrance and exit which limit the wavenumber range of the radiation reaching the detector to one resolution width. Samples for which a very quick measurement is needed, for example, in the eluant from a chromatography column, cannot be studied with instruments of low sensitivity because they cannot scan at speed. However, these limitations may be overcome through the use of a Fourier-transform infrared spectrometer. [Pg.18]

The samples solidified under different cooling rates were analysed by means of an M2000 Fourier-Transform infrared (FTIR) spectrometer manufactured by Midac Co., measuring the absorbance in the range 400-4,000/cm wavenumbers. [Pg.131]

Terahertz (THz) spectroscopy systems utilize far-infrared radiation to extract molecular spectral information in an otherwise inaccessible region of the electromagnetic spectrum where various rotational, vibrational, and translational modes of molecules are located, 0.1-10 THz (Fig. 1). As the wavenumber range is narrowed, THz-radiation can yield more specific information about a particular chemical component within the system. Unlike most spectroscopic techniques, THz instrument measures the wave temporal electric field, which can be Fourier transformed to yield THz pulse amplitude and phase. This added capability allows precise... [Pg.285]

Since FT-IR spectrometry is based on the interference of waves of light (or radiation), first an account of this phenomenon is briefly given, before explaining the Fourier transform method by which an infrared spectrum is obtained from a measured interferogram. Some characteristics of FT-IR spectrometry, namely, wavenumber resolution, measurable wavenumber region, and accurate determination of wavenumbers are discussed. To facilitate the understanding of the description, which inevitably requires some mathematical formulations, many illustrations are provided. [Pg.41]


See other pages where Fourier transform infrared wavenumber measured is mentioned: [Pg.313]    [Pg.344]    [Pg.431]    [Pg.204]    [Pg.202]    [Pg.310]    [Pg.228]    [Pg.79]    [Pg.205]    [Pg.201]    [Pg.603]    [Pg.555]    [Pg.269]    [Pg.337]    [Pg.1006]    [Pg.82]    [Pg.4]    [Pg.127]    [Pg.541]    [Pg.2]    [Pg.551]    [Pg.50]    [Pg.760]    [Pg.994]    [Pg.55]    [Pg.273]    [Pg.344]    [Pg.27]    [Pg.32]    [Pg.32]   