External standardization multiple-point

A multiple-point standardization presents a more difficult problem. Consider the data in Table 5.1 for a multiple-point external standardization. What is the best estimate of the relationship between Smeas and Cs It is tempting to treat this data as five separate single-point standardizations, determining k for each standard and reporting the mean value. Despite its simplicity, this is not an appropriate way to treat a multiple-point standardization. [Pg.117]

Data for Hypothetical Multiple-Point External Standardization... [Pg.117]

Case 3. Linearity demonstrated from 50% of the ICH reporting limit to 150% of the shelf life specification of a related substance, and a significant y-intercept is observed (Figure 3.8). Due to the significant -intercept, a single-point calibration (e.g., high-low or one-point external standard calibration) is not suitable. In this case, multiple-point external standard calibration is the most appropriate. See Section 3.3.3 for more discussion of the significant y-intercept. [Pg.33]

Prerequisites for the Calibration Types. It depends on the design of the analytical procedure as to which regression parameters are meaningful and which results are acceptable. In other words, the model to be used for quantitation must be justified. For a singlepoint calibration (external standardization), a linear function, zero intercept, and the homogeneity of variances are required. The prerequisites for a linear multiple-point calibration are a linear function and in case of an unweighted calibration also the homogeneity of variances. A non-linear calibration requires only a continuous function. With respect to the 100%... [Pg.100]

The multiple-quantum (MQ)/MAS NMR is one of the 2D NMR methods, which is capable of averaging out the second-order quadrupolar interaction in nuclei with spin > 1/2 such as H, "B, O, etc. The "B MQ/ MAS NMR measurements on boron as contained in silyl-carborane hybrid Si-based polymer networks considered here. The molded samples are cut into small pieces to insert them into a 4-mm NMR rotor and spun at 12 kHz in a MAS probe. The observation frequency of the "B nucleus (spin number I = 3/2 and isotope natural abundance = 80.42%) is 96.3 MHz. Excitation of both the echo (—3Q) and anti echo (+3Q) coherences is achieved by using a three-pulse sequence with a zero quantum filter (z-filter). The widths of the first, second, and third pulses are 3.0 4.1 ps, 1.1-1.6 ps, and 19-28 ps, respectively. The z-filter is 20 ps. The recycle delay time is 6-15 s and the data point of FI (vertical) axis is 64 and for each the number of scans is 144. Then, the total measurement time is 15-38 h. The phase cycling used in this experiment consists of 12 phases. Boron phosphate (BPO4 3 = 0 ppm) is used as an external standard for "B. The chemical shift value of BPO4 is —3.60 ppm from BF3 O(C2H5)2 which is used as a standard reference in " B NMR in the liquid state. The transmitter frequency of " B is set on peak of BPO4 for a trustworthy chemical shift after Fourier transform." " ... [Pg.208]

The external standard method is the simplest one and should therefore only be used for simple analytical problems. The injection must be performed with good reproducibility thus it is recommended that the complete loop filling method be used (see Section 4.6). With multiple-point calibration it is not recommended to inject different volumes from a reference stock solution (e.g. 10, 20, 30, 40 and 50 (xl) because it is well possible that neither accuracy nor precision of these injections are high enough. It is better to prepare a number of calibration solutions with different concentrations and to inject equal volumes of them (with complete loop filling or with exactly the same procedure as is used also for the sample). [Pg.294]


See other pages where External standardization multiple-point is mentioned: [Pg.109]    [Pg.1188]    [Pg.390]    [Pg.33]    [Pg.764]    [Pg.441]    [Pg.413]    [Pg.285]    [Pg.273]    [Pg.245]    [Pg.2239]    [Pg.432]    [Pg.810]    [Pg.1366]    [Pg.188]    [Pg.111]   
See also in sourсe #XX -- [ Pg.117 , Pg.117 ]