Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Current to voltage amplifier

Fig. 3. Current to voltage amplifier. The output voltage is given by — Tr-Rfb-... Fig. 3. Current to voltage amplifier. The output voltage is given by — Tr-Rfb-...
It was useful in some cases to use both systems (CVC and DA) in parallel. The resistor was short-circuited in one direction to allow high currents, recorded by the current-to-voltage amplifier, and very small currents in the other direction, recorded by the differential amplifier. Moreover, the differential amplifier could be modified into an auto-ranging system to record current transients after large potential steps [15]. [Pg.1700]

The current-to-voltage amplifier is replaced in some concepts by other systems such as simple resistors between working electrode and ground potential [16, 17], negative resistances (gyrator circuits, [18,19]), logarithmic current amplifiers, and current-photon-converters [20], often in context with IR-drop elimination ( iR-Drop Elimination). [Pg.1700]

The electrochemical circuitry required for SECM is relatively straightforward. Since the interface is not generally externally polarized in SECM measurements of liquid-liquid interfaces, a simple two-electrode system suffices (Fig. 3). A potential is applied to the tip, with respect to a suitable reference electrode, to drive the process of interest at the tip and the corresponding current that flows is typically amplified by a current-to-voltage converter. [Pg.295]

Only about 1 in 10s carbon atoms produces an ion, but ion production is proportional to the number of susceptible carbon atoms entering the flame. In the absence of analyte, 10 14 A flows between the flame tip and the collector, which is held at +200 to 300 V with respect to the flame tip. Eluted analytes produce a current of 10 12 A, which is converted to voltage, amplified, filtered to remove high-frequency noise, and finally converted to a digital signal. [Pg.543]

Figure 6.8 Two-amplifier adder-type potentiostat with current-to-voltage converter. Figure 6.8 Two-amplifier adder-type potentiostat with current-to-voltage converter.
To carry out amperometric or voltammetric experiments simultaneously at different electrodes in the same solution is not difficult. In principle, any number of working electrodes could be studied however, it is unlikely that more than two or three would ever be widely used in practice. The bulk of the solution can have only one controlled potential at a time (if there are significant iR drops, there will be severe control problems with multiple-electrode devices). It is necessary to use a single reference electrode to monitor the difference between this inner solution potential and the inner potential of W1 at the summing point of an operational amplifier current-to-voltage converter (this is the potential of the circuit common see OA-2 in Fig. 6.17). The potential difference between... [Pg.185]

Figure 7.1 (A) Typical controlled-potential circuit and cell OA1, the control amplifier OA2, the voltage follower (Vr = Er) OA3, the current-to-voltage converter. (B) Equivalent circuit of cell Rc, solution resistance between auxiliary and working electrodes Ru, solution resistance between reference and working electrodes, Rs = Rc + Ru and Cdl, capacitance of interface between solution and working electrode. (C) Equivalent circuit with the addition of faradaic impedance Zf due to charge transfer. Potentials are relative to circuit common, and working electrode is effectively held at circuit common (Ew = 0) by OA3. Figure 7.1 (A) Typical controlled-potential circuit and cell OA1, the control amplifier OA2, the voltage follower (Vr = Er) OA3, the current-to-voltage converter. (B) Equivalent circuit of cell Rc, solution resistance between auxiliary and working electrodes Ru, solution resistance between reference and working electrodes, Rs = Rc + Ru and Cdl, capacitance of interface between solution and working electrode. (C) Equivalent circuit with the addition of faradaic impedance Zf due to charge transfer. Potentials are relative to circuit common, and working electrode is effectively held at circuit common (Ew = 0) by OA3.
A potentiostat as a laboratory device is normally equipped with further features such as monitoring of the -> reference electrode, current detection by current-to-voltage-converters, or differential amplifiers, pulse form generators, displays of current and potential, and computer interfaces. See also - IRU potential drop and - IR drop compensation, - Randles. [Pg.545]

Figure 3.5. Engineering drawing of a solid-state photodetector with integrated amplifier, current-to-voltage converter, and temperature controller, (courtesy PerkinElmer Optoelectronics)... Figure 3.5. Engineering drawing of a solid-state photodetector with integrated amplifier, current-to-voltage converter, and temperature controller, (courtesy PerkinElmer Optoelectronics)...
ACIS measurements were performed at frequencies between 1 mHz and 1 kHz using a Solartron Model 1250 Frequency Response Analyzer. Output from the comb specimens was amplified with a Keithley Model 427 Current-to-Voltage Converter before waveform analysis. Reference electrodes were not used owing to the geometry of the encapsulated test specimens. The data reported herein were obtained with a 0.1 V rms amplitude sinusoidal excitation waveform. In one experiment, DC bias was superimposed on this waveform. [Pg.319]

Schematic diagrams of the instrumentation for the ANL capacitive flowmeter are given in Fig. 6.20. A 100-kHz sine-wave oscillator, with stable frequency and amplitude controls, was used to pulse the drive electrode. Each sensing electrode was connected to a current-to-voltage converter preamplifier. The preamplifier outputs were bandpass filtered at 100 kHz 5 Hz and amplitude-demodulated. The demodulated signals were amplified and DC-coupled to a first-order low-pass filter to give density signals. Schematic diagrams of the instrumentation for the ANL capacitive flowmeter are given in Fig. 6.20. A 100-kHz sine-wave oscillator, with stable frequency and amplitude controls, was used to pulse the drive electrode. Each sensing electrode was connected to a current-to-voltage converter preamplifier. The preamplifier outputs were bandpass filtered at 100 kHz 5 Hz and amplitude-demodulated. The demodulated signals were amplified and DC-coupled to a first-order low-pass filter to give density signals.
If fast signals with medium strength need to be recorded, then photodiodes are used in connection with so-called transimpedance amplifiers. In this case, the photodiode is driven in the photocurrent mode, i.e. it is driven in short-cut mode with Ud being nearly equal to zero. This mode requires that any preamplifier to be used is in transimpedance mode, i.e. its input impedance can be matched to that of the photodiode. This mode, also termed current-to-voltage converter, allows one to achieve low input impedance even for high gain. [Pg.195]


See other pages where Current to voltage amplifier is mentioned: [Pg.338]    [Pg.307]    [Pg.549]    [Pg.546]    [Pg.292]    [Pg.6523]    [Pg.257]    [Pg.175]    [Pg.365]    [Pg.290]    [Pg.410]    [Pg.141]    [Pg.1138]    [Pg.1281]    [Pg.338]    [Pg.307]    [Pg.549]    [Pg.546]    [Pg.292]    [Pg.6523]    [Pg.257]    [Pg.175]    [Pg.365]    [Pg.290]    [Pg.410]    [Pg.141]    [Pg.1138]    [Pg.1281]    [Pg.57]    [Pg.669]    [Pg.11]    [Pg.194]    [Pg.339]    [Pg.175]    [Pg.176]    [Pg.177]    [Pg.178]    [Pg.57]    [Pg.57]    [Pg.121]    [Pg.31]    [Pg.2972]    [Pg.443]    [Pg.742]    [Pg.162]    [Pg.59]    [Pg.4782]    [Pg.346]    [Pg.208]    [Pg.229]   
See also in sourсe #XX -- [ Pg.73 ]




SEARCH



Amplifiers

Current-voltage

© 2024 chempedia.info