Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Columns change

Occasionally, samples are run that adsorb onto the packing material. Generally, if one of the performance characteristics of the column changes by 10% or more, it is prudent to clean the column. These performance characteristics are (1) asymmetry factor, retention time, resolution, and theoretical plates. [Pg.134]

Interchanging two adjacent columns changes the sign of the result. [Pg.26]

If you have only two components, start with a nonpolar solvent, and when you are sure the first component is completely off the column, change to a really polar one. With only two components, it doesn t matter what polarity solvent you use to get the second compound off the column. [Pg.215]

Table 5.2. Influence matrix eachfigure represents by how much the variables in thefirst column change when the observation on the top row is left out. Table 5.2. Influence matrix eachfigure represents by how much the variables in thefirst column change when the observation on the top row is left out.
There are several ways to account for variable pressures. If the total pressure of the column changes but not the pressure drop through the trays (the normal situation in heat-integrated columns, particularly with valve trays whose pressure drops are fairly constant), an approximate variable-pressure model can be used. [Pg.141]

We want to use a feedforward controller to make adjustments in steam flow to the reboiler, whenever the feed rate to the column changes, so that bottoms composition is held constant. The feedforward-controller design equation [Eq. (11.28)] gives... [Pg.385]

The experimental order of this study was also changed from the ideal randomised and blocked design. This was justified for reasons of automation where column changes needed to be minimised. As for the aspirin study the validity of this compromise depends on the fact that the repeatability of the method, over a time span such as that required for the ruggedness test, had previously been established. [Pg.228]

The most dramatic main effects observed were due to column changes, the Hypersil column causing a 53% increase in the resolution, as compared to the Spherisorb column. These effects are unexpectedly high due to the close correlation of the specifications for these columns. [Pg.228]

The function of the detector is to sense and respond with an electrical signal when the composition of the gas emerging from the column changes. The type of detector used is dependent on the application. The most widely used detectors are the thermal conductivity, flame ionization, and electron capture. [Pg.335]

The best way to follow column changes is by way of column standard plate counts. For discussion purposes, we will use the four-standard mixture of acetophenone, nitrobenzene, benzene, and toluene described in the discussion on efficiency factor (Chapter 4). Our column will be a Ci8 reverse-phase column run in 70% acetonitrile/water at 254 nm. In an initial run, we obtain four peaks whose interpeak a s double between each pair. After we discuss reverse phase, we will see how these killers affect normal phase columns. [Pg.74]

Column changes are a viable methods development tool in HPLC. Column can be switched at any time. Attach the new column with a slow flow of solvent from the injector outlet line to All the inlet fitting of your column so you don t leave air in the column entrance. Figure on flushing a column with at least six column volumes of a new solvent to re-equilibrate the column before injecting a sample. Plan on ignoring the first injection. Usually, the second, third, and subsequent injections will be reproducible. [Pg.209]

Affinity chromatography techniques have shown less utility in analytical testing than in preparative separations for a variety of reasons, including cost and the difficulty of validating consistent operation as the column changes over time. Protein A affinity has been commonly used to quantitate the total antibody content of either ascites or cell culture fluids. To provide guidance in the development of a purification process, specific immunoaffinity resins are either available or can be readily prepared to quantitate the levels of unrelated protein contaminants. To rapidly determine what the active species in a mixture is, a monoclonal antibody that... [Pg.91]

Column changed, new column has different selectivity from that of old column. [Pg.127]

Indeterminate errors are errors that cannot be eliminated and are inherent in the analytical technique. Determinate errors are errors whose cause and magnitude can be determined, and they include poor sampling technique, decomposition of the column, change in detector response, improper recorder performance, calculation errors, and operator prejudice or error. [Pg.240]

Suppose the barometer in Figure 1 was carried into an open mine 500 meters below sea level. How would the height of the mercury column change Explain why. [Pg.14]

After the component has eluted from the column, change attenuation of the spectrophotometer to 0.1 AUFS (254 nm). [Pg.418]

After the components have eluted from the column, change the refractometer to 32X or the next less-sensitive setting and inject 200 /uL of solution from Activity B-l. [Pg.419]


See other pages where Columns change is mentioned: [Pg.184]    [Pg.383]    [Pg.1334]    [Pg.225]    [Pg.237]    [Pg.449]    [Pg.200]    [Pg.215]    [Pg.168]    [Pg.117]    [Pg.380]    [Pg.184]    [Pg.383]    [Pg.169]    [Pg.49]    [Pg.119]    [Pg.323]    [Pg.43]    [Pg.151]    [Pg.13]    [Pg.289]    [Pg.293]    [Pg.180]    [Pg.475]    [Pg.519]    [Pg.138]    [Pg.12]    [Pg.75]    [Pg.419]   
See also in sourсe #XX -- [ Pg.200 ]




SEARCH



Column change time

Columns length change

Mobile phase column changes

© 2024 chempedia.info