Cobas Ready

At present the Cobas Ready consumes S to 6 xl of serum or plasma per test, depending on the chemistry performed. This volume is automatically pipetted by the built-in autosampler, which avoids the need of manual pipetting and therefore of eventual operator-to-operator pipetting variations. [Pg.22]

Multi-test reagent strips of the Cobas Ready system... [Pg.22]

Fig. 7. Cobas Ready system, open front door. Fig. 7. Cobas Ready system, open front door.
Fig. 8. Cobas Ready, application of a sample on the test strip. Fig. 8. Cobas Ready, application of a sample on the test strip.
A simplified presentation of the reflectometer of the Cobas Ready is shown in Fig. 9. [Pg.25]

Reflectance values on Cobas Ready as on any diffuse reflector are measured as the quantities of light reflected from a reagent layer saturated with a serum sample (/,) or from a white or black reference plate (/ and 4), each normalized to the incident light intensity monitored by the reference photodetector... [Pg.26]

All the reflectivity measurements on the Cobas Ready are bichromatic with... [Pg.26]

Pretreatment of the sample Add 100 nl serum or plasma to 100 al precipitate solution. After good mixing (10 times) allow the mixture to stand for 5 min. at room temperature. Then centrifuge the sample for 9 min. in the Cobas Ready centrifuge. Use the clear supernatant for analysis. [Pg.38]

Lapin, A., Halbritter, E., Ruthner, U., Asboth, F. and Feigl, W. (1990). Erste Erfahrungen mit dem Trockenchemiesystem Cobas Ready in einem Kranken-hauslaboratorium. Ber. GGKC 14, 55. [Pg.50]

Van Dieijen-Visser, M.P. and Janson, P.C.W. (1992). Evaluatie van een nieuwe drogestof-chemie analyzer Cobas Ready/Spotchem. Tijdschr. NVKC 17, 73-79. [Pg.50]

Analyzer Ektachem 700 Ektachem DT-60 Reflotron Cobas Ready, Spotchem Seralyzer Stratus Opus... [Pg.602]

Bilirubin interferes if measurement is effected at an incident angle (i.e. incident on the side of application of the sample) (Cobas Ready, Reflotron, Seralyzer), since it can contribute to influencing the dye that has formed, without having participated in the reaction (e.g., fluorescence). If measurements are... [Pg.609]

The dwelling time covers the period that elapses between the application of the sample to the test slide and its insertion into the measurement device. Since the sample is inserted by the analyser in the Ektachem, Cobas Ready and Stratus instruments, this factor can be neglected as far as these systems are concerned. It is only with the Ektachem DT-60, Reflotron and Seralyzer that dosage of the sample is effected by means of an air displacement pipette via the user. No time error can occur with the Ektachem DT-60 because an inter-... [Pg.610]

In some systems (Cobas Ready, Ektachem, Seralyzer or Stratus) the user... [Pg.611]

A detailed description of various commercially available dry chemistry systems such as Ektachem, Reflotron, Seraly2er, Cobas Ready, Drichem, Opus and Stratus are also included. [Pg.632]


See other pages where Cobas Ready is mentioned: [Pg.21]    [Pg.21]    [Pg.21]    [Pg.23]    [Pg.23]    [Pg.25]    [Pg.25]    [Pg.27]    [Pg.27]    [Pg.29]    [Pg.31]    [Pg.33]    [Pg.35]    [Pg.37]    [Pg.39]    [Pg.41]    [Pg.43]    [Pg.45]    [Pg.47]    [Pg.49]   