Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Calibration, responsibility

There is no widely used detector that has predictable absolute response factors. This is why predicted rather than calibrated response factors are only used when calibration is not practical, such as for unknowns, or when a closely related compound has been calibrated, allowing extrapolation to the desired response factor. [Pg.227]

Calculate the final concentrations of carotenoids in samples by multiplying the peak areas of analytes by the calibration response factors. Apply sample weight and dilution factors to arrive at the concentration of carotenoids in the original sample (i.e., initial concentration). [Pg.860]

Both organic and inorganic polymer materials have been used as solid supports of indicator dyes in the development of optical sensors for (bio)chemical species. It is known that the choice of solid support and immobilization procedure have significant effects on the performance of the optical sensors (optodes) in terms of selectivity, sensitivity, dynamic range, calibration, response time and (photo)stability. Immobilization of dyes is, therefore, an essential step in the fabrication of many optical chemical sensors and biosensors. Typically, the indicator molecules have been immobilized in polymer matrices (films or beads) via adsorption, entrapment, ion exchange or covalent binding procedures. [Pg.191]

Fig. 1 Calibration graphs of ion-selective electrodes and evaluation of selectivity coefficients. 1) Calibration response against —log in solution of free determinand i. The practical limit of detection may be taken as the activity (or the concentration) at the point of intersection of the extrapolated lines as shown. 2) Calibration against —log a in the presence of the interferant), the activity of which is of a known and constant value. Response ) is obtained for —log a = —log The resulting interference can restrict the measuring range, b) The abscissa approximately equal to 60/z mV. Fig. 1 Calibration graphs of ion-selective electrodes and evaluation of selectivity coefficients. 1) Calibration response against —log in solution of free determinand i. The practical limit of detection may be taken as the activity (or the concentration) at the point of intersection of the extrapolated lines as shown. 2) Calibration against —log a in the presence of the interferant), the activity of which is of a known and constant value. Response ) is obtained for —log a = —log The resulting interference can restrict the measuring range, b) The abscissa approximately equal to 60/z mV.
As previously indicated, the determination of a drug concentrations in plasma specimens requires the construction of a calibration response curve. This curve is often constructed as a straight line from the measured peak response ratios (y,) plotted against their respective calibrator concentrations (x,). The drug concentration in a specimen or the apparent (back-calculated) calibrator concentration is obtained from a rearrangement of theequation for the calibration line (without error) x,- = (y,- — b)/m. An example of a calibration curve with back-calculated concentrations is given the first four columns of Table 10. [Pg.3497]

Regardless of the method used to calibrate response function, it is useful to have standard corrected spectra of readily available materials to verify that the... [Pg.284]

The drawbacks of the original Orion liquid membrane electrode model were successfully overcome by simply incorporating the neat exchanger within a thin matrix of PVC (9). Besides having more favourable mechanical and physical characteristics, these novel PVC electrodes had prolonged functional lifetimes and a performance, in terms of Nernstian calibration, response times and selectivity coefficients, on a par with the original Orion 92-20 model. In fact, the mean content of calcium determined with the respective electrodes in tap water on 10 consecutive days was 31.55 and 31.57 ppm (9). [Pg.78]

The calibration response for benzene with a MCT detector may be nonlinear. In the round robin of this test method a linear fit was used for concentrations up to approximately one mass % benzene and a point to point or quadratic fit used for higher concentrations. The region of linearity may vary among instrument types and needs to be determined during calibration. [Pg.1016]

Figure 5 shows the calibration response of a 0.1 ym thick GDPP coated electrode to varying concentrations af dissolved oxygen in 0.9% saline solution, measured over 72 days. The electrodes were kept soaked in the saline solution during this time. Similar trials... [Pg.90]

Figure 5. Calibration responses of a 0.1 ym thick GDPP coated... Figure 5. Calibration responses of a 0.1 ym thick GDPP coated...

See other pages where Calibration, responsibility is mentioned: [Pg.102]    [Pg.599]    [Pg.590]    [Pg.86]    [Pg.242]    [Pg.67]    [Pg.675]    [Pg.102]    [Pg.167]    [Pg.519]    [Pg.91]    [Pg.134]    [Pg.567]    [Pg.29]    [Pg.567]    [Pg.437]    [Pg.738]    [Pg.206]   
See also in sourсe #XX -- [ Pg.96 ]




SEARCH



Absolute Response Calibration

Calibration curve and linear response slope of pH microelectrodes

Calibration detector response

Calibration instrument response function

Calibration response function

© 2024 chempedia.info