Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Calibration, light meters

Figure 3.19. Absolute determination of 8 by in situ autocorrelation. Experiments were performed with a mode locked femtosecond Ti sapphire laser. A prism pair (PC) was used to compensate the group delay dispersion (GDD) of the microscope objective. A long-pass filter eliminates residual argon pump light and Ti sapphire fluorescence. After two sequential beam expanders (BE), the beam was approximately 25 mm in diameter, which was sufficient to overfill the back aperture (10-mm diameter) of the objective. A long-pass dichroic mirror (DC) with reflectivity separates fluorescence from excitation light. The incident power at the sample was measured by recollimating the transmitted beam onto a calibrated power meter. Fluorescence was detected by a photomultiplier tube and recorded as a function of the interferometer delay. (From Ref. [366] with permission of the Optical Society of America.)... Figure 3.19. Absolute determination of 8 by in situ autocorrelation. Experiments were performed with a mode locked femtosecond Ti sapphire laser. A prism pair (PC) was used to compensate the group delay dispersion (GDD) of the microscope objective. A long-pass filter eliminates residual argon pump light and Ti sapphire fluorescence. After two sequential beam expanders (BE), the beam was approximately 25 mm in diameter, which was sufficient to overfill the back aperture (10-mm diameter) of the objective. A long-pass dichroic mirror (DC) with reflectivity separates fluorescence from excitation light. The incident power at the sample was measured by recollimating the transmitted beam onto a calibrated power meter. Fluorescence was detected by a photomultiplier tube and recorded as a function of the interferometer delay. (From Ref. [366] with permission of the Optical Society of America.)...
The light source was a 1 KW Oriel Solar Simulator equipped with AMO filters. Light intensities were measured by a Karl Lambrecht calibrated light probe. The unattenuated power flux delivered by the System was 208 mW/cm. Incident light intensity was varied using neutral density filters. The light source for action spectra measurements consisted of a 150 W Xe lamp and a 14 meter Jarrel Ash monochromator. The action spectra of the cell was measured as described previously (7). [Pg.444]

Place the magnetic stirring bar in the Erlenmeyer flask, and stir it moderately. Monitor the pH of the solution with a properly calibrated pH meter, and slowly add. 5M (42 grams per liter) sodium bicarbonate solution until the pH has risen to the range of 8-8.5. Higher pH will cause racemization. The freebase is then extracted from the water solution with chloroform. Two extractions with 100 ml of chloroform should complete the extraction, but check a third extraction with the black light to ensure that most all of the product iso-lysergic acid hydrazide has been extracted. [Pg.37]

Noises and light in the bedroom can disturb sleep and they may be easily detected using the microphone and light meter or cameras of smartphones. Snore sound is one of the signals usually acquired in a sleep laboratory when a sleep-related breathing disorder is suspected. There is evidence that when a microphone is calibrated and placed at a specified... [Pg.180]

An interferometer can be used to very accurately measure the thermal expansion of solids. Although not utilized commercially to the level of dilatometry, NIST standard materials, which are in turn used to calibrate dilatometers, have had their expansion characteristics determined using interferometry. In fact, the formal definition of the meter is based on interferometric measurements. The operation of the device is based on the principle of interference of monochromatic light. The fundamental relations between wavelength and distance will first... [Pg.186]

Color Meter. The Chroma Meter CR-300 (Minolta, Osaka, Japan) provides both accurate and precise color data (Post et al., 1993). This color meter uses a pulsed xenon light for stable and uniform illumination of the sample. Three photocells measure the photons diifusely reflected by the sample through filters matching the CIE standard observer spectral response. A standard white plate is used for calibration. For analysis, the hand-held detector unit which is protected by a glass window may be placed on top of a small pile of a powdered sample (300 mg), thus standardizing the sample surface. This way, up to one hundred samples can be precisely measured within one hour. Further advantages are the mobility of the unit, and the output of the three color parameters in several color systems (see below). [Pg.32]

As measuring techniques became more precise and the demand for accuracy increased, the standards on which people based their units were improved. In the 18 century, the French invented the metric system, based on a more consistent, systematic, and carefully defined set of standards than had ever been used before. For example, the meter (or metre, from the Greek metron, a measure ) became the standard for length. The first definition for the standard meter was one ten-millionth of the distance from the North Pole to the Equator. This became outdated as the precision of scientist s measuring instruments improved. Today, a meter is defined as the distance light travels in a vacuum in 1/299,792,458 second. Technical instruments for measuring length are calibrated in accordance with this very accurate definition. [Pg.10]

A lux meter, calibrated at intervals prescribed by the manufacturers, is recommended. Because lux is a measurement of light as perceived by the human eye (Figure 6.11) rather than an absolute measurement, a lux meter is not suitable for comparing the output of different types of source. However, it is well suited for checking the evenness of illumination of the sample area and measuring changes in source output with time. [Pg.152]

Samples may be exposed side-by-side with a validated chemical actinometric system to ensure the specified light exposure is obtained, or for the appropriate duration of time when conditions have been monitored using calibrated radiometers/lux meters. An example of an actinometric procedure is provided in the Annex. [Pg.218]

Exposure levels of a product should be confirmed by monitoring that uses calibrated radiometers or lux meters. Validated chemical actinometry can also be used for this purpose. For near-UV light, quinine actinometry is noted in the appendix of the guideline. Quinine actinometry was adopted because its usefulness was confirmed by a collaborative study of the ICH working group892 and by a study carried out by the U.S. Food and Drug Administration.893 However, concern has been expressed about the usefulness of quinine acti-... [Pg.224]

The most common method of measuring the spinning rate is by some optical means, for example, by a photo-sensitive transistor and a light source. The spinning can be monitored on an oscilloscope in which case the quality of spinning can be kept track of as well as the speed. Alternatively, a tachometer can be constructed in a fashion analogous to counting rate meters used in nuclear physics (Elmore and Sands, 1949) and appropriately calibrated. [Pg.489]


See other pages where Calibration, light meters is mentioned: [Pg.362]    [Pg.362]    [Pg.360]    [Pg.361]    [Pg.632]    [Pg.186]    [Pg.460]    [Pg.91]    [Pg.117]    [Pg.282]    [Pg.145]    [Pg.121]    [Pg.210]    [Pg.7]    [Pg.240]    [Pg.573]    [Pg.202]    [Pg.117]    [Pg.131]    [Pg.131]    [Pg.294]    [Pg.329]    [Pg.199]    [Pg.32]    [Pg.349]    [Pg.174]    [Pg.126]    [Pg.2863]    [Pg.241]    [Pg.472]    [Pg.275]    [Pg.85]    [Pg.64]    [Pg.873]    [Pg.90]    [Pg.315]    [Pg.116]    [Pg.66]    [Pg.331]   


SEARCH



Lighting meters

© 2024 chempedia.info