Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Analytical signal detection

Spike recoveries for samples are used to detect systematic errors due to the sample matrix or the stability of the sample after its collection. Ideally, samples should be spiked in the field at a concentration between 1 and 10 times the expected concentration of the analyte or 5 to 50 times the method s detection limit, whichever is larger. If the recovery for a field spike is unacceptable, then a sample is spiked in the laboratory and analyzed immediately. If the recovery for the laboratory spike is acceptable, then the poor recovery for the field spike may be due to the sample s deterioration during storage. When the recovery for the laboratory spike also is unacceptable, the most probable cause is a matrix-dependent relationship between the analytical signal and the concentration of the analyte. In this case the samples should be analyzed by the method of standard additions. Typical limits for acceptable spike recoveries for the analysis of waters and wastewaters are shown in Table 15.1. ... [Pg.711]

A sample is to be analyzed following the protocol shown in Figure 15.2, using a method with a detection limit of 0.05 ppm. The relationship between the analytical signal and the concentration of the analyte, as determined from a calibration curve is... [Pg.723]

The immobilization of the enzymes on various solid carriers gives perfect possibilities to develop not only sensitive and selective, but also rapid and simple test procedures for different compounds determination with visual detection of the analytical signal. [Pg.167]

Signal-to-noise ratio The ratio of the intensity of the analytical signal to that of the noise. This is used in determining the limits of detection and quantitation. [Pg.311]

Sensitivity by itself is not sufficient to completely evaluate an LCEC system for analytical purposes. The minimum detectable quantity (detection limit) is of more practical importance. The detection limit takes into consideration the amount of baseline noise as well as the response to the analyte. The detection limit is then defined as the quantity of analyte which gives a signal-to-noise ratio of three (a S/N of 3 is the generally accepted criterion although other values have been used). For a complete description of an LCEC application, both the sensitivity and detection limit, along with the S/N criteria used, should be provided. [Pg.24]

Mass spectrometry is the only universal multielement method which allows the determination of all elements and their isotopes in both solids and liquids. Detection limits for virtually all elements are low. Mass spectrometry can be more easily applied than other spectroscopic techniques as an absolute method, because the analyte atoms produce the analytical signal themselves, and their amount is not deduced from emitted or absorbed radiation the spectra are simple compared to the line-rich spectra often found in optical emission spectrometry. The resolving power of conventional mass spectrometers is sufficient to separate all isotope signals, although expensive instruments and skill are required to eliminate interferences from molecules and polyatomic cluster ions. [Pg.648]

Several authors [386,387] have discussed the spectroscopic and nonspectroscopic (matrix) interferences in ICP-MS. ICP-MS is more susceptible to nonspectroscopic matrix interferences than ICP-AES [388-390]. Matrix interferences are perceptible by suppression and (sometimes) enhancement of the analyte signal. This enhanced susceptibility has to be related to the use of the mass spectrometer as a detection system. In fact, since both techniques use the same (or comparable) sample introduction systems (nebulisers, spray chambers, etc.) and argon plasmas (torches, generators, etc.), it is reasonable to assume that, as far as these parts are concerned, interferences are comparable. The most severe limitation of ICP-MS consists of polyatomic... [Pg.655]

The generation of analytical signals is a complex process that takes place in several steps. Methods of instrumental analysis often need five steps, namely (1) the genesis, (2) the appearance, (3) the detection and conversion, (4) the registration, and (5) the presentation of signals see Fig. 3.3. [Pg.72]

In principle, there is no fundamental contrast between qualitative, (semi-quantitative), and quantitative analyses. The analytical signal is generated in the same way, only the detection and evaluation is done on the basis of a more rough scale, in qualitative analysis only in form of a yes/no decision. [Pg.111]

There is no clear, specific analyte signal the instrument response (T) does not change adequately with a variation in the analyte value (X). This phenomenon indicates that small changes in analyte concentration are not detected by the measurement instrument. Different or additional instrument response information is required to describe the analyte (the problem is underdetermined). [Pg.381]

The design and implementation of a portable fiber-optic cholinesterase biosensor for the detection and determination of pesticides carbaryl and dichlorvos was presented by Andreou81. The sensing bioactive material was a three-layer sandwich. The enzyme cholinesterase was immobilized on the outer layer, consisting of hydrophilic modified polyvinylidenefluoride membrane. The membrane was in contact with an intermediate sol-gel layer that incorporated bromocresol purple, deposited on an inner disk. The sensor operated in a static mode at room temperature and the rate of the inhibited reaction served as an analytical signal. This method was successfully applied to the direct analysis of natural water samples (detection and determination of these pesticides), without sample pretreatment, and since the biosensor setup is fully portable (in a small case), it is suitable for in-field use. [Pg.371]


See other pages where Analytical signal detection is mentioned: [Pg.224]    [Pg.725]    [Pg.1625]    [Pg.245]    [Pg.160]    [Pg.174]    [Pg.444]    [Pg.162]    [Pg.550]    [Pg.618]    [Pg.143]    [Pg.66]    [Pg.176]    [Pg.256]    [Pg.544]    [Pg.572]    [Pg.63]    [Pg.68]    [Pg.267]    [Pg.301]    [Pg.140]    [Pg.510]    [Pg.665]    [Pg.69]    [Pg.50]    [Pg.237]    [Pg.249]    [Pg.18]    [Pg.133]    [Pg.139]    [Pg.142]    [Pg.146]    [Pg.260]    [Pg.370]    [Pg.195]    [Pg.231]    [Pg.324]    [Pg.343]    [Pg.474]   
See also in sourсe #XX -- [ Pg.47 ]

See also in sourсe #XX -- [ Pg.47 ]




SEARCH



Analyte signal

Analytic signal

Analytical detectability

Analytical signals

Signal detection

Signal detection, analytical model

© 2024 chempedia.info