Linear Wear Rate (mm/year) Based on Radiographs for the Modular Harris-Gallante I Acetabular Component as a Function of Implantation Time... [Pg.85]

Range of Clinical Wear Performance in Modular Acetabular Components... [Pg.85]

Qualitatively, the trends observed for cemented components also appear to be applicable to modular acetabular components, evaluated using current computer-assisted wear measurement techniques (Tables 5.2 and 5.3, and Figure 5.11). The box and whisker plots shown in Figure 5.11 compare the linear... [Pg.85]

Distribution of LWRs and VWRs estimated in a first-generation modular acetabular component design (Harris-Gallante Zimmer Warsaw, IN). These graphs show the tendency for the wear rates to decrease with implantation time. Results are compared based on analysis of A-P radiographs alone (two-dimensional), as well as based on the combined analysis of A-P and lateral radiographs (three-dimensional). (Based on data provided courtesy of J. Martell, M.D.)... [Pg.86]

Barrack R.L. 1996. Concerns with cementless modular acetabular components. Orthopedics 19 741-743. [Pg.87]

Sulzer s second-generation MOM designs incorporated a CoCr articulating surface, but the acetabular component consisted of a modular shell and an UHMWPE liner embedded with a CoCr insert. A cross-section of the contemporary "sandwich" type design, distributed under the trade name METASUL (Centerpulse Orthopedics, Winterthur, Switzerland) is illustrated in Figure 6.5. [Pg.98]

Kurtz, S. M., Ochoa, J. A., White, C. V., Siivastav, S., and Coumoyer, J. (1998), Backside nonconformity and locking restraints affect liner/shell load transfer mechanisms and relative motion in modular acetabular components for total hip replacement, J. Biomech. 31 431-437. [Pg.230]

RANGE OF CLINICAL WEAR PERFORMANCE IN MODULAR ACETABULAR COMPONENTS... [Pg.51]

Kircher J, Bader R, Schroeder B, Mittelmeier W. Extremely high fracture rate of a modular acetabular component with a sandwich polyethylene ceramic insertion for THA a preliminary report Arch Orthop Trauma Surg 2008. [Pg.78]

The head of the femoral component then articulates with an ion-bombarded, HDPE, high walled, acetabular liner which fits iato a screwed ia, machined, titanium, chromium—cobalt—molybdenum or vanadium—aluminum metallic alloy hydroxyapatite-coated acetabular shell/cup. Each of the separate parts of the modular system for total hip arthroplasty is manufactured ia several different sizes. [Pg.189]