Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Zero excitation

The second general approach strives to produce no net excitation of the solvent resonance. In other words, the sequence ultimately returns the solvent magnetisation to the 4-z axis whilst at the same time placing all other magnetisation in the transverse plane prior to acquisition. Section 9.4.3 describes a related approach used in conjunction with pulsed field gradients. [Pg.362]

The binomial sequences aim to improve the zero excitation profile and provide schemes that are less sensitive to spectrometer imperfections. The series may be written 1-1, 1-2-1, 1-3-3-1. .. and so on, where the numbers indieate the relative pulse widths, each separated by a delay x, and the overbar indicates phase inversion of the pulse. For off-resonance spins the pulse elements are additive at the exeitation maximum so for example, should one require 90° off-resonance excitation, 1-1 corresponds to the sequence 45x-t-45 x. Of this binomial series, it turns out that the 1-3-3-1 sequence [66] has good performance and is most tolerant of pulse imperfections by virtue of its symmetry [67]. The trajectory of spins with frequency offset l/2x from the transmitter for a net 90° pulse (1 = 11.25°) is shown in Fig. 9.26. During each [Pg.362]

Whilst maximum excitation occurs at l/2t Hz from the transmitter offset, further nulls occur at offsets of n/x (n = 1, 2, 3,. .. corresponding to complete revolutions during each x) so a judicious choice of x is required to provide excitation over the desired bandwidth. The excitation profiles of the 1-1 and 1-3-1 sequences are shown in Fig. 9.25b and c. Clearly the excitation is non-uniform, so places limits on quantitative measurements, and once again there exists a phase inversion either side of the solvent. Both provide an effective null at the transmitter offset and suppression ratios in excess of 1000-fold can be achieved. [Pg.363]

When implementing this sequence it may be necessary to add attenuation to the transmitter to increase the duration of each pulse so that the shorter elements do not demand very short ( 1 xs) pulses (note the similarity with the requirements for the DANTE hard-pulse selective excitation described above). The binomial sequences can be adjusted to provide an arbitrary overall tip angle by suitable adjustment of the tip angles for each element. For example, inversion of all off-resonance signals can be achieved by doubling all elements relative to the net 90 condition. Exactly this approach has been exploited in the gradient-echo methods described below. [Pg.363]

The most effective approach to date to solvent suppression is the destruction of the net solvent magnetisation by pulsed field gradients (PFGs), so ensuring nothing of this remains observable immediately prior to acquisition. The [Pg.363]


The excited state redox potential of a sensitizer plays an important process. An approximate value of the excited state redox potential potentials of the ground state couples and the zero-zero excitation Equations (13) and (14). The zero-zero energy can be obtained from of the sensitizer 38 role in the electron transfer can be extracted from the energy (E0 0) according to the 77 K emission spectrum... [Pg.725]

Figure 11.19 Average total number of neutrons, vT, as a function of the mass number of the fissioning system for spontaneous and thermal neutron-induced fission. The values for thermal neutron-induced fission have been corrected to zero excitation energy (spontaneous fission) assuming dv jdt = 0.12 MeV-1. Figure 11.19 Average total number of neutrons, vT, as a function of the mass number of the fissioning system for spontaneous and thermal neutron-induced fission. The values for thermal neutron-induced fission have been corrected to zero excitation energy (spontaneous fission) assuming dv jdt = 0.12 MeV-1.
It is possible to shape an excitation profile with weak pulses in such a way as to have zero excitation at a certain frequency. This was first due to Redfield [9],... [Pg.307]

This main difficulty in coupling HPLC to NMR spectroscopy is faced by methods known as solvent suppression techniques, where the large solvent signals are reduced by special pulse sequences, switched prior to the information-selecting and acquisition pulses. Therefore, many efforts have been made to develop effective and minor-disturbing pulse sequences, such as presaturation, zero excitation and PFG-pulse sequences (WET) (see Chapter 1 and the following chapters). Despite the possibility of also suppressing several of the... [Pg.195]

In binomial composite pulse sequences the amplitudes of the pulses are proportional to the binomial coefficients of -t-1 square pulses separated by equal interpulse delays, r. These sequences provide broad regions of near-zero excitation at the solvent frequency while the solute resonances are... [Pg.322]

Generator impedance characteristic with zero Excitation. [Pg.320]

Note that there is no energy dissipating mechanism (zero damping matrix) included in this model. Figure 1.2 shows the free vibration response of the mass M50 in the middle of the chain with an zero initial condition except that the mass at the right end (Afioo) has an initial displacement of 0.01 m. The response appears to be random even though the system is deterministic with a simple initial condition and zero excitation. Figures 1.3 and 1.4 zoom into... [Pg.5]

Total number of prompt neutrons emitted (vj) as a function of the mass of the compound nuclei. Full circles spontaneous fission, blank circles thermal-neutron-induced fission corrected to zero excitation energy using dvj/6Ex = 0.11 MeV - from (Hoffman 1989)... [Pg.259]

A = 1 is then no longer the boundary between positive and negative increments of An a more appropriate criterion is Aq/(Mq +1) where A is the value of A for zero excitation and Mq the trap concentration. Indeed, the resistance increase due to "back diffusion" is expected to occur even more prominently in lifetime than in relaxation semiconductors. In practice, Aq and Mq are not, of course, independent quantities. The net effect of their interrelationship is to make the outcome less dependent on trap concentration than one would otherwise expect. [Pg.325]

The main disadvantage of the phase shift method is the influence of induced emission [see (11.11)] which is a handicap especially when using lasers for excitation [11.17]. The necessary extrapolation to zero excitation intensity is time consuming and limits the accuracy because of the low signal -to -noise ratio at low intensities. [Pg.558]

Absolute values for the ti iplet quantum yield were obtained by extrapolating the slope of the saturation curves to zero excitation intensity and converting the measured laser intensity to the number of incident photons per cm using free chlorophyll a in Triton X-100 at 77K as a calibration standard. For the calculation the following values were assumed triplet quantum yield d>chi = 0.6 and echi = 100 000 M cm" The number of triplets per pigment were approximated by the ratio of the integrals of both the AA (at time zero) and A spectra from 648 to 690 nm for free chlorophyll and from 655 to 695 nm for the complexes. [Pg.413]


See other pages where Zero excitation is mentioned: [Pg.97]    [Pg.195]    [Pg.75]    [Pg.285]    [Pg.11]    [Pg.120]    [Pg.823]    [Pg.32]    [Pg.129]    [Pg.299]    [Pg.4]    [Pg.362]    [Pg.362]    [Pg.125]    [Pg.272]    [Pg.273]    [Pg.103]    [Pg.75]    [Pg.556]    [Pg.262]    [Pg.356]    [Pg.356]    [Pg.277]    [Pg.3643]    [Pg.230]    [Pg.34]    [Pg.187]    [Pg.206]    [Pg.174]    [Pg.64]    [Pg.243]   


SEARCH



Zero excited state

Zero-order approximation excited state

© 2024 chempedia.info