ZAF method

Fortunately, the physics of interaction of energetic electrons and X rays with solid matter is sufficiendy well understood to permit implementation of a quandtadve analysis procedure, the so-called ZAF method, which is based upon a combination of theory and empiricism, to calculate separate correction fiictors for each of the... [Pg.184]

Since the composition of the unknown appears in each of the correction factors, it is necessary to make an initial estimate of the composition (taken as the measured lvalue normalized by the sum of all lvalues), predict new lvalues from the composition and the ZAF correction factors, and iterate, testing the measured lvalues and the calculated lvalues for convergence. A closely related procedure to the ZAF method is the so-called ())(pz) method, which uses an analytic description of the X-ray depth distribution function determined from experimental measurements to provide a basis for calculating matrix correction factors. [Pg.185]

Ema data can be quantitated to provide elemental concentrations, but several corrections are necessary to account for matrix effects adequately. One well-known method for matrix correction is the zaf method (7,31). This approach is based on calculated corrections for major matrix-dependent effects which alter the intensity of x-rays observed at a particular energy after being emitted from the corresponding atoms. The zaf method corrects for differences between elements in electron stopping power and backscattering (the correction), self-absorption of x-rays by the matrix (the a correction), and the excitation of x-rays from one element by x-rays emitted from a different element, or in other words, secondary fluorescence (the f correction). [Pg.285]

The most traditional correction method, called the ZAF method, consists of calculating separately the effects of atomic number (Z), absorption (A) and X-ray fluorescence. The ZAF procedure has been successfully used for some thirty years. In the context of the development of light clement or surface layer analysis, however, the ZAF method reaches its theoretical limits. This is why the current trend is towards methods using a parameterisation of the curve cp(pz) providing more realistic results in extreme cases. [Pg.164]

The ZAF method relies extensively on a computer, similar to the fundamental parameter method. First, the weight fraction of elements in the specimen is estimated from their relative intensities (kl/ fflkl). Then, the (ZiAiFi) factor for each element is estimated and is used for calculating kt. The calculated kt will compare with the measured kt. The iteration of computation continues until the two values converge. [Pg.194]

The ZAF method is based on assumptions that the specimen is in bulk and its surface is microscopically flat. Thus, we should be cautious when a specimen does not meet these assumptions, such as a thin film specimen or a specimen with a rough surface. For example, with a thin film specimen, the matrix effect will be reduced and the intensity-weight fraction relationship will be simplified. [Pg.195]

Quantitative Electron-Probe Microanalysis. (V. D. Scott and G. Love, eds.) John Wiley Sons, New York, 1983. Taken from a short course on the electron microprobe for scientists working in the field. A thorough discussion of EDS and WDS is given, including experimental conditions and specimen requirements. The ZAF correction factors are treated extensively, and statistics, computer programs and Monte Carlo methods are explained in detail. Generally, a very useftd book. [Pg.133]

This method is well adapted to quantitative analysis, where automatic identification of the spectral lines can be made by very sophisticated visual displays. Principal corrections (called ZAF) relate to atomic number (Z), nature of the isotope (A) and fluorescence (F). In semi-quantitative analysis, the software can give an approximate composition of the sample without the need for reference standards. [Pg.279]

General for 2 and 3 Three dimensional intensity data collected on a Bruker SMART CCD diffractometer at 293(2) K MoK, radiation (0.71073 A) reflections corrected for Lorentz and polarisation effects and absorption corrections applied using SADABS structures solved by direct method and successive Fourier synthesis (SHELXS-97 and SHELXL-97) hydrogen atoms placed in calculated positions with fixed isotropic thermal parameters R = [(ZAF)/(ZF )] wR = Z[w(F "-F,2)2]/Z[w(F 2)2]i ... [Pg.339]

The exact measurement of concentration needs a ZAF correction method (see p. 1097). This considers the differences in the stopping power and the backscattering of electrons between specimen and standard (Z correction), in the depth distribution of X-ray generation and absorption (A) of X rays recorded under a takeoff angle of 40-60°, and in the excitation of X-ray quanta of the element of interest by characteristic X rays and continuum of the matrix fluorescence F),... [Pg.1125]


See other pages where ZAF method is mentioned: [Pg.194]    [Pg.196]    [Pg.546]    [Pg.549]    [Pg.194]    [Pg.196]    [Pg.546]    [Pg.549]    [Pg.160]    [Pg.542]    [Pg.248]   
See also in sourсe #XX -- [ Pg.194 ]