Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Valve, sample external loop

There are basically two types of LC sample valve, those with an internal loop and those with an external loop. Valves with an internal loop are normally designed to deliver sample volumes of less than one microliters. Valves with external loops can deliver sample volumes ranging from a few microliters to several milliliters or more. In general, LC sample valves must be able to sustain pressures up to 10,000 p.s.i., although they are likely to operate on a continuous basis, at pressures of 3,000 p.s.i. or less. [Pg.138]

For sample introduction, six-way injection valves are commonly used. The reconstituted sample extract is loaded in an external or internal loop (e.g., 20 pL loop), and introduced into the mobile phase by rotation of the valve. The external loop (>5 pL) is interchangeable in order to adjust the injection volume. For injection volumes below 5 pL, internal loops should be used. The typical injection volume is 1-20 pL onto the column. To... [Pg.172]

There are two types of sample valve commonly used in LC, the internal loop valve and the external loop valve. In order to improve the seating and eliminate leaks, the valve faces are sometimes made from appropriate ceramics. The internal loop valves are largely used with small bore columns, that is to say columns having internal diameters of less than 1.5 mm. The external loop valves are used for larger diameter columns up to semi-preparative columns. [Pg.293]

The external loop valve is the more commonly used sampling system and offers a wider choice of readily adjustable sample sizes. A modified form of the external loop sample valve has become very popular for quantitative LC analysis, a diagram of which is shown in Figure 3. [Pg.293]

The basic difference between this type of valve and the normal external loop sample valve is the incorporation of an extra port at the front of the valve. This port allows the injection of a sample by a syringe directly into the front of the sample loop. Position (A) shows the load position. Injection in the front port causes the sample to flow into the sample loop. The tip of the needle passes through the rotor seal and, on... [Pg.293]

Figure 3. The Modified External Loop Sample Valve... Figure 3. The Modified External Loop Sample Valve...
Although the problems associated with septum injectors can be eliminated by using stop-flow septumless injection, currently the most widely used devices in commercial chromatographs are the microvolume sampling valves (Fig. 8.3) which enable samples to be introduced reproducibly into pressurised columns without significant interruption of the mobile phase flow. The sample is loaded at atmospheric pressure into an external loop in the valve and introduced into the mobile phase by an appropriate rotation of the valve. The volume of sample introduced, ranging from 2 piL to over 100 /iL, may be varied by changing... [Pg.222]

As already stated, the valve system can take two basic forms, the internal loop sampling valve and the external loop sampling valve. A diagram of the internal loop valve which utilizes only four ports is shown in figure 14. [Pg.140]

Where peak dispersion has not been constrained to very small volumes the external sample loop injector can be used and the external loop sample system, which employs six ports, is depicted in figure 15. In the external loop sample valve, three slots are cut in the rotor so that any adjacent pair of ports can be connected. In the loading position shown on the left, the mobile phase supply is connected by a rotor slot to port (4) and the column to port (5) thus allowing mobile phase to flow directly through the column. In this position the sample loop is connected to ports (3) and (6). Sample flows from a syringe into port (1) through the rotor slot to the sample loop at port (6). At the same... [Pg.141]

Figure 30.4, displays the operation of a sample loop in two different modes i.e., (a) sampling mode and (b) injection mode. Here, the sample is loaded at atmospheric pressure into an external loop in the microvolumesampling valve, and subsequently injected into the mobile-phase by a suitable rotation to the valve. However, the volume of sample introduced usually ranges between 2 p/ to over 100 p/ but can be varied either by altering the volume of the sample loop or by employing specific variable-volume sample valves. [Pg.459]

The most common method of sample introduction in HPLC is via a rotary valve, e.g. a Rheodyne valve. A schematic diagram of a rotary valve is shown in Fig. 32.17. In the load position, the sample is introduced via a syringe to fill an external loop of volume 5, 10 or 20 L. While this occurs, the mobile phase passes through the valve to the column. In the inject position, the valve is rotated so that the mobile phase is diverted through the... [Pg.219]

Figure 4.28. The FI A split-loop injection technique, the concept of which is that the (external) loop of an injection valve is split into two sections a and 6), which in the LOAD position are filled with either sample (5) and reagent (/ ), respectively (a), or with a mixture of both (b). In the first variant, the dimensions of sections a and b determine the volumes, which, upon turning of the valve to position INJECT (90° turn of the figure), are propelled into the FI A system by carrier solution (C) in the latter variant the injected volume is determined by the size of section b while the premixed ratio of sample and reagent is a function of the two pumping rates Qr and Qs. Figure 4.28. The FI A split-loop injection technique, the concept of which is that the (external) loop of an injection valve is split into two sections a and 6), which in the LOAD position are filled with either sample (5) and reagent (/ ), respectively (a), or with a mixture of both (b). In the first variant, the dimensions of sections a and b determine the volumes, which, upon turning of the valve to position INJECT (90° turn of the figure), are propelled into the FI A system by carrier solution (C) in the latter variant the injected volume is determined by the size of section b while the premixed ratio of sample and reagent is a function of the two pumping rates Qr and Qs.
As mentioned earlier, external sample loops or internal bores or cavities are used to store the sample. Valves with internal sample storage have the disadvantage that only a fixed volume of sample can be injected, while valves with external sample loops do not have this disadvantage as the sample loop is interchangeable. On the other hand, however, the minimal injection volume is usually smaller with the internal cavity or bore storage. [Pg.71]

The external loop sample valve has three slots cut in the rotor so that any adjacent pair of ports can be connected. In the loading position, shown on... [Pg.179]

There are two types of sample valve in general use in analytical LC the internal loop valve and the external loop valve. The two types of sample valve are shown diagrammatically in Figure 3. [Pg.2635]

The two types of sample injection valve are the external sample loop and the internal sample loop. External sample loop valves are typically used for injecting vapor samples. The volume of sample injected can be as large as desired and as small as the physical distance between ports and tubing diameter permit. External loop sizes typically range from 2 pi up to 10 ml. [Pg.3865]

N2 acts as a pressure release valve to the external loop H. The sample valve time-delay unit R, contains three synchronous timers (1) a variable 3 to 60 synchronous process timer (Chronoset Type CF.D. Robinson Co.) (ii) a miniature, fixed cycle synchronous timer and (iii) a variable 30 s to 12 min synchronous process timer. Timers (i) and (ii) are linked, and are initiated by a push button. Timer (iii) operated independently and again is started by a second push button. The unit also incorporates a re-set button that re-sets timers (i) and (iii). [Pg.166]

Set timer (iii) (Figure 7.1) for 5 min. Open the split heater, move it forward on its travel and close it over the sample tube. Operate the push button for timer (iii). After the set time of 5 min the pilot valve, Mi, is triggered which operated the pneumatic switch valve, F, thus allowing helium carrier gas to flow around the external loop, H, through the sample tube and flushing the liberated volatile constituents on to the gas chromatographic column. The carrier gas remains routed around the external loop, H, until the re-set button is pressed. It is then diverted through the internal loop, G. [Pg.167]


See other pages where Valve, sample external loop is mentioned: [Pg.107]    [Pg.862]    [Pg.215]    [Pg.143]    [Pg.25]    [Pg.45]    [Pg.60]    [Pg.208]    [Pg.90]    [Pg.2823]    [Pg.109]    [Pg.124]    [Pg.259]    [Pg.269]    [Pg.441]    [Pg.25]    [Pg.45]    [Pg.178]    [Pg.2637]    [Pg.2637]    [Pg.2637]    [Pg.2068]    [Pg.429]    [Pg.165]    [Pg.287]    [Pg.643]   
See also in sourсe #XX -- [ Pg.179 ]




SEARCH



Loop valve

Sample external loop

Sample loops

Sample valve

Sampling valve

© 2024 chempedia.info