Tandem mass spectrometry with time-of-flight analyser

4 Tandem Mass Spectrometry with Time-of-Flight Analyser [Pg.134]

If mv is the mass of the precursor and m( the fragment mass, their respective kinetic energies kp and kr are, since they have the same velocity uLl, [Pg.134]

The penetration depth d in the reflectron is given by d = EJqE. Hence, for the precursor and the fragment, the relevant equations are [Pg.134]

TOF spectrometer fitted with a deflection electrode for precursor ion selection. The instrument can be operated in either the linear or the reflectron mode. The selection resolution is about 250, which for example gives a 4Th window at m/z 1000. [Pg.136]

The necessity to step the voltage applied to the reflectron to produce a complete spectrum consumes time, wastes sample and complicates the mass calibration procedure. Indeed, the acquisition of a PSD spectrum is at least 20 times longer than for a classic TOF spectrum and the main part of the sample is lost without contributing to the analysis. [Pg.137]