The variation of sensitivity between different sensors was also checked. Calibration curves with five different sensors were performed. A Relative Standard Deviation of 13, 13 and 42% of calibration slopes (sensitivity) were obtained for Cu, Pb and Cd respectively. These variations should have limited consequence on bias and precision when the standard addition method is used. However, for Cd, variations in the limit of quantification between two electrodes could be expected. Finally, the accuracy of the method was evaluated by the measurement of a SWIFT reference material used during the 2nd SWIFT-WFD Proficiency Testing exercise (Table 4.2.2). The reference value was chosen as the consensus value of the selected data population obtained after excluding the outliers. The performances of the device were estimated according to the Z-score (Z) calculation. Based on this score, results obtained with the SPEs/PalmSens method were consistent with those obtained by all methods for Pb and Cu ( Z < 2) while the result was less satisfactory for Cd (2 < Z < 3). [Pg.266]

Brunori C., Ipolyi I., Pellegrino C., et al., 2007. The SWIFT-WFD Proficiency Testing campaigns in support of implementing the EU Water Framework Directive. Trends Anal. Chern., 23, 993-1004. [Pg.350]