Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Slope detection

The reverse-phase analysis was carried out on a SUPELCOSIL LC-18, 3-/zm particle size, 150 X 4.6-mm ID column (solvent system A, acetonitrile B, acetonitrile-tetrahydrofuran-chloroform (50 27.5 22.5) linear gradient from 30% to 100% of B in 70 min, flow rate 0.5 ml/min) (Fig. 22). The upper part of Fig. 22 shows that various chain lengths (C12 to C24 with one-carbon increment) of PNB-TBDMS-OHFA separated well enough in 30 min for effective recovery of the components by an absorbance slope-detecting fraction collector-detector combination. The separation of the positional isomers present in the used mixture was only minor, and it did not interfere with the fractionation according to chain length. [Pg.202]

Slope Detection. Slope detection is often used for AFMs operated in air. The cantilever is then driven at its free eigenfrequency. After the approach the amplitude decreases by A A due to its being out of tune. The reduced amplitude is detected and used as the feedback signal (see Fig. 65). The z piezoadjustment, necessary to keep the amplitude at the setpoint value (e.g., 80% of Aq),... [Pg.94]

Fig. 65. Illustration of the slope detection principle by means of the resonance curve of the lever. In the vicinity of a sample the resonance ftequency shifts and leads to a reduced amplitude at the excitation frequency. Fig. 65. Illustration of the slope detection principle by means of the resonance curve of the lever. In the vicinity of a sample the resonance ftequency shifts and leads to a reduced amplitude at the excitation frequency.
Compound Linear range (decades) Sensitivity (cps/pg) Log-log slope Detection limit (pg element)... [Pg.67]

Threshold detection is easy to set up but is affected by changes in base line level, caused by long term drift or zero calibration errors. Slope detection is less affected by base line changes but may be inadvertently triggered by a small peak, shoulder, or other signal noise, not present when the set up parameters were entered. [Pg.27]

Tank bottom slope is important because sediment, water, and heavy phases settle at the bottom. Corrosion is usually the most severe at the bottom, and the design of the bottom can have a significant effect on the life of the tank. In addition, if the Hquid stock is changed, it is usually desirable to remove as much as the previous stock as possible. Therefore, designs that allow for the removal of water or stock and the ease of tank cleaning have evolved. In addition, specialized tank bottoms have resulted from the need to monitor and detect leaks. Tank bottoms in contact with the soil or foundations are one of the primary sources of leaks from aboveground tanks. [Pg.315]

The comparison of the results of very different methods has to be judged very precisely, as, e.g., the given thickness of a layer is a function of the limit of detection (EOD) of a method. Additionally, the detected areas vary from about 0.01 up to about 400 mm-. Therefore, the methods with a low level of detection and with a high sensitivity (high slope of the calibration function) give a higher value for the layer thickness. Furthermore, the layers are broadened with time by diffusion. [Pg.411]

FIGURE 6.15 Example of application of method of Lew and Angus [10]. (a) Dose-response data, (b) Clark plot according to Equation 6.27 shown, (c) Data refit to power departure version of Equation 6.27 to detect slopes different from unity (Equation 6.28). (d) Data refit to quadratic departure version of Equation 6.27 to detect deviation from linearity (Equation 6.29). [Pg.115]

Procedures can also be used to analyze straight lines with respect to slope and position, compare sample values to standard population means, compare methods, and detect differences in small samples. [Pg.254]

Once there is an appreciable amount of cells and they are growing very rapidly, the cell number exponentially increases. The optical cell density of a culture can then be easily detected that phase is known as the exponential growth phase. The rate of cell synthesis sharply increases the linear increase is shown in the semi-log graph with a constant slope representing a constant rate of cell population. At this stage carbon sources are utilised and products are formed. Finally, rapid utilisation of substrate and accumulation of products may lead to stationary phase where the cell density remains constant. In this phase, cell may start to die as the cell growth rate balances the death rate. It is well known that the biocatalytic activities of the cell may gradually decrease as they age, and finally autolysis may take place. The dead cells and cell metabolites in the fermentation broth may create... [Pg.82]


See other pages where Slope detection is mentioned: [Pg.144]    [Pg.600]    [Pg.133]    [Pg.171]    [Pg.719]    [Pg.144]    [Pg.95]    [Pg.3868]    [Pg.319]    [Pg.599]    [Pg.647]    [Pg.218]    [Pg.144]    [Pg.600]    [Pg.133]    [Pg.171]    [Pg.719]    [Pg.144]    [Pg.95]    [Pg.3868]    [Pg.319]    [Pg.599]    [Pg.647]    [Pg.218]    [Pg.461]    [Pg.564]    [Pg.912]    [Pg.1561]    [Pg.218]    [Pg.202]    [Pg.369]    [Pg.292]    [Pg.1]    [Pg.284]    [Pg.443]    [Pg.393]    [Pg.552]    [Pg.426]    [Pg.333]    [Pg.172]    [Pg.348]    [Pg.1137]    [Pg.23]    [Pg.221]    [Pg.232]    [Pg.243]    [Pg.246]    [Pg.248]    [Pg.240]    [Pg.612]    [Pg.769]    [Pg.125]   
See also in sourсe #XX -- [ Pg.94 ]




SEARCH



Slope

Sloping

© 2024 chempedia.info