Other expressions derived from the preceding ones and established with a view to replacing one parameter by another or to accommodate simplifications may also be employed to express the resolution. Therefore expression 1.27 is used in this way. [Pg.18]


If there is no or little information on the method s performance characteristics, it is recommended that the method s suitability for its intended use in initial experiments be proven. These studies should include the approximate precision, working range, and detection limits. If the preliminary validation data appear to be inappropriate, the method itself, the equipment, the analysis technique, or the acceptance limits should be changed. In this way method development and validation is an iterative process. For example, in liquid chromatography selectivity is achieved through selection of mobile-phase composition. For quantitative measurements the resolution factor between two peaks should be 2.5 or higher. If this value is not achieved, the mobile phase composition needs further optimization. [Pg.546]