Identify all the controls on the instrument using the instruction manual as a guide. [Pg.129]

Turn on the instrument and allow it to warm up for 20-30 minutes. Permit the sample chamber to purge with the counting gas mixture during this time. [Pg.129]

Count a standard sample for 1 minute and then increase the voltage by 50 V. [Pg.130]

Continue the procedure until the counts begin to again increase and stop. Further voltage increases will result in continuous discharge and damage to the chamber. [Pg.130]