Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Objective calibration function

Most analyses of kinetic data have the object of identifying the constants of a rate equation based on the law of mass action and possibly some mass transfer relation.. The law of mass action Is expressed In terms of concentrations of the participants, so ultimately the chemical composition must be known as a function of time. In the laboratory the chemical composition Is determined by some instrument that is suitably calibrated to provide the needed information. Titration, refractive index, density, chromatography, spectrometry, polarimetry, conductimetry, absorbance, magnetic resonance — all of these are used at one time or another to measure chemical composition. In some cases, the calibration to chemical composition is linear with the reading. [Pg.105]

First, there is the obvious objection that there may be no experimental values for the properties and systems of interest. Second (and almost as obvious) is the possibility that the experimental values are wrong. Third, the experimental values may in fact be derived from experimental measurements by a number of steps that involve assumptions or other theoretical calculations. All of these objections are important, but in one sense they are orthogonal to the real issue what if our calculations contain multiple sources of error that can cancel with one another We know already that any truncated one-particle space and truncated jV-particle space treatment has two sources of error, these two truncations. And there is no reason to suppose that the error from these two sources cannot cancel, indeed, from the early days of large-scale correlated atomic wave functions there is good evidence that they do cancel [35]. Hence even if there are absolutely reliable experimental values for the properties and molecules we want to consider, using them to calibrate theoretical methods may be useless unless we can establish whether we have a cancellation of errors or not. [Pg.345]

The plots of alcohol and water mixtures presented here serve to illustrate the usefulness of XRD for liquid identification from the molecular interference function s(x). The curves shown here differ from the s (x) discussed in Section 2.3.1. in that they portray the square of the ratio of s (x) of the sample to s(x) of a white scatterer , a calibration object of proprietary composition whose scattering characteristics are fairly constant over the x range of interest. Notwithstanding these manipulations, the plots have absolute ordinate scale. [Pg.214]

Mesoscale self-assembly We have not yet modeled MESA by computer, but with the wealth of experimental results we are in a position to develop believable computer simulations calibrated by experiment. The force of attraction between the objects is well understood mathematically in a number of cases [33,120] and in some systems it may be possible to measure these forces experimentally [149]. Some of the problems encountered in modeling molecular systems will also be encountered in modeling MESA. For example, finding global rather than local minima, the availability of computer time limiting how long the assembly can be modeled, and constructing potential functions for interactions that have not been determined... [Pg.38]

Continuous crystallizers must operate steadily at equilibrium to achieve the design requirements. This means the feed rate, production rate, slurry density, operating temperature, liquid level, and so on, should held constant as a function of time. To accomplish this result requires the crystallizer to be isolated from upstream or downstream variations and the instruments need to be continuously calibrated. To help accomphsh this objective, a 12- to 24-hour agitated feed tank needs to be installed before the crystallizer. [Pg.556]

Figure 3.19. Absolute determination of 8 by in situ autocorrelation. Experiments were performed with a mode locked femtosecond Ti sapphire laser. A prism pair (PC) was used to compensate the group delay dispersion (GDD) of the microscope objective. A long-pass filter eliminates residual argon pump light and Ti sapphire fluorescence. After two sequential beam expanders (BE), the beam was approximately 25 mm in diameter, which was sufficient to overfill the back aperture (10-mm diameter) of the objective. A long-pass dichroic mirror (DC) with reflectivity separates fluorescence from excitation light. The incident power at the sample was measured by recollimating the transmitted beam onto a calibrated power meter. Fluorescence was detected by a photomultiplier tube and recorded as a function of the interferometer delay. (From Ref. [366] with permission of the Optical Society of America.)... Figure 3.19. Absolute determination of 8 by in situ autocorrelation. Experiments were performed with a mode locked femtosecond Ti sapphire laser. A prism pair (PC) was used to compensate the group delay dispersion (GDD) of the microscope objective. A long-pass filter eliminates residual argon pump light and Ti sapphire fluorescence. After two sequential beam expanders (BE), the beam was approximately 25 mm in diameter, which was sufficient to overfill the back aperture (10-mm diameter) of the objective. A long-pass dichroic mirror (DC) with reflectivity separates fluorescence from excitation light. The incident power at the sample was measured by recollimating the transmitted beam onto a calibrated power meter. Fluorescence was detected by a photomultiplier tube and recorded as a function of the interferometer delay. (From Ref. [366] with permission of the Optical Society of America.)...
The objective of the ECD and NIMS experiments is to measure the molar response of different compounds as a function of temperature. From these data the fundamental kinetic and thermodynamic properties of the reaction of thermal electrons with molecules and negative ions can be determined. The measurement is carried out in the same manner as the calibration of any detector. Known amounts of a compound are injected into the chromatograph and purified on a column, they then enter the detector. The response of the detector is normalized to the number of moles injected. When obtaining physical parameters, the detector temperature is changed and the procedure repeated. Since the molar response can vary by three to four orders of magnitude, the concentrations of the test molecule and the conditions in the detector at different temperatures must be taken into account. [Pg.76]

The photometric calibration also contributes to the uncertainty of the measured spectrum. Flux standard stars are typically measured at widely spaced wavelengths (50 A is common), and the sensitivity function of the instrument is determined by fitting a low-order polynomial or spline to the flux points. Such fits inevitably introduce low-order wiggles in the sensitivity function, which will vary from star to star. Based on experience, the best spectrophotometric calibration yield uncertainties in the relative fluxes of order 2-3% for widely-spaced emission lines the errors may be better for ratios of lines closer than 20 A apart. Absolute fluxes have much higher uncertainties, of course, especially for narrow-aperture observations of extended objects. [Pg.174]

These values are plotted as a function of sample (object) number in Figure 6.24(a). In Figure 6.24(b) the predicted V5. actual tryptophan concentration plot using the two factor PLSl model is shown. Examining these two plots it should not be surprising that the lower and higher numbered samples have considerable influence on the calibration, they are after all at the extremes of... [Pg.207]

Relatively few studies of the effect of pressure on luminescence Hfetimes have been reported. Most studies have considered the lifetime of the R-lines ( E —> A2) of ruby as a function of pressure with the objective of extending its utility as a pressure calibrant [111, 242 - 247]. Since concentration quenching occurs in ruby, the ambient and high pressure R-line lifetimes depend on the Cr concentration. Dilute ruby (< 0.4 wt% Cr [248]) has a reported R-line lifetime of... [Pg.49]


See other pages where Objective calibration function is mentioned: [Pg.123]    [Pg.34]    [Pg.2875]    [Pg.910]    [Pg.65]    [Pg.71]    [Pg.467]    [Pg.361]    [Pg.74]    [Pg.50]    [Pg.371]    [Pg.113]    [Pg.134]    [Pg.9]    [Pg.118]    [Pg.257]    [Pg.172]    [Pg.452]    [Pg.334]    [Pg.29]    [Pg.290]    [Pg.791]    [Pg.426]    [Pg.118]    [Pg.165]    [Pg.203]    [Pg.11]    [Pg.153]    [Pg.657]    [Pg.241]    [Pg.130]    [Pg.657]    [Pg.186]    [Pg.473]    [Pg.11]    [Pg.985]    [Pg.249]    [Pg.733]    [Pg.330]    [Pg.86]   
See also in sourсe #XX -- [ Pg.96 ]




SEARCH



Object function

Objective function

© 2024 chempedia.info