Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Measurement uniformity

Most of the assumptions are based on idealized models, indicating the limitations of the mathematical methods employed and the quantity and type of experimental data available. For example, the details of the combinatorial entropy of a binary mixture may be well understood, but modeling requires, in large measure, uniformity so the statistical relationships can be determined. This uniformity is manifested in mixing rules and a minimum number of adjustable parameters so as to avoid problems related to the mathematics, eg, local minima and multiple solutions. [Pg.252]

Measuring Uniformity Except for cases in which a coating of one ingredient with another takes place, the theoretical end result of mixing will not be an arrangement in which one type of particle is directly next to a different type. Rather, the theoretical end result when random tumbling takes place will be a random mixture along the lines shown in Fig. 19-8. [Pg.1763]

Now, the next point to understand is the relation between the rate of the reaction (here measured uniformly for all the perovskites studied at an overpotential of 0.3 V) and the strength ofthe OH bond to the transition metal. It is made clear from Fig. 7.111 that the stronger the bond strength, the slower the reaction. This is a determinative... [Pg.563]

III) Kramer s Modulus—Kramer (1935), in studies on tractive forces involved in model channels, developed an expression for measuring uniformity which depends on the ratio of the areas above and below the 50 percent line, as indicated in Figure 13. Kramer s uniformity modulus is the ratio of the area below to the area above this line. The modu-... [Pg.58]

We now consider a 40% silicone oil premixed emulsion dispersed in an aqueous phase. In Fig. 9 the evolution of mean diameter is plotted as a function of the applied shear rate. The dispersed phase volume fraction is kept constant at 75%, while the emulsifier concentration in the continuous medium is varied from 15 wt % to 45 wt %. The error bars show the distribution width deduced from the measured uniformity. At a given shear rate, smaller droplets with lower uniformity are produced (see Fig. 9) when surfactant concentration increases. For example at 45% of Ifralan 205 the uniformity never exceeds 15% whatever the applied shear rate, whereas it is of the order of 25% for 15% of Ifralan 205. Some microscope pictures of the emulsions obtained are given in Fig. 10. To understand the evolution, we may argue that the continuous phase viscosity increases... [Pg.205]

Probably the first search method occurring to an engineer would be to space the n measurements uniformly in the interval (a, 6), i.e., at... [Pg.280]

Bed temperature readily measurable uniform gas circulation through bed uncertain... [Pg.771]

Electrochemical tests are often preferred to mass loss studies as a method of measuring uniform corrosion. Such tests are preferred because in theory they (1) provide a realtime measurement of the metaUic corrosion rate, (2) can provide time-corrosion rate data on a single coupon, and (3) are rapid to perform. Disadvantages of the electrochemical tests include the requirements for comparatively expensive equipment (versus mass loss tests) and higher levels of technical expertise for data analysis. Fmthermore, the data reduction requires the use of conversion-"constants," factors applied to the results of the electrical measurements to convert the data to a corrosion rate. These constants ... [Pg.208]

In concert with the increase in yield stress observed for irradiations performed at lower temperatures (<400°C), a significant reduction in uniform elongation is observed. Fig. 9.8 shows a compilation of measured uniform elongation for numerous FM steels tested at the irradiation temperature. A sharp drop of uniform elongation to... [Pg.344]

For the magnetic measurements, uniform fields could be produced both normal (Hj ) and parallel... [Pg.25]

When detectors were large, we confirmed the detector size and measured uniformity in a very direct way. We moved the image of a small circular aperture, or narrow slit, or a heated wire across the detector and noted the points at which the signal dropped to 50% of its maximum value. If the illuminated area is small compared to the detector, and if the detector response is uniform, and if the spot is symmetric, the 50% points occur when the spot is centered on the edges of the detector. A related method was to place a moveable knife edge in front of a uniform source, and move the image of the knife edge across the detector. [Pg.284]

It should be noted that profiles measured by common profilometers indicate only depth of profile. They do not measure uniformity or coverage of the grit, nor the sharpness of the peaks. Full coverage of the grit blast is indicated by lack of gloss on the metal surface when viewed at a flat grazing angle. [Pg.45]

Figure 9.60 shows a plot of flie measured uniformity value vs. flie model predieted uniformity value. Ideally all flie points would fall along flie solid eurve in the figure. Points are shown for flie model wifli all 6 eoeffieients evaluated and with only 4 coefficients evaluated (C4 = C5 = 0). There is very little differenee in the two sets of points again illustrating the point fliat no dependenee on the square of pressure or square of gas ratio is justified in the model by the experimental data. [Pg.455]


See other pages where Measurement uniformity is mentioned: [Pg.1639]    [Pg.515]    [Pg.29]    [Pg.1460]    [Pg.1958]    [Pg.275]    [Pg.23]    [Pg.1946]    [Pg.503]    [Pg.1643]    [Pg.231]    [Pg.174]    [Pg.32]    [Pg.454]   
See also in sourсe #XX -- [ Pg.65 ]




SEARCH



Blend uniformity measurement

Content uniformity measurements - self calibrating

Self-calibrating high-throughput content uniformity measurements

© 2024 chempedia.info