Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Main mass spectrometer signals

Second, a CO mass spectrometer signal appears when the temperature is above 530 C, further increases with the increasing temperature, and reaches a maximum at 680 C. Meanwhile, a decrease of the hydrogen mass spectrometer signal accompanied by simultaneous water evolution is observed. CO formation is mainly due to carbothermal reaction between reactive carbon atoms or groups on the carbon material [23] and Mo compounds in the presence of hydrogen. [Pg.980]

There are three main reasons for this choice. Firstly, it becomes more and more difficult to obtain recordable, molecular-ion signals from un-derivatized carbohydrates as their M, increases significantly above 3000. Secondly, the mass spectrometers that have been used in all high-mass-carbofiydrate studies published at the time of writing this article are not capable of very sensitive analysis above —3800 mass units (see later). Thirdly, at masses >4000, it is usually not practicable to work at the resolution necessary for adjacent peaks to appear as separate signals in the spectrum. To do so would require that the source and collector slits be narrowed to such a degree that there would be an unacceptable loss in sensitivity. Thus, spectra acquired at mass >4000 are usually composed of unresolved clusters. [Pg.36]

The mass spectroscopy signal of ethyl acetate ester mainly appears during the negative going scan and is delayed compared to the ethanol oxidation Faradic current (Fig. 37a). This delay was explained as an experimental artefact, namely slow ester permeation through the Teflon membrane which establishes the interface between the electrochemical setup and the mass spectrometer due to the large size of ester molecule. [Pg.465]

Although the diode array detects the same material a few seconds after the mass spectrometer, the software may be programmed with this time delay so that the two signals appear to be synchronised. Also, a second synchronisation is needed for the difference in time between the component being detected by the mass spectrometer (from the split flow) and the component arriving at the fraction collector (main flow). This timing is needed to ensure that the trigger from the mass spectrometer to start... [Pg.343]

As shown in Figure 15.1, there are three main components of every mass spectrometer. The ion source is used to produce gas-phase ions by capture or loss of electrons or protons. In the mass analyzer, the ions are separated according to their mlz ratios ions of a particular mlz value reach the detector, and a current signal is produced. This section describes the soft ionization sources, mass analyzers, and detectors that are used in experiments involving biological macromolecules. [Pg.297]

This recalibration will not actually reduce the amount of calculation needed, since one must still use equations 7.9, 7.12, and 7.13 to calculate each p in order to ensure that mass balance holds. In feet, frequent recalibration will add noticeably to the work, since for each recalibrated matrix M it will be necessary to recalculate the L and L+ matrices of equation 7.13. The main benefit of calculating the recalibration factors of equation 7.15 would be to track changes over time in the calibration of the mass spectrometer. Such changes could compensate for, or simply reveal, drift in the mass spectrometer that might need to be investigated. Additionally, of course, any individual set of recalibration factors that seemed particularly unusual might signal seme kind of breakdown of the mass spectrometer system. [Pg.148]

Like an optical spectrometer, a mass spectrometer contains a transducer that converts the beam of ions into an electrical signal that can then be processed, stored in the memory of a computer, and displayed or stored. Unlike mosi optical spectrometers, mass spectrometers require an elaborate vacuum system to main-lain a low pressure in all of Ihe components except the... [Pg.283]

The main challenge for CE detectors is the small diameter of the capillary and the small sample volumes encountered. Detection schemes employed for capillary electrophoresis include measurement of UV absorption, fluorescence and refractive index. Electrochemical signals and conductivity as well as radioactivity from radioisotopes have also been measured. The signals obtained are plotted against the migration time in the form of an electropherogram. In recent years, coupling of CE to a mass spectrometer (CE-MS) has been achieved. [Pg.73]

Time-of-flight (TOP) mass spectrometers are ideally suited for use with the MALDI technique because of their theoretically unlimited mass range, high ion transmission, and for the pulsed nature of the laser used in this method. Mass resolution is the ability of an instrument to separate the signals from ions of similar mass, expressed as the mass of a given ion divided by the full width at half maximum of the peak (fwhm). Resolution in MALDl-TOP MS is mainly restricted by the ionization process, rather than by instrumental limits, because the ions have a certain time-span of formation, a spatial distribution, and a kinetic energy spread. ... [Pg.435]

The main advantage of ion mobility analyzers is the addition of another dimension of separation to mass spectrometry. FAIMS improves signal-to-noise ratios by removing isobaric ions that have three-dimensional shapes that are different from those of the analyte. IMS enables the measurement of the cross-sectional areas of ions in addition to the dimensions of retention time, mass and intensity provided by the chromatograph and mass spectrometer, respectively. IMS can also improve the quality of spectra by separating species that overlap chromatographically and would otherwise give mixed spectra. [Pg.90]


See other pages where Main mass spectrometer signals is mentioned: [Pg.89]    [Pg.73]    [Pg.250]    [Pg.1138]    [Pg.812]    [Pg.587]    [Pg.253]    [Pg.90]    [Pg.367]    [Pg.349]    [Pg.90]    [Pg.117]    [Pg.90]    [Pg.253]    [Pg.117]    [Pg.90]    [Pg.96]    [Pg.97]    [Pg.137]    [Pg.438]    [Pg.117]    [Pg.181]    [Pg.378]    [Pg.138]    [Pg.451]    [Pg.138]    [Pg.159]    [Pg.413]    [Pg.114]    [Pg.295]    [Pg.1073]    [Pg.438]    [Pg.147]    [Pg.35]    [Pg.812]    [Pg.365]    [Pg.368]    [Pg.485]    [Pg.850]    [Pg.270]    [Pg.104]   
See also in sourсe #XX -- [ Pg.119 ]




SEARCH



Mass signals

© 2024 chempedia.info