Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Magnetic fields spectrometers

Firstly, the energy losses of the incident electrons which produce the inner shell excitations may be detected as peaks in electron energy loss spectroscopy (EELS). The elecrons transmitted by the specimen are dispersed in a magnetic field spectrometer and the peaks, due to K, L and other shell excitations giving energy losses in the range of 0-2000eV, may be detected and measured. [Pg.332]

The potential of the suggested technique was also confirmed by FMR (Fig. 10.4). Area under an absorption curve (FMR spectrum intensity) is proportional to magnetic susceptibility x, which relates to magnetization M as M = yfj, where 7/ is an external magnetic field (spectrometer field). Total magnetization of magnetic suspension (M ) is in a proportion to magnetization of individual particle (M and to volume concentration of these particles (q>). [Pg.154]

The most widely used type of trap for the study of ion-molecule reactivity is the ion-cyclotron-resonance (ICR) [99] mass spectrometer and its successor, the Fourier-transfomi mass spectrometer (FTMS) [100, 101]. Figure A3.5.8 shows the cubic trapping cell used in many FTMS instmments [101]. Ions are created in or injected into a cubic cell in a vacuum of 10 Pa or lower. A magnetic field, B, confines the motion in the x-y... [Pg.810]

Figure Bl.7.18. (a) Schematic diagram of the trapping cell in an ion cyclotron resonance mass spectrometer excitation plates (E) detector plates (D) trapping plates (T). (b) The magnetron motion due to tire crossing of the magnetic and electric trapping fields is superimposed on the circular cyclotron motion aj taken up by the ions in the magnetic field. Excitation of the cyclotron frequency results in an image current being detected by the detector electrodes which can be Fourier transfonned into a secular frequency related to the m/z ratio of the trapped ion(s). Figure Bl.7.18. (a) Schematic diagram of the trapping cell in an ion cyclotron resonance mass spectrometer excitation plates (E) detector plates (D) trapping plates (T). (b) The magnetron motion due to tire crossing of the magnetic and electric trapping fields is superimposed on the circular cyclotron motion aj taken up by the ions in the magnetic field. Excitation of the cyclotron frequency results in an image current being detected by the detector electrodes which can be Fourier transfonned into a secular frequency related to the m/z ratio of the trapped ion(s).
Figure Bl.l 1.2. Simplified representation of an NMR spectrometer with pulsed RF and superconducting magnet. The main magnetic field Bq is vertical and centred on the sample. Figure Bl.l 1.2. Simplified representation of an NMR spectrometer with pulsed RF and superconducting magnet. The main magnetic field Bq is vertical and centred on the sample.
The original method employed was to scan eitiier the frequency of the exciting oscillator or to scan the applied magnetic field until resonant absorption occiined. Flowever, compared to simultaneous excitation of a wide range of frequencies by a short RF pulse, the scanned approach is a very time-inefficient way of recording the spectrum. Flence, with the advent of computers that could be dedicated to spectrometers and efficient Fourier transfomi (FT) algoritluns, pulsed FT NMR became the nomial mode of operation. [Pg.1470]

TOF mass spectrometers are very robust and usable with a wide variety of ion sources and inlet systems. Having only simple electrostatic and no magnetic fields, their construction, maintenance, and calibration are usually straightforward. There is no upper theoretical mass limitation all ions can be made to proceed from source to detector. In practice, there is a mass limitation in that it becomes increasingly difficult to discriminate between times of arrival at the detector as the m/z value becomes large. This effect, coupled with the spread in arrival times for any one m/z value, means that discrimination between unit masses becomes difficult at about m/z 3000. At m/z 50,000, overlap of 50 mass units is more typical i.e., mass accuracy is no better than about 50-100 mass... [Pg.191]

Certain regions of a mass spectrometer have no electric or magnetic fields to affect an ion trajectory (field-free regions). Figure 32.3 illustrates three such regions in a conventional double-focusing instrument. [Pg.226]

The system of electric and magnetic fields or lenses is called the ion optics of the mass spectrometer. Electric lenses correct aberrations in the shape of the ion beam. [Pg.405]

In practice, NMR spectrometers vary the magnetic field strength and measure the value at which a constant radio frequency is absorbed. [Pg.463]


See other pages where Magnetic fields spectrometers is mentioned: [Pg.95]    [Pg.36]    [Pg.60]    [Pg.234]    [Pg.481]    [Pg.130]    [Pg.146]    [Pg.735]    [Pg.32]    [Pg.249]    [Pg.305]    [Pg.95]    [Pg.36]    [Pg.60]    [Pg.234]    [Pg.481]    [Pg.130]    [Pg.146]    [Pg.735]    [Pg.32]    [Pg.249]    [Pg.305]    [Pg.252]    [Pg.64]    [Pg.64]    [Pg.1306]    [Pg.1311]    [Pg.1332]    [Pg.1355]    [Pg.1355]    [Pg.1445]    [Pg.1472]    [Pg.1474]    [Pg.1558]    [Pg.1560]    [Pg.1561]    [Pg.1564]    [Pg.1564]    [Pg.1569]    [Pg.2105]    [Pg.155]    [Pg.524]    [Pg.158]    [Pg.175]    [Pg.177]    [Pg.183]    [Pg.189]    [Pg.195]    [Pg.205]    [Pg.226]    [Pg.273]    [Pg.295]   
See also in sourсe #XX -- [ Pg.9 , Pg.10 ]




SEARCH



Magnetic spectrometer

© 2024 chempedia.info