Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Light measurement

The Fresnel equations predict that reflexion changes the polarization of light, measurement of which fonns the basis of ellipsometry [128]. Although more sensitive than SAR, it is not possible to solve the equations linking the measured parameters with n and d. in closed fonn, and hence they cannot be solved unambiguously, although their product yielding v (equation C2.14.48) appears to be robust. [Pg.2838]

The standard unit of light measurement (the light flux) is the lumen, which is the amount of energy (power, watts) emitted or received (Joules per second). At 555 nm, 1 lumen = 0.00147 watts = 0.00147 J.s-1. This is sometimes called a lightwatt . [Pg.118]

Analytical Applications. Chemiluminescence and bioluminescence are useful in analysis for several reasons. (/) Modem low noise phototubes when properly instmmented can detect light fluxes as weak as 100 photons/s (1.7 x 10 eins/s). Thus luminescent reactions in which intensity depends on the concentration of a reactant of analytical interest can be used to determine attomole—2eptomole amounts (10 to 10 mol). This is especially useful for biochemical, trace metal, and pollution control analyses (93,260—266) (see Trace and residue analysis). (2) Light measurement is easily automated for routine measurements as, for example, in clinical analysis. [Pg.274]

Scattering Coefficient measure of the ability of particles to scatter light measured in number proportional to the "amount" of light scattered per unit distance. [Pg.545]

Light is measured in several ways. Until about 1960, most light measurements were obtained by visual comparison. A standard light source of known, but adjustable, brightness was compared by a trained... [Pg.710]

To measure the total light emission of a photoprotein sample, it was necessary to add 1-5 pM Fe2+ several times due to the short effective life of Fe2+ under the conditions involved. The total amount of light measured in this manner was always proportional to the weight of photoprotein used, with both CPA and CPC (4.7 x 1013 photons/mg at 25°C Shimomura and Johnson, 1968d). Thus, the quantum yield is estimated at roughly 0.01 for CPA, and 0.015 for CPC. [Pg.224]

Fig. A.3. Light meter used by the author (Model 8020, Pelagic Electronics). For total light measurements, the signals are integrated with capacitors. Milliammeter reading is automatically reset at full-scale position, and the number of resets is digitally indicated below the meter. The box at the right contains a photomultiplier and sample compartment. Fig. A.3. Light meter used by the author (Model 8020, Pelagic Electronics). For total light measurements, the signals are integrated with capacitors. Milliammeter reading is automatically reset at full-scale position, and the number of resets is digitally indicated below the meter. The box at the right contains a photomultiplier and sample compartment.
Stanley, P. E. (1992). A survey of more than 90 commercially available luminometers and imaging devices for low-light measurements of chemiluminescence and bioluminescence, including instruments for manual, automatic and specialized operation, for HPLC, LC, GLC and microtiter plates. Part I descriptions. T. Biolumin. Chemilumin. 7 77-108. [Pg.439]

Figure 28. Layout of the bench for scattered light measurements (right), and a VIRGO bulk being measured (left). Figure 28. Layout of the bench for scattered light measurements (right), and a VIRGO bulk being measured (left).
We can siunmarize all of the above research carried out over the last two centuries in that a photon is a qucuitum of radiation and a carrier of force between particles, whereas an electron is a quantum of matter. Now, let us examine the more mundane aspects of light measurement including color measurement. [Pg.411]

The main technique that has been used for the measurement of opacity has been to prepare a standard disc of AB cement 1-0 mm thick and aged for 24 hours at 37 °C. This disc, contained in a small trough of water to prevent desiccation, is placed in a reflectometer on a black background. It is then illuminated with diffuse light and the amount of light reflected from it, is measured. The disc is then placed on a white background of 70% reflectivity, and the new amount of reflected light, measured. The contrast ratio 7 o/ o is defined as the Cp., opacity (Crisp et al., 1979). [Pg.380]

Van de Kraats, J. et al. (1996). The pathways of light measured in fundus reflectometry. Vision Research 36 2229-2247. [Pg.85]

A. Ryer, Light Measurement Handbook, International Light, Newburyport, MA, 1997. [Pg.636]

Diffused silicon sensor, 20 655 Diffusely reflected light, measuring, 24 111... [Pg.268]

ICP is an emission technique, which means that it does not use a light source. The light measured is the light emitted by the atoms and monoatomic ions in the atomizer. The ICP atomizer is an extremely hot plasma, which is a high-temperature ionized gas composed of electrons and positive ions confined by a magnetic held. The extremely high temperature means that the atoms and monoatomic ions undergo sufficient excitation (and de-excitation) such that relatively intense emission spectra result. The sample is drawn in with a vacuum mechanism that will be described. The intensity of an emission line is measured and related to concentration. [Pg.247]

Table 2.3 Stray light measurements for various instruments measured at 370 nm... Table 2.3 Stray light measurements for various instruments measured at 370 nm...
The high sensitivity of lock-in amplification can be applied to detect the small periodic changes in transmittance caused by modulated e.xcitation with UV light. Measurements of the amplitude and phase shift of the response signal allow us to determine the spectra and lifetimes of the transient species. [Pg.25]

Light measurement offers the combined capability of rapidly predicting by nondestructive means dust and trash content in cotton and airborne dust level. Of course, the standard error of estimate is not a practical statistic based on only six cottons and is not reported in this feasibility paper. [Pg.83]

Figure 14. Water column analysis involving pump profiling system. In the system pictured, conductivity (an index of salinity) and temperature are profiled. The fluorometer measures ij vivo fluorescence from chlorophyll (from all phytoplankton) and discrete samples are taken for cell counts and nutrient chemistry. Photodetectors can be employed for the measurement of bioluminescence. Light measurements range from Secchi disc readings to more sophisticated transmissometry and spectral radiometry instruments. Figure 14. Water column analysis involving pump profiling system. In the system pictured, conductivity (an index of salinity) and temperature are profiled. The fluorometer measures ij vivo fluorescence from chlorophyll (from all phytoplankton) and discrete samples are taken for cell counts and nutrient chemistry. Photodetectors can be employed for the measurement of bioluminescence. Light measurements range from Secchi disc readings to more sophisticated transmissometry and spectral radiometry instruments.

See other pages where Light measurement is mentioned: [Pg.1880]    [Pg.766]    [Pg.407]    [Pg.717]    [Pg.444]    [Pg.76]    [Pg.710]    [Pg.41]    [Pg.361]    [Pg.369]    [Pg.457]    [Pg.82]    [Pg.74]    [Pg.88]    [Pg.55]    [Pg.208]    [Pg.101]    [Pg.478]    [Pg.539]    [Pg.239]    [Pg.171]    [Pg.139]    [Pg.632]    [Pg.307]    [Pg.52]    [Pg.83]    [Pg.69]    [Pg.120]    [Pg.31]    [Pg.592]   
See also in sourсe #XX -- [ Pg.7 , Pg.164 ]

See also in sourсe #XX -- [ Pg.144 ]

See also in sourсe #XX -- [ Pg.112 ]




SEARCH



© 2024 chempedia.info