Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Light absorption, measurable variables

An interferometric method was first used by Porter and Topp [1, 92] to perfonn a time-resolved absorption experiment with a -switched ruby laser in the 1960s. The nonlinear crystal in the autocorrelation apparatus shown in figure B2.T2 is replaced by an absorbing sample, and then tlie transmission of the variably delayed pulse of light is measured as a fiinction of the delay This approach is known today as a pump-probe experiment the first pulse to arrive at the sample transfers (pumps) molecules to an excited energy level and the delayed pulse probes the population (and, possibly, the coherence) so prepared as a fiinction of time. [Pg.1979]

The next step in this study is to test this control algorithm on the actual laboratory reactor. The major difficulty is the direct measurement of the state variables in the reactor (T, M, I, W). Proposed strategy is to measure total mols of polymer (T) with visible light absorption and monomer concentration (M) with IR absorption. Initiator concentration (I) can be monitored by titrating the n-butyl lithium with water and detecting the resultant butane gas in a thermal conductivity cell. Finally W can be obtained by refractive index measurements in conjuction with the other three measurements. Preliminary experiments indicate that this strategy will result in fast and accurate measurements of the state vector x. [Pg.201]

Optical measurements of airborne combustion aerosols have been carried out for a number of years, usually with light scattering techniques. However, due to the particle size dependence of light scattering and the variable particle size distributions of smokes, it is extremely difficult to relate light scattering properties to particulate mass concentrations. The measurement of light absorption by particles can be directly related to particle mass if two conditions are met ... [Pg.457]

Spreadsheet Summary In Chapter 12 of Applications of Microsoft Excel in Analytical Chemistry, spreadsheets are presented for modeling the effects of chemical equilibria and stray light on absorption measurements. Chemical and physical variables may be changed to observe tlieir effects on instrument readouts. [Pg.734]

Three measurable variables are available for characterizing the degree of light absorption, i.e. the light intensity absorbed by a "dissolved" sample substance. These are ... [Pg.90]

Extinction E is the most frequently used measurable variable for light absorption. It is the decimal logarithm of the ratio of the intensity of the irradiated light (Iq) to the intensity of the light beam leaving the sample solution (I) ... [Pg.91]

An attempt was made to find robust calibration equations for sugar content and acidity by MLR, PCR, and PLS, where the optical parameters were employed as explanatory variables. Normally, chemometrics by NIR spectra employs the absorbance as the explanatory variable, where only wavelength-dependent characteristics of the materials can be considered. In this case, it is very difficult to precisely evaluate the small amount of a constituent such as acid content in a fruit. On the other hand, chemometrics by TOF-NIRS would be related to both wavelength- and time-dependent characteristics as the explanatory variables, where the light absorption and light scattering phenomena in a sample are included. It may therefore be possible to detect the acid content in a fruit on the basis of this new optical concept. The statistical results are summarized in Tables 4.2.1 and 4.2.2. Figure 4.2.9 shows the PLS analysis in a optimum model for acidity in apple. In the case of normal analysis by second-derivative NIR spectra, standard error of calibration (SEC) and correlation coefficient between measured and predicted acidity r were limited to 0.048% and... [Pg.116]

It is slow and tedious for an experiment in which a molecular absorption spectrum is measured. It is slow and tedious because the wavelengths are manually scanned in small increments, and each time the wavelength is changed, the calibration step with the blank needs to be performed due to the variability of light intensity from the light source at the different wavelengths. [Pg.521]

As is the case for LIF, calibration to obtain absolute concentrations is a challenge. In the instrument shown in Fig. 11.45, a calibration source based on the photolysis of water at 185 nm is installed in the inlet. From the absorption cross section of HzO gas at 185 nm, its concentration, the light intensity, and the sample flow rate, the concentration of OH generated by the photolysis can be calculated. However, not only is there significant uncertainty in the absorption cross section for HzO at 185 nm (e.g., see Lazendorf et al., 1997 Hofzumahaus et al., 1997, 1998 and Tanner et al., 1997), but the measured calibration factor was highly variable from day to day, by as much as a factor of two (Tanner et al., 1997). [Pg.602]

Any physical variable giving an accurate measure of the extent to which a reaction has gone toward completion can be nsed to obtain rate data. In this experiment we shall monitor the concentration of unreacted diazonium ion by measuring the absorption of nltraviolet light by the solution. [Pg.284]


See other pages where Light absorption, measurable variables is mentioned: [Pg.214]    [Pg.375]    [Pg.90]    [Pg.166]    [Pg.245]    [Pg.126]    [Pg.409]    [Pg.309]    [Pg.297]    [Pg.357]    [Pg.364]    [Pg.90]    [Pg.504]    [Pg.99]    [Pg.146]    [Pg.222]    [Pg.1789]    [Pg.371]    [Pg.414]    [Pg.141]    [Pg.661]    [Pg.290]    [Pg.391]    [Pg.44]    [Pg.216]    [Pg.139]    [Pg.654]    [Pg.133]    [Pg.680]    [Pg.5]    [Pg.228]    [Pg.321]    [Pg.140]    [Pg.654]    [Pg.140]   
See also in sourсe #XX -- [ Pg.90 ]




SEARCH



Absorption light

Absorption variability

Measurable absorption

Measurable variables of light absorption

Variability measurement

Variability, measures

Variables measurement

© 2024 chempedia.info