Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Dropping speed

During a run, if the supply voltage to a motor terminal drops to 85% of its rated value, then the full load torque of the motor will decrease to 72.25%. Since the load and its torque requirement will remain the same, the motor will star to drop speed until the torque available on its speed-torque curve has a value as high as 100/0.7225 or 138.4% of T to sustain this situation. The motor will now operate at a higher slip, increasing the rotor slip losses also in the same proportion. See equation (1.9) and Figure 1.7. [Pg.11]

If a system disturbance requires extra generating capacity, the generator may drop speed, reducing the frequency proportionately. The under-frequency relay 81-G will detect such a condition and operate a systematic load shedding in a programmed manner in order to meet the load demand. If the abnormal condition persists, the generator is taken off the main supply system. [Pg.514]

At extreme overpotentials, the current is independent of potential. This maximum current is said to be limiting, that is, current a Cbuik- It is termed the diffusion current, /j. The dependence of la on concentration, drop speed, etc., is described by the Ilkovic equation (equation (6.5)), although calibration graphs or standard addition methods (Gran plots) are preferred for more accurate analyses. [Pg.194]

In regard to parison control, a compromise is necessary between the desired net weight and the need to maintain a sufficient safety margin over a set of minimum specifications, which include minimum wall thickness, drop speed, drop strength, dimensional stability, and fluctuations in net weight. Most of these parameters can be directly affected by the molder s ability to control the parison wall thickness. The most common and practical way of doing this has been to adjust the gap between the die and mandrel (Table 4-2). [Pg.179]

Jet speed. Speed can refer to both the speed at which the ink drop is fired, described here as the drop speed, and to the interval between firing events, described here as the drop frequency. The speed of the falting drop may vary according to ink formulation as well as... [Pg.74]

A dropper insert is a dropping mechanism in the neck of the bottle (see Fig. 24.13). Many dropper inserts turn out not to meet the requirements of the Ph. Eur. the dropping speed is too high or they don t provide drops at all. The reproducibility of the drop weight is often poor. The drop size appears to depend for example on the angle at which the bottle is held and on the amount of fluid in the bottle. [Pg.530]

Repeat the drop-speed times as in step 3 and again calculate the average of the 5 or 10 drop times and mean deviation for the wetting torpedo . [Pg.351]

Under a nitrogen atmosphere, acrylonitrile (15.919 g, 0.300 mol), TMEDA (1.047 g, 9.01 mmol), triethylamine (0.912 g, 9.01 mmol), and copper(I) chloride (1.78 g, 18.0 mmol) were placed in a 100 mL three-necked round-bottomed flask equipped with a stir bar, a dropping funnel, a reflux condenser, and a thermometer. Trichlorosilane (44.70 g, 0.330 mol) was added dropwise to this mixture for 1 h. As an exothermic reaction took place, the dropping speed of trichlorosilane was adjusted so that the temperature of the mixture was kept below 60°C. After the addition of trichlorosilane, the mixture was stirred at room temperature for 4 h. The reaction mixture was distilled imder reduced pressure to give trichloro-2-cyanoethylsilane (52.86 g, 93%). Bp 86°C/llmmHg. The distillate solidified on standing at room temperature for several hours. [Pg.464]

The general analysis, while not difficult, is complicated however, the limiting case of the very elongated, essentially cylindrical drop is not hard to treat. Consider a section of the elongated cylinder of volume V (Fig. II-18h). The centrifugal force on a volume element is u rAp, where w is the speed of revolution and Ap the difference in density. The potential energy at distance r from the axis of revolution is then w r Apfl, and the total potential energy for the... [Pg.30]

To the remainder of the casein solution add 0 5 to o 8 g. of finely powdered commercial trypsin, shake to dissolve, and place in a thermostat (or in an incubator) at 40 . After 15 minutes, remove 25 ml. and add a few drops of phenolphthalein it will now be found that the solution remains colourless. Run in carefully Mj 10 NaOH solution until the colour of the solution is just pink, add 5 ml, of neutralised formalin and then titrate against Mj 10 NaOH solution until the pink colour is just restored note the amount required. Remove fiirther quantities (rf 25 ml. at intervals which must be determined by the speed of the reaction. The following will probably make a suitable series i, 2, 3,... [Pg.518]

We start rxn, one drop / second or so C in B. Sometimes we close sep funnel and shake flask B to ensure a constant rate of MeONO generation. Addition speed is limited by equilibrium of pressure between flasks. If it is too much quick, then MeONO gas go through sep. funnel, then we close the sep funnel and wait a bit till generation is low. The addition of C in B takes 1 hour, we close sep funnel and shake a bit B to finish reaction. If rxn (A) climbs temp too much, we can add ice in the water bath. I ve monitorized temp touching a part of solution that was out of water bath. At the final part may be water is to much cool, so we can take it out. After the addition of C in B we wait one more hour. [Pg.85]

The function of aeration in a wastewater treatment system is to maintain an aerobic condition. Water, upon exposure to air, tends to estabUsh an equihbrium concentration of dissolved oxygen (DO). Oxygen absorption is controlled by gas solubiUty and diffusion at the gas—hquid interface. Mechanical or artificial aeration may be utilised to speed up this process. Agitating the water, creating drops or a thin layer, or bubbling air through water speeds up absorption because each increases the surface area at the interface. [Pg.339]

The fundamental case for pressure filters may be made using equation 10 for dry cake production capacity Y (kg/m s) derived from Darcy s law when the filter medium resistance is neglected. Eor the same cycle time (same speed), if the pressure drop is increased by a factor of four, production capacity is doubled. In other words, filtration area can be halved for the same capacity but only if is constant. If increases with pressure drop, and depending how fast it increases, the increased pressure drop may not give much more capacity and may actually cause capacity reductions. [Pg.393]

Measurement Requirements. Any analysis of measurement requirements must begin with consideration of the particular accuracy, repeatabihty, and range needed. Depending on the appHcation, other measurement considerations might be the speed of system response and the pressure drop across the flow meter. For control appHcations repeatabihty may be the principal criterion conversely for critical measurements, the total installed system accuracy should be considered. This latter includes the accuracy of the flow meter and associated readout devices as well as the effects of piping, temperature, pressure, and fluid density. The accuracy of the system may also relate to the required measurement range. [Pg.56]

The feed is normally introduced to the top hearth where the rabble arms and teeth attached to the central shaft rotate and spiral soflds across the hearth to the center, where an opening is provided and the soflds drop to the next hearth. The teeth of the rabble arms on the hearth spiral the soflds toward the outside to ports that let the soflds drop down to the next hearth. Soflds continue downward, traversing each hearth until they reach the bottom and the ash is discharged. The primary advantage of this system is the long residence time in the furnace controlled by the speed of the central shaft and pitch of the teeth. [Pg.46]


See other pages where Dropping speed is mentioned: [Pg.415]    [Pg.166]    [Pg.529]    [Pg.415]    [Pg.166]    [Pg.529]    [Pg.20]    [Pg.30]    [Pg.42]    [Pg.438]    [Pg.1904]    [Pg.10]    [Pg.81]    [Pg.434]    [Pg.699]    [Pg.65]    [Pg.233]    [Pg.365]    [Pg.31]    [Pg.420]    [Pg.355]    [Pg.206]    [Pg.410]    [Pg.105]    [Pg.106]    [Pg.106]    [Pg.108]    [Pg.108]    [Pg.257]    [Pg.389]    [Pg.94]    [Pg.100]    [Pg.16]    [Pg.69]   
See also in sourсe #XX -- [ Pg.529 , Pg.530 ]




SEARCH



© 2024 chempedia.info