Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Dispersion apparant

An alternative technique for the measurement of sc is to study the photopotential. The theory of this is discussed in some detail in Sect. 7, but the essential features of the measurement are shown in Fig. 16. After equilibration in the dark, when the potential of the electrode at open circuit becomes equal to the redox potential Vredox, the light is turned on and the electrode potential changes at open circuit in such a way that the bands become flat. There are many problems with this technique and it is considerably less reliable than a properly conducted a.c. experiment, but it may give a reasonably accurate picture if surface recombination is small (vide infra). Some results for p-GaAs in aqueous solution are shown in Fig. 17 and the S values derived are of the order 0.7, though the dispersion apparent in Fig. 17 makes a quantitative interpretation difficult. [Pg.89]

Selective flocculation. Suppose X23 is positive but now the dispersion medium is adjusted to be a worse than 0-solvent for, say, polymer 2 (Xi2>2 but Xi3 < 2)- Homoflocculation of the particles coated by polymer 2 will ensue. Provided that xii is not excessively large, heteroflocculation would be expected to be absent. This implies that it should be possible under such conditions to flocculate selectively one species of sterically stabilized particles in a mixed dispersion. Apparently polymer coated particles are able to discriminate between self and not-selF when undergoing Brownian collisions in such dispersion media, coagulation only occurring between particles of the same type. Of course, if the dispersion medium were made a worse than 0-solvent for both polymers (xi2. Xi3>tX both 2-2 and 3-3 homoflocculation would occur. Heteroflocculation would only be possible if... [Pg.318]

These experiments yield T2 which, in the case of fast exchange, gives the ratio (Aoi) /k. However, since the experiments themselves have an implicit timescale, absolute rates can be obtained in favourable circumstances. For the CPMG experiment, the timescale is the repetition time of the refocusing pulse for the Tjp experiment, it is the rate of precession around the effective RF field. If this timescale is fast witli respect to the exchange rate, then the experiment effectively measures T2 in the absence of exchange. If the timescale is slow, the apparent T2 contains the effects of exchange. Therefore, the apparent T2 shows a dispersion as the... [Pg.2106]

Colloidal dispersions often display non-Newtonian behaviour, where the proportionality in equation (02.6.2) does not hold. This is particularly important for concentrated dispersions, which tend to be used in practice. Equation (02.6.2) can be used to define an apparent viscosity, happ, at a given shear rate. If q pp decreases witli increasing shear rate, tire dispersion is called shear tliinning (pseudoplastic) if it increases, tliis is known as shear tliickening (dilatant). The latter behaviour is typical of concentrated suspensions. If a finite shear stress has to be applied before tire suspension begins to flow, tliis is known as tire yield stress. The apparent viscosity may also change as a function of time, upon application of a fixed shear rate, related to tire fonnation or breakup of particle networks. Thixotropic dispersions show a decrease in q, pp with time, whereas an increase witli time is called rheopexy. [Pg.2673]

After the analyzer of a mass spectrometer has dispersed a beam of ions in space or in time according to their various m/z values, they can be collected by a planar assembly of small electron multipliers. There are two types of multipoint planar collectors an array is used in the case of spatial separation, and a microchannel plate is used in the case of temporal separation. With both multipoint assemblies, all ions over a specified mass range are detected at the same time, or apparently at the same time, giving these assemblies distinct advantages over the single-point collector in the analysis of very small quantities of a substance or where ions are produced intermittently during short time intervals. [Pg.410]

NTU p is the "exterior apparent" overall gas-phase number of transfer units calculated neglecting axial dispersion simply on the basis of equation 56, whereas NTU stands for the higher real number of transfer units (Nq ) which is actually required under the influence of axial dispersion. The correction factor ratio can be represented as a function of those parameters that are actually known at the outset of the calculation... [Pg.35]

The recommended rapid design procedure consists of the following steps (/) The apparent is calculated using equation 56. (2) The extent of axial dispersion is estimated from Hterature correlations for each phase in terms of Pe numbers and transformed into values. (3) The correction... [Pg.36]

Rates of nitration determined over a range of temperatures in two-phase dispersions have been used to calculate energies of activation from 59—75 kj/mol (14—18 kcal/mol). Such energies of activation must be considered as only apparent, since the tme kinetic rate constants, NO2 concentrations, and interfacial area all change as temperature is increased. [Pg.34]

Phenomena at Liquid Interfaces. The area of contact between two phases is called the interface three phases can have only aline of contact, and only a point of mutual contact is possible between four or more phases. Combinations of phases encountered in surfactant systems are L—G, L—L—G, L—S—G, L—S—S—G, L—L, L—L—L, L—S—S, L—L—S—S—G, L—S, L—L—S, and L—L—S—G, where G = gas, L = liquid, and S = solid. An example of an L—L—S—G system is an aqueous surfactant solution containing an emulsified oil, suspended soHd, and entrained air (see Emulsions Foams). This embodies several conditions common to practical surfactant systems. First, because the surface area of a phase iacreases as particle size decreases, the emulsion, suspension, and entrained gas each have large areas of contact with the surfactant solution. Next, because iaterfaces can only exist between two phases, analysis of phenomena ia the L—L—S—G system breaks down iato a series of analyses, ie, surfactant solution to the emulsion, soHd, and gas. It is also apparent that the surfactant must be stabilizing the system by preventing contact between the emulsified oil and dispersed soHd. FiaaHy, the dispersed phases are ia equiUbrium with each other through their common equiUbrium with the surfactant solution. [Pg.234]

In applying this concept, the factor of particle size must be continuously borne in mind. A heterodisperse system can reach a steady state wherein the smaller particles are agglomerated and the larger particles are dispersed, giving the apparent effect of an equiUbrium. In ideal monodisperse systems under steady conditions, however, no such effects are noted. [Pg.532]

Axial Dispersion Effects In adsorption bed calculations, axial dispersion effects are typically accounted for by the axial diffusionhke term in the bed conservation equations [Eqs. (16-51) and (16-52)]. For nearly linear isotherms (0.5 < R < 1.5), the combined effects of axial dispersion and mass-transfer resistances on the adsorption behavior of packed beds can be expressed approximately in terms of an apparent rate coefficient for use with a fluid-phase driving force (column 1, Table 16-12) ... [Pg.1516]

In practice, experimental peaks can be affected by extracolumn retention and dispersion factors associated with the injector, connections, and any detector. For hnear chromatography conditions, the apparent response parameters are related to their corresponding true column value by... [Pg.1532]


See other pages where Dispersion apparant is mentioned: [Pg.195]    [Pg.95]    [Pg.173]    [Pg.245]    [Pg.341]    [Pg.182]    [Pg.195]    [Pg.95]    [Pg.173]    [Pg.245]    [Pg.341]    [Pg.182]    [Pg.188]    [Pg.723]    [Pg.129]    [Pg.131]    [Pg.79]    [Pg.317]    [Pg.36]    [Pg.33]    [Pg.34]    [Pg.405]    [Pg.302]    [Pg.420]    [Pg.423]    [Pg.472]    [Pg.55]    [Pg.8]    [Pg.257]    [Pg.438]    [Pg.490]    [Pg.461]    [Pg.121]    [Pg.345]    [Pg.1225]    [Pg.1643]    [Pg.1652]    [Pg.1853]    [Pg.1857]    [Pg.128]    [Pg.10]    [Pg.32]    [Pg.6]    [Pg.193]   
See also in sourсe #XX -- [ Pg.307 ]




SEARCH



Apparent Dispersion from Detector Sensor Volume

Apparent dispersion, sensor

Detector apparent dispersion

Dispersion coefficient, apparent

© 2024 chempedia.info