1 Copper ions in the digest are subsequently dried, ashed and vaporised by electrical heating in a graphite furnace. The copper atoms thus formed absorb radiation from a hollow cathode lamp. The absorbance is measured at a wavelength of 327.4 nm. [Pg.84]

1 This procedure yields a standard curve that is linear up to approximately 50 pg/L Cu. [Pg.84]

2 The detection limit is approximately 1.2 pg/L in the digest The determination limit is approximately 3.6 pg/L (0.4 respectively 1.2 mg/kg in the dried plant material). [Pg.84]

1 To avoid interferences as much as possible, background correction is necessary (e.g. Zeeman background correction). [Pg.84]

Whereas the flame AAS determination normally uses 324.7 nm as wavelength for measurement, the graphite furnace determination has at 327.4 nm an improved curve [Pg.84]