Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Detector linear range

If the purity of the synthetic standards is questionable, the following is suggested use the (impure) ng/yl standards to obtain retention time and the detector linear range. Then, during the analysis, add the appropriate phenyltin chloride directly to a test tube and derivatize with lithium aluminium hydride, eliminating the usual extraction and concentration steps. Since the conversion occurs quantitatively, peak heights from these derivatization standards can be used to construct a standard curve. The procedure calls for... [Pg.327]

The amount of sample injected must not overload the column stationary phase nor exceed the detector linear range. A narrow toiling range sample will require a smaller amount injected than a wider toiling range sample. [Pg.449]

We take intensity I to mean x-ray quanta per second measured by a detector. An instantaneous detector measures / directly that is, the detector gives a reading always proportional to intensity over the linear range of the detector. The reading could appear on an ammeter after suitable amplification. [Pg.43]

Detector Linearity Linear Dynamic Range Detector Noise Level... [Pg.158]

There are a number of properties of a detector that determine whether they may be used for a particular analysis, with the most important being (a) the noise obtained during the analysis, (b) its limit of detection, (c) its linear range, and (d) its dynamic range. The last three are directly associated with the analyte being determined. [Pg.40]

When the linear range is exceeded, the introduction of more analyte continues to produce an increase in response but no longer is this directly proportional to the amount of analyte present. This is referred to as the dynamic range of the detector (see Figure 2.6). At the limit of the dynamic range, the detector is said to be saturated and the introduction of further analyte produces no further increase in response. [Pg.42]

Figure 2.6 Detector response curve showing (a) ideal behaviour, (b) real behaviour, (c) its linear range, (d) its dynamic range, (e) the noise level, and (f) the limit of detection at three times the noise level. Figure 2.6 Detector response curve showing (a) ideal behaviour, (b) real behaviour, (c) its linear range, (d) its dynamic range, (e) the noise level, and (f) the limit of detection at three times the noise level.
Transverse moving head of four-quadrant position detectors (Fig. 2), precise measuring lateral displacement which corresponds to lateral voltage F/ when it varies at the linear range of 10 V, thus we can compute the rotated angle a of the reflection ray ... [Pg.190]

A second source of error may be in the detector. Detector linearity is an idealization useful over a certain concentration range. While UV detectors are usually linear from a few milliabsorbance units (MAU) to 1 or 2 absorbance units (AU), permitting quantitation in the parts per thousand level, many detectors are linear over only one or two decades of operation. One approach in extending the effective linear range of a detector is high-low injection.58 In this approach, an accurate dilution of a stock sample solution is prepared. The area of the major peak is estimated with the dilution, and the area of the minor peak is estimated with the concentrated stock. This method, of course, relies on linear recovery from the column. Another detector-related source of error that is a particular source of frustration in communicating... [Pg.155]

Fluorescence detectors can be made much more sensitive than uv absorbance detectors for favourable solutes (such as anthracene) the noise equivalent concentration can be as low as 10 12 g cm-3. Because both the excitation wavelength and the detected wavelength can be varied, the detector can be made highly selective, which can be very useful in trace analysis. The response of the detector is linear provided that no more than about 10% of the incident radiation is absorbed by the sample. This results in a linear range of 103-104. [Pg.64]

Refractive index detectors are not as sensitive as uv absorbance detectors. The best noise levels obtainable are about 1CT7 riu (refractive index units), which corresponds to a noise equivalent concentration of about 10-6 g cmT3 for most solutes. The linear range of most ri detectors is about 104. If you want to operate them at their highest sensitivity you have to have very good control of the temperature of the instrument and of the composition of the mobile phase. Because of their sensitivity to mobile phase composition it is very difficult to do gradient elution work, and they are generally held to be unsuitable for this purpose. [Pg.73]

Some commercially available detectors have a number of detection modes built into a single unit. Fig. 2.4o is a diagram of the detector used in the Perkin Elmer 3D system, which combines uv absorption, fluorescence and conductivity detection. The uv function is a fixed wavelength (254 nm) detector, and the fluorescence function can monitor emission above 280 nm, based on excitation at 254 nm. The metal inlet and outlet tubes act as the electrodes in the conductance cell. The detection modes can be operated independently or simultaneously, using a multichannel recorder. In the conductivity mode, using NaCl, a linear range of 103 and a noise equivalent concentration of 5 x 10 8 g cm-3 have been obtained. [Pg.74]

Detector Minimum detectable Linear range Temperature Remarks... [Pg.101]


See other pages where Detector linear range is mentioned: [Pg.111]    [Pg.111]    [Pg.111]    [Pg.111]    [Pg.212]    [Pg.570]    [Pg.571]    [Pg.578]    [Pg.398]    [Pg.225]    [Pg.240]    [Pg.48]    [Pg.284]    [Pg.86]    [Pg.49]    [Pg.146]    [Pg.150]    [Pg.155]    [Pg.292]    [Pg.664]    [Pg.808]    [Pg.16]    [Pg.20]    [Pg.194]    [Pg.51]    [Pg.54]    [Pg.380]    [Pg.194]    [Pg.326]    [Pg.340]    [Pg.432]    [Pg.102]    [Pg.102]    [Pg.104]    [Pg.104]    [Pg.112]    [Pg.113]   
See also in sourсe #XX -- [ Pg.119 ]

See also in sourсe #XX -- [ Pg.163 , Pg.164 ]




SEARCH



Detector, characteristics linear range

Detector, linear

Detectors linearity

Detectors range

Detectors, HPLC linear range

Linear range

Linear range of detectors

Linear range, chromatographic detectors

Linearity range

Specifications detector linear range

The Linear Dynamic Range of a Detector

© 2024 chempedia.info