Columns dynamic coating

Open tubular columns can be coated internally with the appropriate stationary phase dissolved in a suitable solvent, or, alternatively, with monomeric materials that are subsequently polymerized to form a relatively rigid polymer film on the column walls. There are two basic methods for depositing the stationary phase as a surface film on a capillary column, dynamic coating and static coating, the latter being the most commonly used. [Pg.125]

Factors Controlling Selectivity Open Tubular Columns Dynamic Coating Static Coating Column Regeneration Capillary Column Design and Choice Stationary Phases for the GC Separation of Chiral Substances Synopsis References Chapter 6... [Pg.544]

Compared to wall-coated columns, dynamically coated SCOT columns have higher capacity ratios. [Pg.223]

Poole, S. K., Patel, S., Dehring, K., Workman, H., Dong, J. Estimation of octanol-water partition coefficients for neutral and weakly acidic compounds by microemulsion electrokinetic chromatography using dynamically coated capillary columns. J. Chromatogr. [Pg.355]

Carbowax 20M, polysiloxanes, and N-cyclo-3-azetidinol are the most widely used sutetances for the thermal degradation method [143,180,192-194]. In the case of the Carbowax treatment deactivation can be carried out in either of two ways. The column can be dynamically coated with a solution of Carbowax 20H in a volatile solvent, excess solvent evaporated with a stream of nitrogen, the column ends sealed and the column heated at about... [Pg.77]

FIGURE 14.9 Separation of Mn(II), Zn(II), Cd(II), and Pb(II) using (a) a dynamically coated and (b) a pre-coated MTB ODS reversed-phase column. (From Pauli, B. and Haddad, P.R., Trends Anal. Chem., 18, 107, 1999. Copyright 1999. With permission from Elsevier.)... [Pg.404]

Fig. 5.1. Separation of chlorophenol acetates (2-0.02 ng). Peaks 1 = 2-chloro 2 = 3-chloro 3 = 4-chloro 4 = 2,6-dichloro 5 = 2,5-dichloro 6 = 2,4-dichloio 7 = 3,4-dichloro 8 = 2,3-dichIoro 9 = 3,5-dichloro 10 = 2,4,6-trichloro 11 = 2,4,5-trichloro 12 = 2,3,4,6-tetrachloro 13 = pentachloro-phenol acetates. Conditions Pyrex glass column (25 m X 0.35 mm I.D.), dynamically coated with SE-30 temperature programme, 3°C/min (95-180°C) helium flow-rate, 2-3 ml/min splitting flow-rate, 0-60 ml/min. (Reproduced from J. Chromatogr. 131 (1977) 412.)... Fig. 5.1. Separation of chlorophenol acetates (2-0.02 ng). Peaks 1 = 2-chloro 2 = 3-chloro 3 = 4-chloro 4 = 2,6-dichloro 5 = 2,5-dichloro 6 = 2,4-dichloio 7 = 3,4-dichloro 8 = 2,3-dichIoro 9 = 3,5-dichloro 10 = 2,4,6-trichloro 11 = 2,4,5-trichloro 12 = 2,3,4,6-tetrachloro 13 = pentachloro-phenol acetates. Conditions Pyrex glass column (25 m X 0.35 mm I.D.), dynamically coated with SE-30 temperature programme, 3°C/min (95-180°C) helium flow-rate, 2-3 ml/min splitting flow-rate, 0-60 ml/min. (Reproduced from J. Chromatogr. 131 (1977) 412.)...
Columns for LLPC are prepared by using solvent evaporation, direct coating, precipitation, or a dynamic coating technique [3], In the solvent-evaporation technique, the nonvolatile hquid stationary phase is dissolved in a volatile solvent and mixed with the dry sup-... [Pg.961]

R. M. Cassidy and S. Elchuk, Dynamically coated columns for the separation of metal ions and anions by ion chromatography, AnaZ. Chem. 54 1558 (1982). [Pg.983]

Akatsuka, K., Yoshida, Y., Nobuyama, N., Hoshi, S., Nakamura, S. and Kato, T., 1998, Preconcentration of Trace Cadmium from Seawater Using a Dynamically Coated Column of Quaternary Ammonium Salt on C18-bonded Silica Gel and Determination by Graphite-furnace Atomic Absorption Spectrometry , Analytical Sciences, 14, 529-533. [Pg.28]


See other pages where Columns dynamic coating is mentioned: [Pg.84]    [Pg.400]    [Pg.233]    [Pg.84]    [Pg.400]    [Pg.233]    [Pg.418]    [Pg.78]    [Pg.79]    [Pg.80]    [Pg.595]    [Pg.596]    [Pg.598]    [Pg.598]    [Pg.72]    [Pg.395]    [Pg.398]    [Pg.399]    [Pg.402]    [Pg.404]    [Pg.1216]    [Pg.120]    [Pg.154]    [Pg.249]    [Pg.49]    [Pg.126]    [Pg.182]    [Pg.651]    [Pg.430]    [Pg.217]    [Pg.292]    [Pg.961]    [Pg.1067]    [Pg.1067]    [Pg.1067]    [Pg.298]    [Pg.264]    [Pg.170]    [Pg.183]    [Pg.184]    [Pg.185]    [Pg.112]    [Pg.147]   