Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Baseline calibrations

Of all the requirements that have to be fulfilled by a manufacturer, starting with responsibilities and reporting relationships, warehousing practices, service contract policies, airhandUng equipment, etc., only a few of those will be touched upon here that directly relate to the analytical laboratory. Key phrases are underlined or are in italics Acceptance Criteria, Accuracy, Baseline, Calibration, Concentration range. Control samples. Data Clean-Up, Deviation, Error propagation. Error recovery. Interference, Linearity, Noise, Numerical artifact. Precision, Recovery, Reliability, Repeatability, Reproducibility, Ruggedness, Selectivity, Specifications, System Suitability, Validation. [Pg.138]

Ar as carrier gas. The library is redundant and mirrored at the diagonal while the diagonal itself does not contain any catalytically active material to allow baseline calibration. (2,1) element is pure Rh, (12,1) element is pure Pt and (12,11) element is pure Pd. The numbers in the x,y-plane indicate the catalyst composition and the numbers in z-direction indicate CO signal intensity [43] (by courtesy of Elsevier Ltd.). [Pg.458]

It is important to have some validation tests when things start to (or appear to) go wrong. These validation tests can be used to check the calibration and performance of the DSC system. They should be run on a periodic basis, and the results kept as an historical log of the DSC system performance. There are three validation tests that are advisable to run on a regular basis baseline, calibration, and performance. The data from the validation tests are invaluable for assessing system performance when a strange or erroneous result presents itself. [Pg.46]

Full temperature range calibration Enthalpy calibration Baseline calibration Check flow meters and gas bottles Back-up data and methods Monthly... [Pg.50]

Calibration is a potentially complex subject as there can be cell asymmetries in terms of heat capacity, thermal resistances, and other effects such as convection and emissivity. However, here we shall confine ourselves to a simple procedure that will be effective in most cases. We can consider that calibration of an MTDSC consists of three steps (1) heat flow calibration (by calculation of a cell constant), (2) baseline calibration, and (3) heat capacity calibration (10). [Pg.112]

For the underlying signal, baseline calibration is carried out in essentially the same manner as in a conventional DSC the baseline (obtained with empty matched pans) underlying heat flow can simply be subtracted from the underlying heat flow for the sample. An alternative is to mathematically fit a polynomial through this baseline and subtract this fitted curve from subsequent runs. The benefit of the latter approach is that it reduces noise. [Pg.113]

The DSC system is usually calibrated in several ways. Baseline calibration is performed with no pans in place. The calibration measures the baseline slope and offset over the temperature range of interest. The computer system controlling the DSC stores these values and subtracts ba.scline slope and offset from subsequent sample runs to minimize their effects. [Pg.903]

According to the earthquake record monitored by Chengdu station (CD2) of Chinese National Earthquake Bereau, the Wenchuan earthquake lasted for about 160 seconds and strong part are nearly 20 s. In this paper, earthquake wave of the front 20 seconds is used. Seismosignal software is used to make the filtration with range of 0.1-10 Hz and then do the linear baseline calibration. The velocity time history should be first transferred to velocity-stress time history and then set in the bottom viscous boundary. Stress time history is shown in Figure 7. [Pg.149]

After baseline calibration is performed, heat flow calibration is done by melting a known quantity of a material with a well-known heat of fusion. Indium is the most often used standard. Indium is placed in the sample pan and scanned against an empty reference pan. The area of the melting peak is related to the known enthalpy of fusion by a calibration factor known as the celt constant. This procedure also calibrates the temperature axis from the known melting temperature of indium. Temperature calibration should also be performed over a wider temperature range, by measuring the melting points of several well-known standards. [Pg.987]

Figure 25 contains plots of the pure component spectra for the two calibrations. It is apparent that, in the absence of the extraneous absorbances from Component 4, CLS is now able to do a good job of estimating the pure component spectra. However, even with nonzero intercepts, CLS is unable to remove the sloping baseline from the spectra. Both calibrations distributed most of the baseline effect onto the spectrum for Component 2 and some onto the Component 3 spectrum. [Pg.68]

Artifact removal and/or linearization. A common form of artifact removal is baseline correction of a spectrum or chromatogram. Common linearizations are the conversion of spectral transmittance into spectral absorbance and the multiplicative scatter correction for diffuse reflectance spectra. We must be very careful when attempting to remove artifacts. If we do not remove them correctly, we can actually introduce other artifacts that are worse than the ones we are trying to remove. But, for every artifact that we can correctly remove from the data, we make available additional degrees-of-freedom that the model can use to fit the relationship between the concentrations and the absorbances. This translates into greater precision and robustness of the calibration. Thus, if we can do it properly, it is always better to remove an artifact than to rely on the calibration to fit it. Similar reasoning applies to data linearization. [Pg.99]

LOD) calculate and display the limits of detection and quantitation LOD, LOQ. [Note This form of calculating the LOD or LOQ was chosen because the results are influenced not only by the noise on the baseline, but also by the calibration design from the educational point of view this is more important than the consideration whether any agency has officially adopted this or that LOD-model. For a comparison, see Figs. 2.14, 2.15, and 4.31]. [Pg.375]

Relatively long time-constants of 1 to 60 s are used in order to minimize the noise inherent in c.d. spectra. Thus, the scanning rate is low, and as c.d. instruments tend to drift with time, they should be calibrated daily, and baselines should be measured for each sample. [Pg.78]

The detector was calibrated by pxm ing solutions of sodium dichromate of known absorbance through the sample port of the detector. The solutions were prepared in the carrier fluid which served as reference. The recorder response was measured as the ultimate height reached on the chart paper above the baseline when the sample fluid was switched to a sodiiam dichromate solution of known absorbance. The calibration was insensitive to flow-rate variations. [Pg.52]

Peedc-to-peak resolution in SEC can be calculated by the ratio of peak separation at the peak maxiaut to the sum of the baseline peak widths. This general definition of resolution is less useful in SEC, where a measure of the ability of the column to separate solutes of different molecular weight is required. For this purposes, we define a new term, the specific resolution factor, R, which relates peak resolution to sample molecular weight, assuming all measurements are made within the linear region of the molecular weight calibration curve, equation (4.41)... [Pg.739]

The IR spectra are finally analyzed to determine the effluent concentration from each reactor channel. The quantification of species concentration is performed using either univariate or multivariate calibration methods. For non-overlapping peaks, like CO, C02, and N20, we can use univariate calibration. This is simply performed by baseline correction, the peak areas and/or peak heights and then converting these values... [Pg.329]

The baseline is chosen to be a straight line connecting these two points in the chromatogram. Additional schemes for dealing with the baseline are available for specimens containing large amounts of low molecular weight species. The results obtained for such specimens are, however, dubious because of problems associated with calibration and because of interference from injection artifacts... [Pg.135]

Figure 5. A typical chromatogram. The solid curve is the part of the data that is used to characterize the molecular weight distribution. The dashed portions represent data that are not used. The solid straight line represents the baseline under the chromatographic peak. The vertical lines define the limits of the calibration function for the column set. Figure 5. A typical chromatogram. The solid curve is the part of the data that is used to characterize the molecular weight distribution. The dashed portions represent data that are not used. The solid straight line represents the baseline under the chromatographic peak. The vertical lines define the limits of the calibration function for the column set.

See other pages where Baseline calibrations is mentioned: [Pg.389]    [Pg.389]    [Pg.114]    [Pg.903]    [Pg.333]    [Pg.831]    [Pg.81]    [Pg.43]    [Pg.389]    [Pg.389]    [Pg.114]    [Pg.903]    [Pg.333]    [Pg.831]    [Pg.81]    [Pg.43]    [Pg.141]    [Pg.240]    [Pg.756]    [Pg.757]    [Pg.44]    [Pg.277]    [Pg.82]    [Pg.115]    [Pg.138]    [Pg.26]    [Pg.300]    [Pg.372]    [Pg.222]    [Pg.930]    [Pg.1152]    [Pg.638]    [Pg.45]    [Pg.208]    [Pg.185]    [Pg.187]    [Pg.197]    [Pg.139]   
See also in sourсe #XX -- [ Pg.651 ]




SEARCH



Baseline

© 2024 chempedia.info