Use a methanol or acetonitrile and water mixture without buffer (e.g., 50%) as the flush solvent for most autosamplers. Degas it to avoid bubbles. [Pg.262]

Fill the vial with enough sample solution for all injections. Note that some antosamplers nse a side-port needle that might reqnire at least a 0.5 mL sample volnme in a 2-mL vial (e.g., Waters Alliance). [Pg.262]

Purge the injector daily and before sample analysis to remove bubbles in the sampling syringe. This is critical for sampling precision. [Pg.262]

Typical injection volnmes for pharmacentical assays range 5-50 pL. Injection sizes of 5pL might lead to poorer precision. [Pg.262]

Put a warning label ontside the antosampler if a non-standard sampling syringe or loop is installed. [Pg.262]