Experiments 10-27 are designed to check the autosampler injection precision, pump repeatability and detector/system linearity. One programs the system to automatically inject multiple replicate volumes of a certified test standard. One typically injects 6-10 replicates per volume. The standard component s peak areas are used for calculated injection precision (reproducibility) and system linearity whereas, the retention times are used to calculate pump repeatability. [Pg.329]

Eor injection precision test details and calculations, refer to the procedure described earlier under Autosampler in section (j) of the OQ guidelines. [Pg.329]

It is recommended that OQ test the following on an HPLC system flow accuracy, pump compositional accuracy, pressure pulsations, column oven temperature accuracy/stability, detector noise/drift and wavelength accuracy, autosampler injection precision and carryover. [Pg.333]

For autosampler precision, 10 consecutive lO-pL injections of an eth-ylparaben solution (20 J,g/mL) are used (Figure 6). A Waters Symmetry column packed with 5- J,m particles is used. The manufacturer s specification for peak area precision at 0.5% RSD is adopted as the acceptance criterion. This stringent precision criterion is required for precise assay testing of drug substances typically specified at 98-102% purity. The linearity test is performed by single injections of 5, 10, 40, and 80 pL of the... [Pg.296]

Finally, Experiment 28 is used to check for any autosampler/ system sample carryover. A blank water sample is injected immediately following the highest injected volume of the preceding precision/repeatability/linearity test. Any carryover is then calculated as in the procedure described earlier under Autosampler in section (j) of the OQ guidelines. [Pg.330]

An autosampler allows the automatic sample injection from sample vials or 96-well microplates (Figure 4.7a). An autosampler reduces labor cost and increases productivity and testing precision. [Pg.85]

Flow precision Determine retention time RSD of six 10-pL injections of ethylparaben (same as in autosampler test) RSD <+0.5%... [Pg.228]