Starting a FIA system

It is always advisable to perform a number of functional tests and dispersion measurements on a newly designed manifold. This is helpful if the user is not familiar with the technique. The time so spent will result in a better understanding of the physical functions of the system so that problems, connected with the chemical reactions on which the analytical procedure is based, will be identified and readily solved. [Pg.299]

As repeatedly emphasized in this book, the dispersion coefficient or D is the key parameter in FIA, as it indicates the degree of dilution of the original sample solution and thus the extent to which the injected [Pg.299]

Carrier Stream Solution. 1 x 10 M aqueous borax solution. [Pg.301]

Dye Solution. A dye stock solution is prepared by dissolving 0.400 g of bromothymol blue in 25 mL of 96% ethanol, making the final volume up to 100 mL with the 1 x 10 M borax solution. The dye solution used is then prepared by mixing 1 mL of the stock dye solution with 199 mL of the 1 X 10 M borax solution (the absorbancy of this solution m a 10-mm-long cell is 1.2 A). [Pg.301]