Spectrum synchronous, luminescence

Commercial spectrometers, such as the Perkin-Elmer MPF-43A fluorescence spectrometer, that allow interlocking of excitation and emission monochromators lately have become available for utilizing this underexploited analytical technique. The synchronous luminescence technique reduces the complexity of the luminescence spectrum of a compound compared with a conventionally obtained luminescence spectrum. One can, therefore, better tackle the analysis of fairly complex mixtures without resorting to techniques that are expensive or excessively time consuming. [Pg.86]

Some luminescence instruments allow simultane-ously scanning both the excitation and the emission wavelengths with a sntall w avelength difference between them. The spectrum that results is known as a synchronous spectrum. A luminescence signal Is ob-... [Pg.410]

Molecular fluorescence spectroscopy is a commonly employed analytical method that is sensitive to certain chemical properties of FA (9-13). Fulvic acid s molecular fluorescence is principally due to conjugated unsaturated segments and aromatic moieties present in the macromolecule (14). Several types of fluorescence spectra can be measured, including an excitation emission matrix or total luminescence spectrum, constant offset synchronous fluorescence, excitation spectra, and emission spectra, furnishing the researcher with useful data. The ability to resolve and select multiple fluorescent species makes these approaches extremely useful for studying FA relative to its chemical reactivity. [Pg.109]

The excitation wavelength selector can be either a filter or a monochromator. Filters offer better detection Hmits, but do not provide spectral scanning capabilities. Often, a filter is used in the excitation beam along with a monochromator in the emission beam to allow emission spectra to be acquired. FuU emission and excitation spectral information can be acquired only if monochromators are used in both the excitation and emission beams. In modern instruments with array detectors, a polychromator is used in the emission beam instead of a monochromator. Recent research instraments are able to scan both wavelengths automatically and combine all data into a 2D excitation—emission spectrum. In lifetime spectrometers, a pulsed light source and a gated detector are synchronized in order to measure the time dependence of the luminescence emission. [Pg.67]


See other pages where Spectrum synchronous, luminescence is mentioned: [Pg.88]    [Pg.599]    [Pg.432]    [Pg.88]    [Pg.19]    [Pg.1342]    [Pg.741]    [Pg.442]   
See also in sourсe #XX -- [ Pg.410 ]

See also in sourсe #XX -- [ Pg.410 ]