Reconstruction, image filtered backprojection method

The Filtered Backprojection (FBP) method may be used to process by reconstructing the original image from its projections in two steps Filtering and Backprojection. [Pg.176]

The first point has not been clearly recognized or appreciated in the early days of the method, so terms like radon transformation aad filtered backprojection were introduced [Herl, Manl], In practical realizations of image reconstruction from projections, however, numerical frers must, indeed, be used [Herl]. [Pg.201]

Figure 4.10. Comparison of filtered backprojection (FBP) and iterative (OSEM) methods with attenuation correction, (a) Lungs, (b) Normal Liver, (c) Liver with tumor, (d) Breast. FBP images with attenuation corrections are noisier than OSEM images with attenuation correction. (Reprinted by permission of Society of Nuclear Medicine from Riddell C et al (2001) Noise reduction in oncology FDG PET images by iterative reconstructions a quantitative assessment. J Nucl Med 42 1316)... Figure 4.10. Comparison of filtered backprojection (FBP) and iterative (OSEM) methods with attenuation correction, (a) Lungs, (b) Normal Liver, (c) Liver with tumor, (d) Breast. FBP images with attenuation corrections are noisier than OSEM images with attenuation correction. (Reprinted by permission of Society of Nuclear Medicine from Riddell C et al (2001) Noise reduction in oncology FDG PET images by iterative reconstructions a quantitative assessment. J Nucl Med 42 1316)...
The filtered backprojection can be applied to 3D image reconstruction with some manipulations. The 3D data sinograms are considered to consist of a set of 2D parallel projections, and the FBP is applied to these projections by the Fourier method. The iteration methods also can be generally applied to the 3D data. However, the complexity, large volume, and incomplete sampling of the data due to the finite axial length of the scanner are some of the factors that limit the use of the FBP and iterative methods directly in 3D reconstruction. To circumvent these difficulties, a modified method of handling 3D data is commonly used, which is described below. [Pg.82]


See other pages where Reconstruction, image filtered backprojection method is mentioned: [Pg.137]    [Pg.138]    [Pg.71]    [Pg.82]    [Pg.611]    [Pg.175]   
See also in sourсe #XX -- [ Pg.72 ]