Pulse Sampling or Gated Detection

While TCSPC andFD fluorometry are the dominant methods used by biochemists, there are alternative methods for measuring intensity decays. Prior to the introduction of [Pg.116]

Gated detection can be accomplished in two ways. One method is to turn on or gate the g n of the detector for a short period during the intensity decay. Surprisingly, thb can be accomplished on a timescale adequate for measurement of nanosecond lifetimes. Alternatively, the detector can be on during the entire decay, and the electric cal pulse measured with a sampling oscilloscope. Such devices can sample electrical signals with a resolution of tens of picoseconds. [Pg.117]

In recent years there has been a reintroduction of g ed detection methods. The time resolution can be good but is not con arable to that of a laser source and an MCT PMT. Typical instrument response functions are close to 3 [Pg.117]