Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Probe tip capacitance

So the measurement becomes less invasive. But with this setting, you also end up worsening the signal-to-noise ratio. In particular, when the scope increases its gain automatically (to compensate for the 10 1 mode), it ends up bringing up its own noise floor too. So the 1 1 mode is less noisy, inherently so. Unfortunately the probe tip capacitance of 30pF can create its own problems and can often quell the very noise you are trying to measure. [Pg.205]

In SFM, the probe tip is mounted on a highly sensitive, cantilever-type spring. The force of interaction between the sample and the tip can be calculated from the spring constant and the measured deflection of the cantilever. The deflection is sensed using the STM principle (Vignette 1.8) or capacitance or optical methods. The SFM can be operated in the contact regime or like the SFA. In the latter mode, one can measure van der Waals forces (see Chapter 10), ion-ion repulsion forces (see Chapter 11), and capillary forces and frictional forces, among others. In contrast to STM, the SFM can be used for both conductors and... [Pg.55]

An equivalent circuit of the three-electrode cell discussed in Chapters 6 and 7 is illustrated in Figure 9.1. In this simple model, Rr is the resistance of the reference electrode (including the resistance of a reference electrode probe, i.e., salt bridge), Rc is the resistance between the reference probe tip and the auxiliary electrode (which is compensated for by the potentiostat), Ru is the uncompensated resistance between the reference probe and the working-electrode interphase (Rt is the total cell resistance between the auxiliary and working electrodes and is equal to the sum of Rc and Ru), Cdl is the double-layer capacitance of the working-electrode interface, and Zf is the faradaic impedance of the electrode reaction. [Pg.268]

Electrical measurement of the dielectric constant is done through the fabrication of metal—oxide—semiconductor capacitor structures, where the ULK serves as the dielectric of the capacitor. A doped Si wafer is used as the substrate, on which the ULK film is deposited. This ULK film is subjected to CMP, say, or any other process whose impact on ULK characteristics needs to be quantified. An aluminum film is deposited on the backside of the Si wafer to form one of the capacitor contacts. Using a shadow mask, aluminum dots of varying diameters are evaporated onto the surface of the ULK film, to form the other terminal of the capacitor. Each aluminum dot is probed to measure its capacitance (at about 100 kHz). Evaporation through a shadow mask allows for the formation of metal contacts without altering the dielectric further— as would be the case if reactive-ion-etch were used to form the contacts. (It should be noted that more complex process flows can be used to eliminate concerns such as dot-size variation, the effect of probe-tip impact on the dielectric being tested, etc.) The results of electrical measurement of the k-value increase post-CMP of the variety... [Pg.102]

Fluid-level detectors which use capacitive measurement ensure sufficient reagent and sample volumes. These detectors are integrated into the reagent probe and sample tip and require no maintenance. [Pg.40]

The cantilever bending-technique requires a sensitive displacement detection such as a capacitance probe (Klokholm 1976, 1977), optical interferometry (Sontag and Tam 1986), a tunnelling tip (Wandass et al. 1988) or angular detection (e.g. laser beam deflection, Son-tag and Tam 1986 Trippel 1977 Tam and Schroeder 1988 Betz 1997 Sander et al. 1998). [Pg.106]

Figure 16.1 shows the system setup of the sndm using the lc lumped constant resonator probe [4], In the figure, Cs(t) denotes the capacitance of the specimen under the center conductor (the tip) of the probe. Cs (t) is a function of time because of the nonlinear dielectric... [Pg.304]

Apart from resistance, capacitive effects should also be taken into consideration. The best design of a Luggin probe is one with a narrow capillary at its tip with thin walls to prevent shielding, but with thick walls in the main body which widen rapidly away from the tip to reduce resistance in the control loop [21]. [Pg.46]

Nanocell is the smallest electrochemical cell developed by Sugimura and Nakagiri [11] and further developed and utilized for ENT by BloeB et al. [10]. The nanocell consists of two electrodes distance between electrodes is generally maintained in the order of less than 1 nm. In between two electrodes, absorbed water film acts as an electrolyte whose volume is maintained by vapor pressure and ranges from 10 to 10 cm. Double layer capacitance is not formed across the solid liquid interface in the nanocell due to the much smaller inter-electrode gap and hence, generated hydrogen ion and hydroxyl ion recombine immediately. Nanotip of microtool such as tip of scanning probe microscope (SPM) or AFM tip is most suitable for the formation of electrochemical nanoceU. [Pg.244]

The probe motion sensor controls the force acting between the tip and the surface feedbacking a correction signal for the vertical positioning of the probe relatively to the sample, to keep the force or the distance constant. An optical beam deflection system (optical lever) is often applied for this purpose, ensuring low noise, stability, and versatiUty. The trend now is to replace the optical lever with self-sensing means like, for instance, piezoelectric, piezoresistive, or capacitive. [Pg.64]

Figure 7.42 Kelvin probe principles. Tip and specimen are capacitively coupled. Charges on the tip changes with probe oscillation. Vs = voltage difference between tip and sample. Figure 7.42 Kelvin probe principles. Tip and specimen are capacitively coupled. Charges on the tip changes with probe oscillation. Vs = voltage difference between tip and sample.
Fig. 11.5 - Simple potentiostat for maintaining a constant potential between a reference electrode, RE, and working electrode, WE is the solution resistance between the reference electrode probe and the counter electrode, CE the uncompensated resistance between the Luggin tip and the working electrode and Qi the double layer capacitance of the working electrode. Fig. 11.5 - Simple potentiostat for maintaining a constant potential between a reference electrode, RE, and working electrode, WE is the solution resistance between the reference electrode probe and the counter electrode, CE the uncompensated resistance between the Luggin tip and the working electrode and Qi the double layer capacitance of the working electrode.

See other pages where Probe tip capacitance is mentioned: [Pg.182]    [Pg.167]    [Pg.167]    [Pg.355]    [Pg.182]    [Pg.167]    [Pg.167]    [Pg.355]    [Pg.333]    [Pg.8]    [Pg.8]    [Pg.205]    [Pg.143]    [Pg.316]    [Pg.318]    [Pg.336]    [Pg.190]    [Pg.190]    [Pg.251]    [Pg.563]    [Pg.136]    [Pg.305]    [Pg.217]    [Pg.187]    [Pg.347]    [Pg.353]    [Pg.300]    [Pg.410]    [Pg.425]    [Pg.456]    [Pg.509]    [Pg.287]    [Pg.370]    [Pg.340]    [Pg.350]    [Pg.425]    [Pg.39]    [Pg.592]   
See also in sourсe #XX -- [ Pg.167 , Pg.190 ]

See also in sourсe #XX -- [ Pg.167 , Pg.190 ]

See also in sourсe #XX -- [ Pg.167 , Pg.190 ]




SEARCH



Capacitance probes

Probe tip

© 2024 chempedia.info