Practical Considerations in the Design of Solid Phase FI Optosensing Systems

2 Practical Considerations in the Design of Solid Phase FI Optosensing Systems [Pg.125]

Techniques and manifolds used for other on-line soiption separation methods described in the previous sections may be used to advantage for solid phase optosensing systems. [Pg.125]

A flow-cell/column packed with C 8 used by Ruzicka et al.[60,61] for the determination of phosphate by absorptiometry is shown in Fig. 4.12. The light was transmitted radially through the column, which had an i.d. of only 2 mm, at a level where the bulk of the analyte was retained. [Pg.125]

Owing to light attenuation in such systems, the light source of the spectrophotometer may have to be modified to provide higher intensities. With double-beam spectrophotometers equipped with photomultipliers, normally the detection system should be sensitive enough for such applications, however, the reference beam will have to be attenuated by a suitable means to balance the two beams. [Pg.125]