For high-frequency measurements, normal photomultipliers are too slow, and microchannel plate photomultipliers are required. However, internal crosscorrelation is not possible with the latter and an external mixing circuit must be used. [Pg.180]

The time resolution of a phase fluorometer using the harmonic content of a pulsed laser and a microchannel plate photomultiplier is comparable to that of a single-photon counting instrument using the same kind of laser and detector. [Pg.180]


Phase Fluorometers using the Harmonic Content of a Pulsed Laser... [Pg.237]

Historically, the first instrument for the determination of lifetime was a phase fluorometer (designed by Gaviola in 1926) operating at a single frequency. Progress in instrumentation enabled variable modulation frequency by employing a cw laser (or a lamp) and an electro-optic modulator (0.1-250 MHz), or by using the harmonic content of a pulsed laser source (up to 2 GHz). These two techniques will now be described. [Pg.178]