New Aftematives to Applanation Tonometry. In an effort to address the biomechanical variables associated with the measurement of lOP, new lOP measuring devices have been developed and introduced.Two notable instruments are the dynamic contour (Pascal) tonometer and the ocular response analyzer (Reichert) tonometer. These technologies... [Pg.673]

Ocular Response Analyzer. The ocular response analyzer applies force to the cornea in the form of a collimated air pulse, with an electrooptical system used to monitor changes in curvature during corneal deformation. The cornea moves inward with the air pulse and then returns to normal curvature. The curvature detection system records two pressure values at inward and outward applanation events. Corneal biomechanical properties create a damping effect that manifests as a difference between the two pressures. Averaging these two pressures provides a Goldmann-correlated lOE The difference between these two pressure values is referred to as corneal hysteresis. [Pg.674]

Luce DA. Introduction to the ocular response analyzer. Optometric Glaucoma Society E-Journal 2006 2(1). [Pg.697]