All of the units are installed in a single 19-in portable rack and the total weight is about 80 kg. [Pg.84]

The Larmor frequency rf reference signal is generated using a DDS (direct digital synthesizer) board (FSW01, DST Inc., Asaka, Japan). This board can generate coherent and spectrally pure rf signals from 5 to 200 MHz, of which the frequency is controlled via the ISA bus. This board has an essential role in the NMR lock process. [Pg.84]

The data acquisition is performed using an ADC board (PC-414G3, DATEL, USA). Because the ADC board has 14-bit resolution, simultaneous two-channel acquisition capability up to 1 MHz sampling rate and 32 Kword FIFO buffer memory, it has sufficient performance for most MRI applications. [Pg.84]

The PC system runs under a Windows95/98/2000 operating system and is capable of real-time image acquisition, reconstruction and display (16, 17). [Pg.85]

The rf transmitter amplifies an rf pulse signal of about 1 mW up to several W or up to several kW. The amplifier should work in a linear mode (class AB) because excitation pulse shape for slice selection must be reproduced. Class AB rf transmitters such as these with blanking gates are widely available commercially. [Pg.86]


Figure 2.2.3 shows a typical block diagram for a compact MRI system. The square box surrounded by the dotted lines shows the electrical sub-system, which is stored in a single portable rack as shown in Figure 2.2.2. The electrical sub-system is thus often called the (portable) MRI console. In accord with Figure 2.2.3, an MR image acquisition process is described as follows. Figure 2.2.3 shows a typical block diagram for a compact MRI system. The square box surrounded by the dotted lines shows the electrical sub-system, which is stored in a single portable rack as shown in Figure 2.2.2. The electrical sub-system is thus often called the (portable) MRI console. In accord with Figure 2.2.3, an MR image acquisition process is described as follows.
Figure 2.2.9 shows a typical portable MRI console. The MRI console consists of a digital system (PC), MRI transceiver, 3 channel gradient driver and rf transmitter. [Pg.83]

The bottom image in Figure 2.2.15 is a maximum intensity projection of a 100 xm cubic-resolution 3D-image dataset of a blue berry acquired with this system. The network structure of the fruit is clearly visualized. If the MRI console and permanent magnet can be made smaller, a true desktop or handy M R microscope will then be constructed. [Pg.88]

Because the major cost of a commercial MRI is the magnet, there have been attempts to develop portable MRI consoles, that is, the requisite instrumentation for MRI minus the magnet, that could be transported to a medical MRI facility so that a scientific experiment could be set up independently of the medical MRI console (see Chapter 2.2). There are also efforts to develop compact NMR/MRI instruments for non-traditional uses. Granular studies would easily fit into this category. [Pg.506]


See other pages where MRI Console is mentioned: [Pg.83]    [Pg.84]    [Pg.83]    [Pg.84]    [Pg.180]    [Pg.319]    [Pg.30]    [Pg.218]    [Pg.73]   