Mass Spectrometer for Isotope Ratio Analysis

Isobaric interferences from other isotope species at the target masses require the measurement of more masses than just the targeted isotope ratio for a necessary correction. For the measurement of 8 C, three collectors at m/z 44,45 and [Pg.280]

46 are necessary (see Table 2.49). Algorithms for a fully automated correction of isobaric ion contributions are implemented in modern isotope ratio MS data systems, for example, for the isobaric interferences of and on m/z46 of CO2, see Table 2.50 (Craig, 1957). Other possible interferences such as CO on Nj, N2O on CO2 can be taken care of by the interface technology. [Pg.281]

Standardization in isotope ratio monitoring measurements should be done exclusively using the principle of Identical Treatment of reference and sample material , the IT Principle . Mostly, isotopic referencing is made with a co-injected peak of standard gas. [Pg.281]